510(k) Report

510(k) · Cleared

Recalled · Class III · 2012

Stratus CS Acute Care D-Dimer Testpak, Dilpak, Calpak

K063356510(k) TraditionalCleared Mar 23, 2007136-day reviewClass II21 CFR 864.7320Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dade Behring, Inc. · DAP Fibrinogen And Fibrin Split Products, Antigen, Antiserum, Control

Cleared on Mar 23, 2007 as substantially equivalent to K022976 (Dade Behring, Inc., 2003) and 2 other predicates after a 136-day review. Its family tree reaches back 4 generations to 1986. 1 later device descends from it. It has been recalled once.

Decided
Mar 23, 2007
Review
136 days
Class
Class II
Product code
DAP
Recalls
1
Predicates
3

Its longest line of descent runs back 4 generations to K862156 (1986). 5 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 4 generations backK862156 Asserachrom D-DI KitAmerican Bioproducts Co. · 1986
  2. 3 generations backK973819 Vidas D-Dimer (DD) AssayBiomerieux Vitek, Inc. · 1998
  3. 2 generations backK022976 Stratus CS D-Dimer (DDMR) Method (Testpak & Dilpak), Models Cddmr &…Dade Behring, Inc. · 2003 · Recalled
  4. 1 generation backK051597 Stratus CS D-DimerDade Behring, Inc. · 2005 · Recalled
  5. This deviceK063356 Stratus CS Acute Care D-Dimer Testpak, Dilpak, CalpakDade Behring, Inc. · 2007
  6. Descendant · cites it directly · 0 of its ownK081732 Innovance D-Dimer and Innovance D-Dimer ControlsSiemens Healthcare Diagnostics · 2008

Most influential descendants

Showing 1 of 1 · See all
  • K081732Innovance D-Dimer and Innovance D-Dimer ControlsSiemens Healthcare Diagnostics · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K862156Asserachrom D-DI KitAmerican Bioproducts Co.1986—
−3K973819Vidas D-Dimer (DD) AssayBiomerieux Vitek, Inc.1998—
−2K022976Stratus CS D-Dimer (DDMR) Method (Testpak & Dilpak), Models Cddmr & Cddmr-DDade Behring, Inc.20031
−1K051597Stratus CS D-DimerDade Behring, Inc.20051
0 (this device)K063356Stratus CS Acute Care D-Dimer Testpak, Dilpak, CalpakDade Behring, Inc.2007—
+1K081732Innovance D-Dimer and Innovance D-Dimer ControlsSiemens Healthcare Diagnostics2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
FIG. 1 — Lineage of K063356, 1986–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DAP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DAP device

  1. #1 K253985 INNOVANCE® D-Dimer 2.0 (SMN 10873803)score 13.0 · 5 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022976Stratus CS D-Dimer (DDMR) Method (Testpak & Dilpak), Models Cddmr & Cddmr-D
Summary page 2
…ilPak 3. Identification of the Predicate Device * Dade Behring Stratus® CS DDMR TestPak and DilPak K022976/K051597 * Dade Behring Stratus® CS DDMR CalPak K022977 FDA Guidance Document(s):…
Dade Behring, Inc.20031
*K051597Stratus CS D-Dimer
Summary page 2
…3. Identification of the Predicate Device * Dade Behring Stratus® CS DDMR TestPak and DilPak K022976/K051597 * Dade Behring Stratus® CS DDMR CalPak K022977 FDA Guidance Document(s):…
Dade Behring, Inc.20051
*K022977Stratus CS D-Dimer Calpak Calibrator, Model Cddmr-C
Summary page 2
…* Dade Behring Stratus® CS DDMR TestPak and DilPak K022976/K051597 * Dade Behring Stratus® CS DDMR CalPak K022977 FDA Guidance Document(s): '"Bundling Multiple Devices or Multiple Indications in a Single…
Dade Behring, Inc.2002—

Also cleared against K022976: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081732Innovance D-Dimer and Innovance D-Dimer ControlsSiemens Healthcare Diagnostics20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 7, 2012Class IIIZ-2283-2012Nonconforming Material/ComponentSiemens has confirmed that the Stratus(R) CS Acute Care(TM) D-Dimer CalPak calibrator lots 502058002 and 501318002 may exhibit SLOPE or CALCV error messages (instrument errors), which indicate calibration failures.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 4.9 years on median.
  • 2 of the 5 ancestors have a recall of any class.
  • Of its descendants, K081732 (Siemens Healthcare Diagnostics, 2008) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K022976.

How does its category compare?

All DAP devices

DAP Fibrinogen And Fibrin Split Products, Antigen, Antiserum, Control: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · DAP

1in 2017

2007: 22008: 32009: 12010: 22011: 12012: 12014: 12015: 12016: 22017: 12026 (partial year): 1
2007high 3 (2008) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DAP

127in 2008

2008: 127 days median
2007high 127 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DAP

2in 2025

2015: 12021: 12025: 2
2007high 2 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DAP

2in 2025

2007: 42008: 102009: 132010: 92011: 142012: 102013: 82014: 92015: 382016: 202017: 172018: 52019: 92020: 112021: 72022: 302023: 42024: 12025: 2
2007high 38 (2015) · low 1 (2024)2025

MAUDE reports naming product code DAP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—04
20083127010
20091—013
20102—09
20111—014
20121—010
20130—08
20141—09
20151—138
20162—020
20171—017
20180—05
20190—09
20200—011
20210—17
20220—030
20230—04
20240—01
20250—22
20261—0—

What is it indicated for?

The Stratus® CS Acute CareTM D-dimer (DDMR) method is an in vitro diagnostic test for the quantitative measurement of cross-linked fibrin degradation products (D-dimer) in human citrated or heparinized plasma. The Stratus® CS Acute CareTM DDMR method is intended for use as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)]. This method is for use by trained health care professionals in the…

Read the full statement (156 words)

The Stratus® CS Acute CareTM D-dimer (DDMR) method is an in vitro diagnostic test for the quantitative measurement of cross-linked fibrin degradation products (D-dimer) in human citrated or heparinized plasma. The Stratus® CS Acute CareTM DDMR method is intended for use as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)]. This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings.

The Stratus® CS Acute CareTM D-dimer Calibrator (DDMR CalPak) Catalog. No. CDDMR-C, is an in vitro diagnostic product intended to be used for calibration of the Stratus® CS Acute CareTM D-dimer (DDMR) method.

The Stratus® CS Acute CareTM D-dimer Dilution Pak (DDMR DilPak), Catalog. No. CDDMR-D, is an in vitro diagnostic product intended to be used in conjunction with the Acute CareTM DDMR TestPak, Catalog. No. CDDMR for the measurement of samples with elevated levels of D-Dimer.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K063356: Stratus CS Acute Care D-Dimer Testpak, Dilpak, Calpak.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K063356

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026