510(k) Report

510(k) · Cleared

Cleared

X3C 1600 Plus Digital Radiographic System

K071402510(k) SpecialCleared Jun 8, 200718-day reviewClass II21 CFR 892.1680

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Imaging Dynamics Company, Ltd. · KPR System, X-Ray, Stationary

Cleared on Jun 8, 2007 as substantially equivalent to K062586 (Imaging Dynamics Company, Ltd., 2006) after an 18-day review. Its family tree reaches back 1 generation to 2006. 1 later device descends from it.

Decided
Jun 8, 2007
Review
18 days
Class
Class II
Product code
KPR
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K062586 (2006). 1 later device descends from it, 1 citing it directly.

Descendants, by generation 1 generation forward: 1 descendant K062586 · Xplorer 1600 Plus · Imaging Dynamics Company, Ltd. · 20060
  1. Root · 1 generation backK062586 Xplorer 1600 PlusImaging Dynamics Company, Ltd. · 2006
  2. This deviceK071402 X3C 1600 Plus Digital Radiographic SystemImaging Dynamics Company, Ltd. · 2007
  3. Descendant · cites it directly · 0 of its ownK111098 Innovaxion FP, Innovaxion FP 1600 Plus, MagellanImaging Dynamics Company, Ltd. · 2011

Most influential descendants

Showing 1 of 1 · See all
  • K111098Innovaxion FP, Innovaxion FP 1600 Plus, MagellanImaging Dynamics Company, Ltd. · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K062586Xplorer 1600 PlusImaging Dynamics Company, Ltd.2006—
0 (this device)K071402X3C 1600 Plus Digital Radiographic SystemImaging Dynamics Company, Ltd.2007—
+1K111098Innovaxion FP, Innovaxion FP 1600 Plus, MagellanImaging Dynamics Company, Ltd.2011—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
FIG. 1 — Lineage of K071402, 2006–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062586Xplorer 1600 Plus
Summary page 1
…1600 Plus Digital Radiographic System, Device Classification: Class Il, 892.1680 (KPR), 892.1630 (MQB), Predicate Device: Xplorer 1600 Plus Digital Radiographic System (K062586) Device Description: The X3C 1600 Plus is a modification to Xplorer 1600 Plus where the Xplorer 1000 digital radiographic detect…
Imaging Dynamics Company, Ltd.2006—
†K992955Xplorer 1000Imaging Dynamics Corp.2000—
†K070079X3C Digital Radiographic DetectorImaging Dynamics Company, Ltd.2007—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111098Innovaxion FP, Innovaxion FP 1600 Plus, MagellanImaging Dynamics Company, Ltd.20110

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K111098 (Imaging Dynamics Company, Ltd., 2011) has gone furthest, with 0 descendants of its own.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The X3C 1600 PLUS is intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. The X3C 1600 Plus (510k submission device) is not intended for mammography.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K071402: X3C 1600 Plus Digital Radiographic System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K071402

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026