Fetch Aspiration Catheter

K070363510(k) TraditionalCleared May 11, 200793-day reviewClass II21 CFR 870.5150

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Possis Medical, Inc. · QEZ Aspiration Thrombectomy Catheter

Cleared on May 11, 2007 as substantially equivalent to K042937 (Vascular Solutions, Inc., 2005) and 2 other predicates after a 93-day review. Its family tree reaches back 3 generations to 2003. 1 of the 8 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it.

Decided
May 11, 2007
Review
93 days
Class
Class II
Product code
QEZ
Recalls
None
Predicates
3

Its longest line of descent runs back 3 generations to K032763 (2003). 5 ancestors in all. 2 later devices descend from it, 1 citing it directly. 1 of the 8 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2005: 1 other ancestor2008: 1 descendant cleared2011: 1 descendant cleared, 1 with a Class I recall1K032763 · Vascular Solutions Pronto Extraction Catheter · Vascular Solutions, Inc. · 2003 · recalledK042937 · Modification to Vascular Solutions Pronto Extraction Catheter · Vascular Solutions, Inc. · 2005K062172 · Fetch Aspiration Catheter · Possis Medical, Inc. · 2006200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant, 1 with a Class I recall1 K062172 · Fetch Aspiration Catheter · Possis Medical, Inc. · 2006K042937 · Modification to Vascular Solutions Pronto Extraction Catheter · Vascular Solutions, Inc. · 2005K032763 · Vascular Solutions Pronto Extraction Catheter · Vascular Solutions, Inc. · 2003 · recalled0
  1. Root · 3 generations backK032763 Vascular Solutions Pronto Extraction CatheterVascular Solutions, Inc. · 2003 · Recalled
  2. 2 generations backK042937 Modification to Vascular Solutions Pronto Extraction CatheterVascular Solutions, Inc. · 2005
  3. 1 generation backK062172 Fetch Aspiration CatheterPossis Medical, Inc. · 2006
  4. This deviceK070363 Fetch Aspiration CatheterPossis Medical, Inc. · 2007
  5. Descendant · cites it directly · 1 of its ownK081989 Modification to Fetch Aspiration CatheterPossis Medical, Inc. · 2008
  6. Descendant · 2 generations on · 0 of its ownK101354 Fetch 2 Aspiration Catheter Model 109400-001Medrad Interventional/Possis · 2011 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K081989Modification to Fetch Aspiration CatheterPossis Medical, Inc. · 2008 · cites it directly1descendant
  • K101354Fetch 2 Aspiration Catheter Model 109400-001Medrad Interventional/Possis · 2011 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K032763Vascular Solutions Pronto Extraction CatheterVascular Solutions, Inc.20031
−2K042937Modification to Vascular Solutions Pronto Extraction CatheterVascular Solutions, Inc.2005—
−1K062172Fetch Aspiration CatheterPossis Medical, Inc.2006—
0 (this device)K070363Fetch Aspiration CatheterPossis Medical, Inc.2007—
+1K081989Modification to Fetch Aspiration CatheterPossis Medical, Inc.2008—
+2K101354Fetch 2 Aspiration Catheter Model 109400-001Medrad Interventional/Possis2011Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200510
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201111
FIG. 1 — Lineage of K070363, 2003–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QEZ device

  1. #1 K253925 Aventus Thrombectomy Systemscore 17.9 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042937Modification to Vascular Solutions Pronto Extraction Catheter
Summary page 5
…Possis Medical, Inc. Possis Medical, Inc. Performance Standards: None have been developed for this device, per Section 514 Predicate Devices: Pronto Extraction Catheter (K042937), Pronto V3 Extraction Catheter (K063371) and FETCH Aspiration Catheter (K062172) Device Description The FETCH Aspiration Cath…
Vascular Solutions, Inc.2005—
*K063371Pronto V3 Extraction Catheter, Model 5003
Summary page 5
…Performance Standards: None have been developed for this device, per Section 514 Predicate Devices: Pronto Extraction Catheter (K042937), Pronto V3 Extraction Catheter (K063371) and FETCH Aspiration Catheter (K062172) Device Description The FETCH Aspiration Catheter is a rapid exchange, low-profile tip,…
Vascular Solutions, Inc.2006—
*K062172Fetch Aspiration Catheter
Summary page 5
…veloped for this device, per Section 514 Predicate Devices: Pronto Extraction Catheter (K042937), Pronto V3 Extraction Catheter (K063371) and FETCH Aspiration Catheter (K062172) Device Description The FETCH Aspiration Catheter is a rapid exchange, low-profile tip, dual lumen catheter that uses a 0.014”…
Possis Medical, Inc.2006—

Also cleared against K042937: 3 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081989Modification to Fetch Aspiration CatheterPossis Medical, Inc.20081

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K101354Fetch 2 Aspiration Catheter Model 109400-001Medrad Interventional/Possis · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 4 years and passes through 2 companies; a new generation was cleared every 1.1 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • Of its descendants, K081989 (Possis Medical, Inc., 2008) has gone furthest, with 1 descendant of its own.
  • 3 other devices also cite its predicate K042937.

How does its category compare?

All QEZ devices

QEZ Aspiration Thrombectomy Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · QEZ

5in 2025

2007: 112008: 62009: 42010: 32011: 72012: 52013: 42014: 12016: 12017: 32018: 12019: 32020: 32021: 32022: 22023: 12024: 12025: 52026 (partial year): 6
2007high 11 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QEZ

108in 2025

2007: 79 days median2008: 72 days median2009: 83 days median2010: 33 days median2011: 118 days median2012: 90 days median2013: 144 days median2017: 112 days median2019: 94 days median2020: 120 days median2021: 53 days median2025: 108 days median2026 (partial year): 98 days median
2007high 144 (2013) · low 33 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QEZ

No recall names QEZ since 2007.

Adverse event reports per year · QEZ

21in 2025

2021: 22022: 182023: 142024: 172025: 21
2007high 21 (2025) · low 0 (2007)2025

MAUDE reports naming product code QEZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007117900
200867200
200948300
201033300
2011711800
201259000
2013414400
20141—00
20150—00
20161—00
2017311200
20181—00
201939400
2020312000
202135302
20222—018
20231—014
20241—017
20255108021
20266980—

What is it indicated for?

The FETCH™ Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K070363: Fetch Aspiration Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K070363

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026