510(k) Report

510(k) · Cleared

Cleared

D101 Kids Infant Hollow Fiber Membrane Oxygenator with Integrated Hardshell Cardiotomy/Venous Reservoir

K072091510(k) AbbreviatedCleared Oct 29, 200791-day reviewClass II21 CFR 870.4350

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sorin Group Italia S.R.L. · DTZ Oxygenator, Cardiopulmonary Bypass

Cleared on Oct 29, 2007 as substantially equivalent to K001021 (Dideco S.P.A., 2000) and 1 other predicate after a 91-day review. Its family tree reaches back 2 generations to 2000. 1 of the 18 devices in its family, counting itself, has a Class I recall. 12 later devices descend from it.

Decided
Oct 29, 2007
Review
91 days
Class
Class II
Product code
DTZ
Recalls
None
Predicates
2

Its longest line of descent runs back 2 generations to K991737 (2000). 5 ancestors in all. 12 later devices descend from it, 5 citing it directly. 1 of the 18 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 2 other ancestors2001: 1 other ancestor2011: 3 descendants cleared2023: 2 descendants cleared2010: 1 descendant cleared2013: 2 descendants cleared2014: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared2021: 1 descendant cleared2026: 1 descendant clearedK991737 · Lilliput Phospholypidic Inert Surface in Oxygenation (PH.I.S.I.O.) Infant/Newborn Hollow Fiber Oxygenator · Dideco S.P.A. · 2000 · recalledK061031 · D 100 L001 PH.I.S.I.O.: Newborn Hollow Fiber Membrane Oxygenator with Integrated Hardshell Cardiotomy/Venous Reservoir W · Sorin Group Italia S.R.L. · 2006200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors 1 generation forward: 5 descendants2 generations forward: 3 descendants, 1 with a Class I recall13 generations forward: 2 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant K061031 · D 100 L001 PH.I.S.I.O.: Newborn Hollow Fiber Membrane Oxygenator with Integrated Hardshell Cardiotomy/Venous Reservoir W · Sorin Group Italia S.R.L. · 2006K991737 · Lilliput Phospholypidic Inert Surface in Oxygenation (PH.I.S.I.O.) Infant/Newborn Hollow Fiber Oxygenator · Dideco S.P.A. · 2000 · recalled0+5
  1. Root · 2 generations backK991737 Lilliput Phospholypidic Inert Surface in Oxygenation (PH.I.S.I.O.)…Dideco S.P.A. · 2000 · Recalled
  2. 1 generation backK061031 D 100 L001 PH.I.S.I.O.: Newborn Hollow Fiber Membrane Oxygenator…Sorin Group Italia S.R.L. · 2006
  3. This deviceK072091 D101 Kids Infant Hollow Fiber Membrane Oxygenator with Integrated…Sorin Group Italia S.R.L. · 2007
  4. Descendant · cites it directly · 6 of its ownK103191 Quadrox-I Pediatric Microporous Membrane Oxygenator with Integrated…Maquet Cardiopulmonary, AG · 2011 · Recalled
  5. Descendant · cites it directly · 6 of its ownK100278 Quadrox-ID Pediatric Diffusion Membrane Oxygenator with Bioline…Maquet Cardiopulmonary, AG · 2010 · Recalled
  6. Descendant · 2 generations on · 4 of its ownK132829 Quadrox-I, Quadrox-ID, Quadrox-IR OxygenatorsMaquet Cardiopulmonary, AG · 2013 · Recalled
  7. Descendant · 3 generations on · 2 of its ownK150267 QUADROX-i Adult/Small Adult Oxygenators;QUADROX-iD Adult OxygenatorsMaquet Cardiopulmonary, AG · 2015 · Recalled
  8. Descendant · 4 generations on · 1 of its ownK203424 Quantum Perfusion Blood Oxygenator ECC VT200-E1, Quantum Perfusion…Qura S.R.L · 2021

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2010: 12011: 32013: 22014: 12015: 12021: 12023: 22026: 1
2010high 3 (2011) · low 0 (2012)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K103191Quadrox-I Pediatric Microporous Membrane Oxygenator with Integrated Arterial Filter with…Maquet Cardiopulmonary, AG · 2011 · cites it directly6descendants
  • K100278Quadrox-ID Pediatric Diffusion Membrane Oxygenator with Bioline Coating, Models BE-Hmod…Maquet Cardiopulmonary, AG · 2010 · cites it directly6descendants
  • K132829Quadrox-I, Quadrox-ID, Quadrox-IR OxygenatorsMaquet Cardiopulmonary, AG · 2013 · 2 generations on4descendants
  • K150267QUADROX-i Adult/Small Adult Oxygenators;QUADROX-iD Adult OxygenatorsMaquet Cardiopulmonary, AG · 2015 · 3 generations on2descendants
  • K203424Quantum Perfusion Blood Oxygenator ECC VT200-E1, Quantum Perfusion Blood Oxygenator ECC…Qura S.R.L · 2021 · 4 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K991737Lilliput Phospholypidic Inert Surface in Oxygenation (PH.I.S.I.O.)…Dideco S.P.A.20001
−1K061031D 100 L001 PH.I.S.I.O.: Newborn Hollow Fiber Membrane Oxygenator with…Sorin Group Italia S.R.L.2006—
0 (this device)K072091D101 Kids Infant Hollow Fiber Membrane Oxygenator with Integrated Hardshell…Sorin Group Italia S.R.L.2007—
+1K103191Quadrox-I Pediatric Microporous Membrane Oxygenator with Integrated Arterial…Maquet Cardiopulmonary, AG20114
+1K100278Quadrox-ID Pediatric Diffusion Membrane Oxygenator with Bioline Coating…Maquet Cardiopulmonary, AG20104
+2K132829Quadrox-I, Quadrox-ID, Quadrox-IR OxygenatorsMaquet Cardiopulmonary, AG20131
+3K150267QUADROX-i Adult/Small Adult Oxygenators;QUADROX-iD Adult OxygenatorsMaquet Cardiopulmonary, AG20152
+4K203424Quantum Perfusion Blood Oxygenator ECC VT200-E1, Quantum Perfusion Blood…Qura S.R.L2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201130
201320
201411
201510
202110
202320
202610
FIG. 1 — Lineage of K072091, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTZ device

  1. #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001021Lilliput Phospholypidic Inert Surface in Oxygenation (PH.I.S.I.O.) Infant Hollow Fibre…
Summary page 1
…Surface In Oxygenation (Ph.I.S.hO.) Infant Hollow Fiber Oxygenator (K001021) D100 L001 PhJ.S.I.O. Newborn Hollow fiber…
Dideco S.P.A.2000—
*K061031D 100 L001 PH.I.S.I.O.: Newborn Hollow Fiber Membrane Oxygenator with Integrated…
Summary page 1
…ed Hardshell CardiotomyNenous Reservoir with phosphorylcholine coating (K061031) DEVICE DESCRIPTION: The D101 KIDS Hollow Fiber Membrane Oxygenator With Integrated Hardshell CardiotomyNenous Reservoir with…
Sorin Group Italia S.R.L.2006—

Also cleared against K001021: 1 other device →

What descends from it? (12)

See all 12

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103191Quadrox-I Pediatric Microporous Membrane Oxygenator with Integrated Arterial Filter with…Maquet Cardiopulmonary, AG20116
K100278Quadrox-ID Pediatric Diffusion Membrane Oxygenator with Bioline Coating, Models BE-Hmod…Maquet Cardiopulmonary, AG20106
K102919Neonatal Venous Hardshell Cardiotomy Reservoir, Pediatric Venous Hardshell Carsiotomy…Maquet Cardiopulmonary, AG20111
K231652D100 Kids; D101 KidsSorin Group Italia S.R.L.20230
K231773Quantum Perfusion Blood Oxygenator with Integrated AF and Sensor VT75-C1U, Quantum…Qura S.R.L20230

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K141432Neonatal Venous Hardshell Cardiotomy Reservior, Pediatric Venous Hardshell Cardiotomy Reservior, Neonatal…Maquet Cardiopulmonary, AG · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 6.3 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • Of its descendants, K103191 (Maquet Cardiopulmonary, AG, 2011) has gone furthest, with 6 descendants of its own.
  • 1 other device also cites its predicate K001021.

How does its category compare?

All DTZ devices

DTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · DTZ

2in 2025

2007: 42008: 32009: 92010: 22011: 42012: 92013: 92014: 52015: 92016: 32017: 22018: 32019: 52020: 52021: 52022: 22023: 52024: 32025: 22026 (partial year): 3
2007high 9 (2009) · low 2 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTZ

124in 2024

2007: 55 days median2008: 88 days median2009: 210 days median2011: 160 days median2012: 90 days median2013: 28 days median2014: 28 days median2015: 134 days median2016: 62 days median2018: 72 days median2019: 90 days median2020: 93 days median2021: 336 days median2023: 145 days median2024: 124 days median2026 (partial year): 137 days median
2007high 336 (2021) · low 28 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTZ

1in 2025

2007: 72008: 52010: 12011: 132012: 142014: 22016: 12018: 2, 2 Class I2019: 12020: 32021: 1, 1 Class I2022: 62025: 1
2007high 14 (2012) · low 0 (2009)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DTZ

976in 2025

2007: 1342008: 1052009: 522010: 1192011: 1902012: 2152013: 3462014: 7642015: 6582016: 4432017: 3882018: 3202019: 4892020: 4422021: 5952022: 5592023: 7382024: 5352025: 976
2007high 976 (2025) · low 52 (2009)2025

MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074557134
20083885105
20099210052
20102—1119
2011416013190
201299014215
20139280346
20145282764
201591340658
20163621443
20172—0388
20183722 (2)320
20195901489
20205933442
202153361 (1)595
20222—6559
202351450738
202431240535
20252—1976
202631370—

What is it indicated for?

The D101 KIDS Infant Hollow Fiber Membrane Oxygenator is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 2.5 liters/minute. It provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery.…

Read the full statement (267 words)

The D101 KIDS Infant Hollow Fiber Membrane Oxygenator is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 2.5 liters/minute. It provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacuum assisted) during normal operation, to always assure the proper oxygenation capability of the device. The D101 KIDS should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

The D101 KIDS Infant Hollow Fiber Membrane Oxygenator is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 2.5 liters/minute. It provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The D101 KIDS should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

D121 KIDS HVR has been specifically designed for cardiovascular procedures requiring cardiopulmonary by-pass. It collects venous blood and it defoams, filters and stores the blood from the operating field through thoracic, intracardiac and general suction. The D121 KIDS HVR should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072091: D101 Kids Infant Hollow Fiber Membrane Oxygenator with Integrated Hardshell Cardiotomy/Venous Reservoir.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072091

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026