510(k) · Cleared
Recalled · 2008Tenderflow Periatric Arterial Cannula
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Terumo Cardiovascular Systems Corp. · DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Cleared on Mar 13, 2007 as substantially equivalent to K930620 (C.R. Bard, Inc., 1993) after a 98-day review. Its family tree reaches back 1 generation to 1993. It has been recalled 24 times.
- Decided
- Mar 13, 2007
- Review
- 98 days
- Class
- Class II
- Product code
- DWF
- Recalls
- 24
- Predicates
- 1
Its longest line of descent runs back 1 generation to K930620 (1993). No later device cites it as a predicate yet.
- Root · 1 generation backK930620 Bard (R) William Harvey (R) Arterial Straigh CannulaC.R. Bard, Inc. · 1993 · Recalled
- This deviceK063618 Tenderflow Periatric Arterial CannulaTerumo Cardiovascular Systems Corp. · 2007
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K930620 | Bard (R) William Harvey (R) Arterial Straigh Cannula | C.R. Bard, Inc. | 1993 | 2 |
| 0 (this device) | K063618 | Tenderflow Periatric Arterial Cannula | Terumo Cardiovascular Systems Corp. | 2007 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DWF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DWF device
- #1 K251258 MC2™ Two-Stage Venous Cannula; MC2X™…score 14.6 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K930620 | Bard (R) William Harvey (R) Arterial Straigh CannulaSummary page 1…fied as Class II devices Predicate Device: The unmodified device is identified as Terumo's L Series 1863 Arterial Cannula. This device was originally cleared by FDA with K930620, dated May 8, 1993. This device is legally marketed and has not been the subject of Premarket Notification 510(k) clearance. The… | 1993 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 16, 2008 | Not classified | Z-0637-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0638-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0639-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0640-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0630-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0633-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0634-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0636-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0641-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0631-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0632-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| Oct 16, 2008 | Not classified | Z-0635-2009 | Device Design | The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it. |
| May 22, 2007 | Not classified | Z-0877-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0881-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0879-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0880-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0883-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0885-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0874-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0876-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0878-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0884-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0875-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
| May 22, 2007 | Not classified | Z-0882-2007 | Other | The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line). |
The 5 latest of 24 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Its one ancestor has a recall.
How does its category compare?
All DWF devicesDWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · DWF
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DWF
89in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DWF
4in 2024
Class I (3 since 2007) Class II/III not classified
Adverse event reports per year · DWF
1,020in 2025
MAUDE reports naming product code DWF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 25 | 121 |
| 2008 | 6 | 95 | 25 | 126 |
| 2009 | 8 | 148 | 21 | 99 |
| 2010 | 3 | 62 | 26 | 197 |
| 2011 | 5 | 85 | 13 | 212 |
| 2012 | 10 | 110 | 37 | 233 |
| 2013 | 9 | 110 | 14 (1) | 238 |
| 2014 | 14 | 110 | 6 | 325 |
| 2015 | 9 | 46 | 3 (1) | 218 |
| 2016 | 3 | 32 | 3 | 144 |
| 2017 | 5 | 127 | 2 (1) | 101 |
| 2018 | 5 | 30 | 10 | 118 |
| 2019 | 3 | 120 | 2 | 179 |
| 2020 | 3 | 30 | 0 | 142 |
| 2021 | 5 | 358 | 3 | 162 |
| 2022 | 2 | — | 0 | 276 |
| 2023 | 2 | — | 0 | 470 |
| 2024 | 3 | 29 | 4 | 401 |
| 2025 | 4 | 89 | 0 | 1,020 |
| 2026 | 8 | 81 | 20 | — |
What is it indicated for?
The TenderFlowTM Pediatric Arterial Cannula is a surgically invasive device
intended for short term use and indicated for cannulation and perfusion of the
ascending aorta during cardiopulmonary bypass surgery.
These devices are indicated for up to 6 hours of use.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K063618 · printed Nov 4, 2026
Cite this page
510(k) Report. “K063618: Tenderflow Periatric Arterial Cannula.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K063618
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026