Synerad Omni, Models AN6255 and AN6265

K071430510(k) SpecialCleared Jun 21, 200729-day reviewClass II21 CFR 892.1680

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Analogic Corp. · KPR System, X-Ray, Stationary

Cleared on Jun 21, 2007 as substantially equivalent to K043025 (Analogic Corp., 2004) after a 29-day review. Its family tree reaches back 6 generations to 1990.

Decided
Jun 21, 2007
Review
29 days
Class
Class II
Product code
KPR
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K900935 (1990). 10 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 6 generations backK900935 DU Pont Linx Laser Imager W/the Linx NetworkingE.I. Dupont DE Nemours & Co., Inc. · 1990
  2. 5 generations backK964250 Linx Networking SystemSterling Diagnostic Imaging, Inc. · 1996
  3. 4 generations backK980970 Iirad Directray Operator ConsoleSterling Diagnostic Imaging, Inc. · 1998
  4. 3 generations backK033345 AN5150 and AN7150 Digital Radiology SystemsAnalogic Corp. · 2003
  5. 2 generations backK040995 Synerad Multi Digital X-Ray System, Model AN6150Analogic Corp. · 2004 · Recalled
  6. 1 generation backK043025 Synerad Omni, Digital X-Ray System, Model AN6250Analogic Corp. · 2004
  7. This deviceK071430 Synerad Omni, Models AN6255 and AN6265Analogic Corp. · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K900935DU Pont Linx Laser Imager W/the Linx NetworkingE.I. Dupont DE Nemours & Co., Inc.1990—
−5K964250Linx Networking SystemSterling Diagnostic Imaging, Inc.1996—
−4K980970Iirad Directray Operator ConsoleSterling Diagnostic Imaging, Inc.1998—
−3K033345AN5150 and AN7150 Digital Radiology SystemsAnalogic Corp.2003—
−2K040995Synerad Multi Digital X-Ray System, Model AN6150Analogic Corp.20041
−1K043025Synerad Omni, Digital X-Ray System, Model AN6250Analogic Corp.2004—
0 (this device)K071430Synerad Omni, Models AN6255 and AN6265Analogic Corp.2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710
200020
FIG. 1 — Lineage of K071430, 1990–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K043025Synerad Omni, Digital X-Ray System, Model AN6250
Summary page 1
…ces to which equivalence is being claimed are: The Analogic AN6250 Digital Radiology System. The predicate system was cleared under Premarket Notification K043025. Analogic Corporation May 21, 2007 AN6255 and AN6265 Digital R…
Analogic Corp.2004—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 2.6 years on median.
  • 2 of the 10 ancestors have a recall of any class.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The AN6255 and AN6265 are digital X-ray general radiography systems intended for use by qualified/trained doctor or technician and are designed to perform radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities and other body parts excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or lying in the prone or supine position. Images from these systems are…

Read the full statement (107 words)

The AN6255 and AN6265 are digital X-ray general radiography systems intended for use by qualified/trained doctor or technician and are designed to perform radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities and other body parts excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or lying in the prone or supine position.

Images from these systems are available for preview by the doctor/X-ray technician on the operator's workstation within seconds of the x-ray exposure. Digital (DICOM) images can be stored on electronic media, or exported to a (DICOMIPACS) network, clinical review station or to a film printer.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071430: Synerad Omni, Models AN6255 and AN6265.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K071430

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026