510(k) Report

510(k) · Cleared

Cleared

Stryker Intradiscal RF Generator

K063489510(k) TraditionalCleared May 23, 2007187-day reviewClass II21 CFR 882.4400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Synergetics, Inc. · GXD Generator, Lesion, Radiofrequency

Cleared on May 23, 2007 as substantially equivalent to K033981 (Smith & Nephew, Inc., 2004) after a 187-day review. Its family tree reaches back 4 generations to 1990.

Decided
May 23, 2007
Review
187 days
Class
Class II
Product code
GXD
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K896183 (1990). 9 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK896183 KTP/532 Surgical Laser for Percutaneous DiskectomyLaserscope · 1990
  2. 3 generations backK974464 Oratec Spinecath Intradiscal CatheterOratec Interventions, Inc. · 1998
  3. 2 generations backK993854 Ora-50 S Autotemp Electrothermal Spine System and AccessoriesOratec Interventions, Inc. · 1999
  4. 1 generation backK033981 Smith & Nephew Electrothermal 20S Spine Generator, Model 7209975Smith & Nephew, Inc. · 2004
  5. This deviceK063489 Stryker Intradiscal RF GeneratorSynergetics, Inc. · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K896183KTP/532 Surgical Laser for Percutaneous DiskectomyLaserscope1990—
−3K974464Oratec Spinecath Intradiscal CatheterOratec Interventions, Inc.1998—
−2K993854Ora-50 S Autotemp Electrothermal Spine System and AccessoriesOratec Interventions, Inc.1999—
−1K033981Smith & Nephew Electrothermal 20S Spine Generator, Model 7209975Smith & Nephew, Inc.2004—
0 (this device)K063489Stryker Intradiscal RF GeneratorSynergetics, Inc.2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199410
199510
199820
FIG. 1 — Lineage of K063489, 1990–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GXD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GXD device

  1. #1 K251247 GX1 Radiofrequency Generator Kit…score 11.0 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033981Smith & Nephew Electrothermal 20S Spine Generator, Model 7209975
Summary page 1
…ntradiscal RF Generator Common Name: Radiofrequency Lesion Generator Classification Name: Generator, Radiofrequency Lesion (GXD) Equivalent to: K033981 Smith & Nephew ElectroThermal 20S Spine Generator Device Description: The Stryker Intra…
Smith & Nephew, Inc.2004—
†K993967Oratec Spinecath Intradiscal Catheter, Model 920002Oratec Interventions, Inc.1999—

Also cleared against K033981: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 4 companies; a new generation was cleared every 4.2 years on median.
  • None of the 9 ancestors has a recall.
  • 3 other devices also cite its predicate K033981.

How does its category compare?

All GXD devices

GXD Generator, Lesion, Radiofrequency: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · GXD

1in 2025

2007: 42008: 22010: 12011: 22017: 22019: 12020: 22022: 12023: 22024: 22025: 1
2007high 4 (2007) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GXD

185in 2007

2007: 185 days median
2007high 185 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GXD

1in 2023

2010: 12016: 12019: 12023: 1
2007high 1 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GXD

88in 2025

2007: 102008: 102009: 122010: 692011: 892012: 772013: 1192014: 642015: 672016: 512017: 832018: 1562019: 1952020: 1382021: 1542022: 842023: 542024: 632025: 88
2007high 195 (2019) · low 10 (2007)2025

MAUDE reports naming product code GXD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074185010
20082—010
20090—012
20101—169
20112—089
20120—077
20130—0119
20140—064
20150—067
20160—151
20172—083
20180—0156
20191—1195
20202—0138
20210—0154
20221—084
20232—154
20242—063
20251—088
20260—0—

What is it indicated for?

The Stryker Intradiscal RF Generator is intended for the coagulation and decompression of disc material to treat symptomatic patients with annular disruption of contained herniated discs. The Stryker Intradiscal RF Generator will be used with the previously cleared Stryker RF Intradiscal Adapter and catheters such as Smith & Nephew Spinecath TM & Acutherm TM catheters.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K063489: Stryker Intradiscal RF Generator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K063489

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026