Incircle BI-Polar RF Ablation System

K070711510(k) TraditionalCleared Apr 24, 200741-day reviewClass II21 CFR 878.4400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Rfa Medical, Inc. · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Apr 24, 2007 as substantially equivalent to K040785 (Boston Scientific Corp, 2004) and 3 other predicates after a 41-day review. Its family tree reaches back 1 generation to 2001. 3 later devices descend from it.

Decided
Apr 24, 2007
Review
41 days
Class
Class II
Product code
GEI
Recalls
None
Predicates
4

Its longest line of descent runs back 1 generation to K010060 (2001). 4 ancestors in all. 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2005: 1 other ancestor2004: 1 other ancestor2008: 1 descendant cleared2007: 1 descendant cleared2009: 1 descendant clearedK010060 · Rita Starburst Xli Electrosurgical Device · Rita Medical Systems · 2001 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 2 descendants2 generations forward: 1 descendant K010060 · Rita Starburst Xli Electrosurgical Device · Rita Medical Systems · 2001 · recalled0
  1. Root · 1 generation backK010060 Rita Starburst Xli Electrosurgical DeviceRita Medical Systems · 2001 · Recalled
  2. This deviceK070711 Incircle BI-Polar RF Ablation SystemRfa Medical, Inc. · 2007
  3. Descendant · cites it directly · 2 of its ownK070443 Intracept BI-Polar RF Probe, Model 04814Relievant Medsystems, Inc. · 2007
  4. Descendant · 2 generations on · 0 of its ownK083856 Intracept Flexible BI-Polar RF Probe and Curved Instrument Set…Relievant Medsystems, Inc. · 2009
  5. Descendant · cites it directly · 0 of its ownK082954 Encage Bipolar RF ProbeTrod Medical SA · 2008

Most influential descendants

Showing 3 of 3 · See all
  • K070443Intracept BI-Polar RF Probe, Model 04814Relievant Medsystems, Inc. · 2007 · cites it directly2descendants
  • K083856Intracept Flexible BI-Polar RF Probe and Curved Instrument Set, Models: FG0018, FG0026Relievant Medsystems, Inc. · 2009 · 2 generations on0descendants
  • K082954Encage Bipolar RF ProbeTrod Medical SA · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K010060Rita Starburst Xli Electrosurgical DeviceRita Medical Systems20013
0 (this device)K070711Incircle BI-Polar RF Ablation SystemRfa Medical, Inc.2007—
+1K070443Intracept BI-Polar RF Probe, Model 04814Relievant Medsystems, Inc.2007—
+2K083856Intracept Flexible BI-Polar RF Probe and Curved Instrument Set, Models…Relievant Medsystems, Inc.2009—
+1K082954Encage Bipolar RF ProbeTrod Medical SA2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200410
200510
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200810
200910
FIG. 1 — Lineage of K070711, 2001–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040785Concerto Bipolar Needle Electrode
Summary page 2
…The RFA Medical InCircle' TM Bi-Polar RF Ablation System is substantially equivalent to the Boston Scientific ConcertoTM Bipolar Needle Electrode (K040785, K050361, and K060419) and Rita Medical Systems StarBurstTM Xli Electrosurgical Device (K010060). The InCircle has a simi…
Boston Scientific Corp2004—
*K050361Modification to Concerto Bipolar Needle Electrode
Summary page 2
…e RFA Medical InCircle' TM Bi-Polar RF Ablation System is substantially equivalent to the Boston Scientific ConcertoTM Bipolar Needle Electrode (K040785, K050361, and K060419) and Rita Medical Systems StarBurstTM Xli Electrosurgical Device (K010060). The InCircle has a similar inten…
Boston Scientific Corp2005—
*K060419Concerto Bipolar Needle Electrode
Summary page 2
…InCircle' TM Bi-Polar RF Ablation System is substantially equivalent to the Boston Scientific ConcertoTM Bipolar Needle Electrode (K040785, K050361, and K060419) and Rita Medical Systems StarBurstTM Xli Electrosurgical Device (K010060). The InCircle has a similar intended u…
Boston Scientific Corp2006—
*K010060Rita Starburst Xli Electrosurgical Device
Summary page 2
…he Boston Scientific ConcertoTM Bipolar Needle Electrode (K040785, K050361, and K060419) and Rita Medical Systems StarBurstTM Xli Electrosurgical Device (K010060). The InCircle has a similar intended use, materials of construction, and principles of operation as the predicate…
Rita Medical Systems20013

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K070443Intracept BI-Polar RF Probe, Model 04814Relievant Medsystems, Inc.20072
K082954Encage Bipolar RF ProbeTrod Medical SA20080

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K070443 (Relievant Medsystems, Inc., 2007) has gone furthest, with 2 descendants of its own.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The RFA Medical InCircleTM Bi-Polar RF Ablation System is intended to be used in conjunction with radiofrequency (RF) generators for the thermal coagulation of soft tissues.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K070711: Incircle BI-Polar RF Ablation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K070711

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026