# K063672: Fibula Plate Prolock with Angular Stability

- Pathway: 510(k) (Traditional)
- Applicant: I.T.S. Implantat-Technologie-Systeme GmbH
- Decided: Feb 2, 2007 after 53 days
- Product code: HRS (Plate, Fixation, Bone)
- Device class: 2
- Regulation: 21 CFR 888.3030
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 4 generations back to K000684 (2000). 23 ancestors in all.
25 later devices descend from it; 3 cite it directly.
0 of 49 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K000684: Small Fragment Dynamic Compression Locking (DCL) System, Synthes (Usa) (2000) Summary page 1.
- K051986: Synthes 4.5 MM LCP Straight Reconstruction Plates, Synthes (Usa) (2005) Summary page 1.
- K013248: Synthes LCP Distal Tibia Plates, Synthes (Usa) (2001) Summary page 1.
- K011335: Synths One-Third Tubular DCL Plate, Synthes (Usa) (2001) Summary page 1.
- K001945: Synthes (Usa) Medial Distal Tibia Plates, Synthes (Usa) (2000) Summary page 1.
- K040593: Periarticular Locking Plates and Screws, 2358, 2359 Series, Zimmer, Inc. (2004) Summary page 1.
- K033639: Acumed Lower Extremity Congruent Plate System, Acumed, LLC (2004) Summary page 1.
- (no number) Synthes Semi-Tubular Plate (Pre-Amendment): named in the summary without a submission number

## Cited in the summary but kept off the family tree
- K010334: Wako Lipid Calibrator, Wako Chemicals USA, Inc. (2001). Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree

## Most influential descendants
- K162674: BioPro Foot Plating Systems (2017), 19 descendants of its own
- K193222: Tyber Medical Anatomical Plating System (2020), 11 descendants of its own
- K203817: Tyber Medical Anatomical Plating System (2021), 10 descendants of its own
- K222465: Tyber Medical Anatomical Plating System (2023), 9 descendants of its own
- K183011: Additive Orthopaedics Patient Specific 3D Locking Lattice Plates (2019), 5 descendants of its own

## Recalls of this device
- Jan 23, 2026 · Class II · Z-1601-2026: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

## Indications for use
The I.T.S. Fibula Plate PROlock with Angular Stability is a titanium implant fracture fixation system for repairing bone fractures located from the middle to the distal third of the fibula. Indications for Use include metaphysial and diaphysial fracture fixation of acute fractures, malunions, and non-unions of the distal fibula. Other indications include corrective osteotomy and open and closed fractures. The system is not intended for spinal use.

Source page: https://510kreport.com/device/K063672
Methodology: https://510kreport.com/methodology
