510(k) Report

510(k) · Cleared

Recalled · 2010

Dorado Pta Balloon Dilatation Catheter

K072283510(k) TraditionalCleared Sep 19, 200734-day reviewClass II21 CFR 870.1250Recalls 15

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bard Peripheral Vascular, Inc. · DQY Catheter, Percutaneous

Cleared on Sep 19, 2007 as substantially equivalent to K014212 (C.R. Bard, Inc., 2002) and 3 other predicates after a 34-day review. Its family tree reaches back 4 generations to 2000. 96 later devices descend from it. It has been recalled 15 times.

Decided
Sep 19, 2007
Review
34 days
Class
Class II
Product code
DQY
Recalls
15
Predicates
4

Its longest line of descent runs back 4 generations to K001134 (2000). 8 ancestors in all. 96 later devices descend from it, 7 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2002: 1 other ancestor2001: 1 other ancestor2005: 1 other ancestor2008: 2 descendants cleared2025: 6 descendants cleared2010: 2 descendants cleared2011: 8 descendants cleared2012: 7 descendants cleared2015: 7 descendants cleared2009: 1 descendant cleared2019: 4 descendants cleared2017: 5 descendants cleared2014: 6 descendants cleared2016: 9 descendants cleared2022: 7 descendants cleared2020: 5 descendants cleared2023: 2 descendants cleared2024: 8 descendants cleared2013: 6 descendants cleared2026: 6 descendants cleared2018: 2 descendants cleared2021: 3 descendants clearedK001134 · Gazelle Balloon Dilatation Catheter · Boston Scientific Corp · 2000K021735 · Boston Scientific Ultra-Soft SV Balloon Dilatation Catheter · Boston Scientific Scimed, Inc. · 2002 · recalledK050389 · Ultra-Soft SV Balloon Dilatation Catheter · Boston Scientific Corp · 2005K053116 · Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-Xxxxxx · Boston Scientific Corp · 2005200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 1 other ancestor 1 generation forward: 7 descendants2 generations forward: 26 descendants3 generations forward: 29 descendants4 generations forward: 17 descendants5 generations forward: 10 descendants6 generations forward: 6 descendants7 generations forward: 1 descendant K053116 · Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-Xxxxxx · Boston Scientific Corp · 2005K050389 · Ultra-Soft SV Balloon Dilatation Catheter · Boston Scientific Corp · 2005K021735 · Boston Scientific Ultra-Soft SV Balloon Dilatation Catheter · Boston Scientific Scimed, Inc. · 2002 · recalledK001134 · Gazelle Balloon Dilatation Catheter · Boston Scientific Corp · 20000+5
  1. Root · 4 generations backK001134 Gazelle Balloon Dilatation CatheterBoston Scientific Corp · 2000
  2. 3 generations backK021735 Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc. · 2002 · Recalled
  3. 2 generations backK050389 Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp · 2005
  4. 1 generation backK053116 Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-XxxxxxBoston Scientific Corp · 2005
  5. This deviceK072283 Dorado Pta Balloon Dilatation CatheterBard Peripheral Vascular, Inc. · 2007
  6. Descendant · cites it directly · 44 of its ownK083657 Conquest Pta Balloon Dilatation CatheterC.R. Bard, Inc. · 2008 · Recalled
  7. Descendant · cites it directly · 36 of its ownK103459 Vascutrak Pta Dilatation CatheterC.R. Bard, Inc. · 2010
  8. Descendant · cites it directly · 35 of its ownK082579 Evercross .035 Otw Pta Dilatation CatheterEv3, Inc. · 2008 · Recalled
  9. Descendant · 2 generations on · 29 of its ownK112182 Angiosculpt Pta Scoring Balloon CatheterAngioscore, Inc. · 2011
  10. Descendant · 3 generations on · 28 of its ownK122685 Angiosculpt Pta Scoring Balloon CatheterAngioscore, Inc. · 2013

Showing 5 of 96 descendants. See all descendants

Descendants cleared per year

2008: 22009: 12010: 22011: 82012: 72013: 62014: 62015: 72016: 92017: 52018: 22019: 42020: 52021: 32022: 72023: 22024: 82025: 62026: 6
2008high 9 (2016) · low 1 (2009)2026

Most influential descendants

Showing 5 of 96 · See all
  • K083657Conquest Pta Balloon Dilatation CatheterC.R. Bard, Inc. · 2008 · cites it directly44descendants
  • K103459Vascutrak Pta Dilatation CatheterC.R. Bard, Inc. · 2010 · cites it directly36descendants
  • K082579Evercross .035 Otw Pta Dilatation CatheterEv3, Inc. · 2008 · cites it directly35descendants
  • K112182Angiosculpt Pta Scoring Balloon CatheterAngioscore, Inc. · 2011 · 2 generations on29descendants
  • K122685Angiosculpt Pta Scoring Balloon CatheterAngioscore, Inc. · 2013 · 3 generations on28descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K001134Gazelle Balloon Dilatation CatheterBoston Scientific Corp2000—
−3K021735Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc.200223
−2K050389Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp2005—
−1K053116Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-XxxxxxBoston Scientific Corp2005—
0 (this device)K072283Dorado Pta Balloon Dilatation CatheterBard Peripheral Vascular, Inc.2007—
+1K083657Conquest Pta Balloon Dilatation CatheterC.R. Bard, Inc.20082
+1K103459Vascutrak Pta Dilatation CatheterC.R. Bard, Inc.2010—
+1K082579Evercross .035 Otw Pta Dilatation CatheterEv3, Inc.20081
+2K112182Angiosculpt Pta Scoring Balloon CatheterAngioscore, Inc.2011—
+3K122685Angiosculpt Pta Scoring Balloon CatheterAngioscore, Inc.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200210
200510
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200820
200910
201020
201180
201270
201360
201460
201570
201690
201750
201820
201940
202050
202130
202270
202320
202480
202560
202660
FIG. 1 — Lineage of K072283, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQY device

  1. #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K014212Bard Conquest Pta Balloon Dilatation Catheter
Summary page 1
…Classification: Class II Classification Panel: Cardiovascular Predicate Devices: ConquestTM PTA Balloon Dilatation Catheter (K014212, cleared January 17, 2002), manufactured by Bard Peripheral Vascular, Inc. Ultraverse ® Small Vessel PTA Balloon…
C.R. Bard, Inc.2002—
*K012913Bard Utaverse Small Vessel Pta Balloon Dilatation Catheter
Summary page 1
…Catheter (K014212, cleared January 17, 2002), manufactured by Bard Peripheral Vascular, Inc. Ultraverse ® Small Vessel PTA Balloon Dilatation Catheter (K012913, cleared September 28, 2001), manufactured by Bard Peripheral Vascular, Inc.…
C.R. Bard, Inc.2001—
*K053116Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-Xxxxxx
Summary page 2
…e.%ot3 Page 20 * Sterling TM Over-The-Wire PTA Balloon Dilatation Catheter (K053116, cleared December 16, 2005), manufactured by Boston Scientific Corporation. * PolarCath TM Peripheral Dila…
Boston Scientific Corp2005—
*K060572Modification to Polarcath Peripheral Dilatation System
Summary page 2
…atation Catheter (K053116, cleared December 16, 2005), manufactured by Boston Scientific Corporation. * PolarCath TM Peripheral Dilatation System (K060572, cleared March 15, 2006), manufactured by Boston Scientific Corporation. Device Description: The subject device, the Do…
Boston Scientific Corporation2006—

Also cleared against K014212: 3 other devices →

What descends from it? (96)

See all 96

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K083657Conquest Pta Balloon Dilatation CatheterC.R. Bard, Inc.200844
K103459Vascutrak Pta Dilatation CatheterC.R. Bard, Inc.201036
K082579Evercross .035 Otw Pta Dilatation CatheterEv3, Inc.200835
K103751Mustang Balloon Dilatation CathetersBoston Scientific201115
K120660Conquest 40 Pta Dilatation CatheterBard Peripheral Vascular, Inc.20123
K120722Rival Pta Dilatation CatheterBard Peripheral Vascular, Inc.20121
K250219Dorado™ PTA Balloon Dilatation CatheterBard Peripheral Vascular, Inc.20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 4, 2010Not classifiedZ-2136-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2122-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2124-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2131-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2135-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2129-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2123-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2128-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2132-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2133-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2125-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2134-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2126-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2127-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
May 4, 2010Not classifiedZ-2130-2010Nonconforming Material/ComponentThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 1.9 years on median.
  • 1 of the 8 ancestors has a recall of any class.
  • Of its descendants, K083657 (C.R. Bard, Inc., 2008) has gone furthest, with 44 descendants of its own.
  • 3 other devices also cite its predicate K014212.

How does its category compare?

All DQY devices

DQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · DQY

21in 2025

2007: 162008: 312009: 372010: 222011: 362012: 302013: 322014: 192015: 272016: 222017: 262018: 252019: 302020: 202021: 182022: 182023: 182024: 162025: 212026 (partial year): 11
2007high 37 (2009) · low 16 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQY

115in 2025

2007: 85 days median2008: 66 days median2009: 73 days median2010: 74 days median2011: 71 days median2012: 87 days median2013: 92 days median2014: 127 days median2015: 118 days median2016: 117 days median2017: 113 days median2018: 106 days median2019: 111 days median2020: 120 days median2021: 81 days median2022: 74 days median2023: 114 days median2024: 87 days median2025: 115 days median2026 (partial year): 30 days median
2007high 127 (2014) · low 66 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQY

1in 2025

2007: 62008: 122009: 172010: 322011: 62012: 22013: 36, 1 Class I2014: 32015: 32016: 22, 7 Class I2017: 82018: 22019: 12020: 22021: 22023: 172024: 12025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 36 (2013) · low 0 (2022)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · DQY

3,056in 2025

2007: 3132008: 4962009: 8192010: 1,2932011: 1,2112012: 1,1912013: 1,3522014: 1,4972015: 2,3372016: 2,2522017: 2,4892018: 2,8552019: 2,9022020: 2,8892021: 3,0062022: 2,6432023: 3,0492024: 3,1902025: 3,056
2007high 3,190 (2024) · low 313 (2007)2025

MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200716856313
2008316612496
2009377317819
20102274321,293
2011367161,211
2012308721,191
2013329236 (1)1,352
20141912731,497
20152711832,337
20162211722 (7)2,252
20172611382,489
20182510622,855
20193011112,902
20202012022,889
2021188123,006
2022187402,643
202318114173,049
2024168713,190
2025211151 (1)3,056
202611302 (1)—

What is it indicated for?

The DoradoTM PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the renal, iliac, femoral, popliteal, tibial, peroneal, and subclavian arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature. This catheter is not for use in the coronary arteries.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072283: Dorado Pta Balloon Dilatation Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072283

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026