Ksea Clearvision II, Model 40334120

K072410510(k) TraditionalCleared May 9, 2008256-day reviewClass II21 CFR 882.1480

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

KARL STORZ Endoscopy-America, Inc. · GWG Endoscope, Neurological

Cleared on May 9, 2008 as substantially equivalent to K982594 (Xomed, Inc., 1998) and 1 other predicate after a 256-day review. Its family tree reaches back 2 generations to 1990. 4 later devices descend from it.

Decided
May 9, 2008
Review
256 days
Class
Class II
Product code
GWG
Recalls
None
Predicates
2

Its longest line of descent runs back 2 generations to K902683 (1990). 3 ancestors in all. 4 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2018: 1 descendant cleared2026: 2 descendants cleared2020: 1 descendant clearedK902683 · Scopescrubber (TM) · Healthtek Microsurgical, Inc. · 1990K982594 · Endo-Scrub/Endo-Scrub 2 · Xomed, Inc. · 1998 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant K982594 · Endo-Scrub/Endo-Scrub 2 · Xomed, Inc. · 1998 · recalledK902683 · Scopescrubber (TM) · Healthtek Microsurgical, Inc. · 19900
  1. Root · 2 generations backK902683 Scopescrubber (TM)Healthtek Microsurgical, Inc. · 1990
  2. 1 generation backK982594 Endo-Scrub/Endo-Scrub 2Xomed, Inc. · 1998 · Recalled
  3. This deviceK072410 Ksea Clearvision II, Model 40334120KARL STORZ Endoscopy-America, Inc. · 2008
  4. Descendant · cites it directly · 2 of its ownK180735 Endomat SelectKarl Storz SE & CO. KG · 2018
  5. Descendant · 2 generations on · 1 of its ownK201355 Endomat Select, IBS Shaver Software, LicenseKarl Storz SE & CO. KG · 2020
  6. Descendant · 3 generations on · 0 of its ownK261706 Endomat Select (UP220)Karl Storz SE & CO. KG · 2026
  7. Descendant · cites it directly · 0 of its ownK253679 Hydro Irrigation SystemStryker Instruments · 2026

Most influential descendants

Showing 4 of 4 · See all
  • K180735Endomat SelectKarl Storz SE & CO. KG · 2018 · cites it directly2descendants
  • K201355Endomat Select, IBS Shaver Software, LicenseKarl Storz SE & CO. KG · 2020 · 2 generations on1descendant
  • K261706Endomat Select (UP220)Karl Storz SE & CO. KG · 2026 · 3 generations on0descendants
  • K253679Hydro Irrigation SystemStryker Instruments · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K902683Scopescrubber (TM)Healthtek Microsurgical, Inc.1990—
−1K982594Endo-Scrub/Endo-Scrub 2Xomed, Inc.19981
0 (this device)K072410Ksea Clearvision II, Model 40334120KARL STORZ Endoscopy-America, Inc.2008—
+1K180735Endomat SelectKarl Storz SE & CO. KG2018—
+2K201355Endomat Select, IBS Shaver Software, LicenseKarl Storz SE & CO. KG2020—
+3K261706Endomat Select (UP220)Karl Storz SE & CO. KG2026—
+1K253679Hydro Irrigation SystemStryker Instruments2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202010
202620
FIG. 1 — Lineage of K072410, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GWG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWG device

  1. #1 K211202 1688 4K Camera System, L11 LED Light Source…score 22.5 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982594Endo-Scrub/Endo-Scrub 2
Summary page 1
…Classification Name: Neurological endoscope. Regulation Number: 21 CFR 882.1480 Product Code: GWG Predicate Device(s): Xomed Endo-Scrub/Endo-Scrub 2 (K982594) KSEA Clearvision® Lens Irrigation System (K013838) Device Description: The KSEA Clearvision® II is a…
Xomed, Inc.19981
*K013838Ksea Clearvision Lens Irrigation System
Summary page 1
…CFR 882.1480 Product Code: GWG Predicate Device(s): Xomed Endo-Scrub/Endo-Scrub 2 (K982594) KSEA Clearvision® Lens Irrigation System (K013838) Device Description: The KSEA Clearvision® II is a microprocessor-controlled pump device. The device p…
KARL STORZ Endoscopy-America, Inc.2002—

Also cleared against K982594: 2 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180735Endomat SelectKarl Storz SE & CO. KG20182
K253679Hydro Irrigation SystemStryker Instruments20260

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 9.7 years on median.
  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K180735 (Karl Storz SE & CO. KG, 2018) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K982594.

How does its category compare?

All GWG devices

GWG Endoscope, Neurological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · GWG

2in 2025

2008: 22013: 12014: 22016: 22017: 42018: 42019: 32020: 22021: 32023: 52024: 12025: 22026 (partial year): 4
2007high 5 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWG

269in 2023

2017: 126 days median2018: 128 days median2019: 156 days median2021: 106 days median2023: 269 days median2026 (partial year): 217 days median
2007high 269 (2023) · low 106 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWG

2in 2011

2009: 12011: 2
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GWG

33in 2025

2007: 32008: 52010: 12011: 42012: 72013: 32014: 192015: 372016: 252017: 122018: 162019: 102020: 192021: 212022: 122023: 92024: 92025: 33
2007high 37 (2015) · low 0 (2009)2025

MAUDE reports naming product code GWG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—03
20082—05
20090—10
20100—01
20110—24
20120—07
20131—03
20142—019
20150—037
20162—025
20174126012
20184128016
20193156010
20202—019
20213106021
20220—012
2023526909
20241—09
20252—033
202642170—

What is it indicated for?

The KSEA Clearvision II is a lens irrigation system for cleaning the distal lens of the telescope and maintaining clear visualization without removing the scope from the surgical site during endoscopic neurosurgical procedures using an endonasal approach.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072410: Ksea Clearvision II, Model 40334120.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K072410

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026