510(k) Report

510(k) · Cleared

Recalled · 2012

Varisource IX HDR Afterloader

K071467510(k) TraditionalCleared Jul 16, 200748-day reviewClass II21 CFR 892.5700Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Varian Medical Systems, Inc. · JAQ System, Applicator, Radionuclide, Remote-Controlled

Cleared on Jul 16, 2007 as substantially equivalent to K061582 (Varian Medical Systems, Inc., 2006) after a 48-day review. Its family tree reaches back 2 generations to 1995. 1 later device descends from it. It has been recalled 3 times.

Decided
Jul 16, 2007
Review
48 days
Class
Class II
Product code
JAQ
Recalls
3
Predicates
1

Its longest line of descent runs back 2 generations to K945383 (1995). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2012: 1 descendant clearedK945383 · High Dose Rate Afterloader · Varian Assoc., Inc. · 1995K061582 · Varisource 200 HDR Afterloader · Varian Medical Systems, Inc. · 20061995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant K061582 · Varisource 200 HDR Afterloader · Varian Medical Systems, Inc. · 2006K945383 · High Dose Rate Afterloader · Varian Assoc., Inc. · 19950
  1. Root · 2 generations backK945383 High Dose Rate AfterloaderVarian Assoc., Inc. · 1995
  2. 1 generation backK061582 Varisource 200 HDR AfterloaderVarian Medical Systems, Inc. · 2006
  3. This deviceK071467 Varisource IX HDR AfterloaderVarian Medical Systems, Inc. · 2007
  4. Descendant · cites it directly · 0 of its ownK122236 Varisourse IX Brahytherapy Afterloader Varisource IX (T)…Varian Medical Systems, Inc. · 2012 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K122236Varisourse IX Brahytherapy Afterloader Varisource IX (T) Brachytherapy AfterloaderVarian Medical Systems, Inc. · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K945383High Dose Rate AfterloaderVarian Assoc., Inc.1995—
−1K061582Varisource 200 HDR AfterloaderVarian Medical Systems, Inc.2006—
0 (this device)K071467Varisource IX HDR AfterloaderVarian Medical Systems, Inc.2007—
+1K122236Varisourse IX Brahytherapy Afterloader Varisource IX (T) Brachytherapy…Varian Medical Systems, Inc.20121

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
FIG. 1 — Lineage of K071467, 1995–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAQ device

  1. #1 K241853 Titanium Fletcher-Style Applicator Set Defined…score 14.9 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061582Varisource 200 HDR Afterloader
Summary page 1
…R §892.5700 Class 11 Product Code: JAQ 3. Predicate Device to claim substantial equivalence: VariSource K061582 4. Description of the Device: The VaniSource iX Series High Dose Rate Afterloader system is a computer controlled rem…
Varian Medical Systems, Inc.2006—

Also cleared against K061582: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122236Varisourse IX Brahytherapy Afterloader Varisource IX (T) Brachytherapy AfterloaderVarian Medical Systems, Inc.20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 5, 2012Not classifiedZ-1200-2012Device DesignComponents of medical device used in brachytherapy may be defective and cause a shifted dose distribution to patients.
Jan 5, 2012Not classifiedZ-1199-2012Device DesignComponents of medical device used in brachytherapy may be defective and cause a shifted dose distribution to patients.
Mar 1, 2010Not classifiedZ-0325-2012Software designVariSource iX console software may cause incorrect dwell times to be delivered to patient and may result in a mistreatment.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 2 companies; a new generation was cleared every 11.4 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K122236 (Varian Medical Systems, Inc., 2012) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K061582.

How does its category compare?

All JAQ devices

JAQ System, Applicator, Radionuclide, Remote-Controlled: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · JAQ

6in 2025

2007: 112008: 102009: 102010: 22011: 22012: 72013: 62014: 72015: 112016: 62017: 72018: 72019: 22020: 42022: 12023: 22024: 102025: 62026 (partial year): 2
2007high 11 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAQ

91in 2025

2007: 48 days median2008: 34 days median2009: 27 days median2012: 89 days median2013: 100 days median2014: 116 days median2015: 183 days median2016: 154 days median2017: 229 days median2018: 252 days median2020: 62 days median2024: 126 days median2025: 91 days median
2007high 252 (2018) · low 27 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAQ

1in 2019

2008: 52009: 12010: 22011: 132012: 52013: 12019: 12026 (partial year): 3
2007high 13 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAQ

3in 2025

2007: 132008: 322009: 352010: 362011: 422012: 342013: 432014: 182015: 72016: 32017: 52018: 22019: 82020: 82021: 52022: 52023: 32024: 12025: 3
2007high 43 (2013) · low 1 (2024)2025

MAUDE reports naming product code JAQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071148013
20081034532
20091027135
20102—236
20112—1342
2012789534
20136100143
20147116018
20151118307
2016615403
2017722905
2018725202
20192—18
202046208
20210—05
20221—05
20232—03
20241012601
202569103
20262—3—

What is it indicated for?

The VariSource iX is indicated, in the treatment of both benign and malignant disease, for both curative and palliative intent, in the delivery of remote-controlled high dose-rate Brachytherapy for conditions anywhere in the body when brachytherapy treatment is indicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071467: Varisource IX HDR Afterloader.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K071467

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026