510(k) Report

510(k) · Cleared

Recalled · 2009

Biomet Microfixation Lactosorb Pectus Stabilizer

K071577510(k) TraditionalCleared Sep 6, 200790-day reviewClass II21 CFR 888.3030Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biomet Microfixation, Inc. · HRS Plate, Fixation, Bone

Cleared on Sep 6, 2007 as substantially equivalent to K981789 (Walter Lorenz Surgical, Inc., 1998) and 2 other predicates after a 90-day review. Its family tree reaches back 3 generations to 1997. It has been recalled once.

Decided
Sep 6, 2007
Review
90 days
Class
Class II
Product code
HRS
Recalls
1
Predicates
3

Its longest line of descent runs back 3 generations to K972420 (1997). No later device cites it as a predicate yet.

  1. Root · 3 generations backK972420 Lorenz Pectus Support BarWalter Lorenz Surgical, Inc. · 1997
  2. 2 generations backK981789 Lorenz Pectus Support Bar StabilizerWalter Lorenz Surgical, Inc. · 1998
  3. 1 generation backK061384 Lorenz Pectus Support Bar, Titanium; Lorenz Pectus Support Bar…Walter Lorenz Surgical, Inc. · 2006
  4. This deviceK071577 Biomet Microfixation Lactosorb Pectus StabilizerBiomet Microfixation, Inc. · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K972420Lorenz Pectus Support BarWalter Lorenz Surgical, Inc.1997—
−2K981789Lorenz Pectus Support Bar StabilizerWalter Lorenz Surgical, Inc.1998—
−1K061384Lorenz Pectus Support Bar, Titanium; Lorenz Pectus Support Bar Stabilizer…Walter Lorenz Surgical, Inc.2006—
0 (this device)K071577Biomet Microfixation Lactosorb Pectus StabilizerBiomet Microfixation, Inc.2007—
FIG. 1 — Lineage of K071577, 1997–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HRS device

  1. #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K981789Lorenz Pectus Support Bar Stabilizer
Summary page 1
…quivalence: The Biomet Microfixation LactoSorb®9 Pectus Stabilizer is believed to be substantially equivalent in application and function to the Lorenz Pectus Stabilizer K98 1789, Lorenz Pectus Support Bar K972420 and Lorenz Pectus Support Bar System K061384. DEPARTMENT OF HEALTH & HUM…
Walter Lorenz Surgical, Inc.1998—
*K972420Lorenz Pectus Support Bar
Summary page 1
…ant and does not require removal. Intended Use: The Biomet Microfixation LactoSorb® Pectus Stabilizer is intended to be used with Lorenz Pectus Support Bars cleared via K972420 and K06 1384 for repairing Pectus Excavatumn and other sternal deformities when additional stabilization is necessary. Materials…
Walter Lorenz Surgical, Inc.1997—
*K061384Lorenz Pectus Support Bar, Titanium; Lorenz Pectus Support Bar Stabilizer, Titanium
Summary page 1
…not require removal. Intended Use: The Biomet Microfixation LactoSorb® Pectus Stabilizer is intended to be used with Lorenz Pectus Support Bars cleared via K972420 and K06 1384 for repairing Pectus Excavatumn and other sternal deformities when additional stabilization is necessary. Materials Lac…
Walter Lorenz Surgical, Inc.2006—

Also cleared against K981789: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 20, 2009Not classifiedZ-1323-2009Device DesignThis recall is based on the occurrence of known adverse effects identified in the device instructions for use. Bending, fracture, loosening, rubbing and migration of the devices can occur as a result of excessive activity, trauma or load…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 1.3 years on median.
  • None of the 3 ancestors has a recall.
  • 2 other devices also cite its predicate K981789.

How does its category compare?

All HRS devices

HRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · HRS

54in 2025

2007: 352008: 352009: 472010: 492011: 432012: 592013: 562014: 762015: 562016: 442017: 572018: 402019: 462020: 482021: 402022: 472023: 402024: 542025: 542026 (partial year): 51
2007high 76 (2014) · low 35 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HRS

90in 2025

2007: 63 days median2008: 87 days median2009: 85 days median2010: 100 days median2011: 88 days median2012: 106 days median2013: 89 days median2014: 57 days median2015: 97 days median2016: 109 days median2017: 122 days median2018: 116 days median2019: 89 days median2020: 88 days median2021: 87 days median2022: 144 days median2023: 120 days median2024: 96 days median2025: 90 days median2026 (partial year): 82 days median
2007high 144 (2022) · low 57 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HRS

1in 2025

2007: 22008: 112009: 32010: 162011: 52012: 42013: 112015: 72016: 52017: 32018: 22019: 252020: 12021: 92022: 32023: 12025: 12026 (partial year): 3
2007high 25 (2019) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HRS

2,306in 2025

2007: 3392008: 3112009: 4352010: 4912011: 7732012: 1,6632013: 2,6432014: 2,4132015: 3,0952016: 2,2762017: 2,2202018: 2,6632019: 3,3592020: 2,7902021: 2,4422022: 1,7002023: 2,0492024: 1,8492025: 2,306
2007high 3,359 (2019) · low 311 (2008)2025

MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735632339
2008358711311
200947853435
20104910016491
201143885773
20125910641,663
20135689112,643
2014765702,413
2015569773,095
20164410952,276
20175712232,220
20184011622,663
20194689253,359
2020488812,790
2021408792,442
20224714431,700
20234012012,049
2024549601,849
2025549012,306
202651823—

What is it indicated for?

The Biomet Microfixation LactoSorb® Pectus Stabilizer is intended to be used with Lorenz Pectus Support Bars cleared via K972420 and K061384 for repairing Pectus Excavatum and other sternal deformities when additional stabilization is necessary.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071577: Biomet Microfixation Lactosorb Pectus Stabilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K071577

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026