## Predicates on the family tree (from the FDA summary)
- None. The FDA summary names no predicate by number.

## Recalls of this device
- Jul 21, 2009 · Not classified · Z-0853-2011: Mechanical Failure-- Patient Support System's table top may unexpectedly descend several inches.
- Jun 22, 2009 · Not classified · Z-1810-2009: Targeting accuracy out of specification, Error alert does not render the system down, which may result in mistreatment in the wrong area.
- Mar 16, 2009 · Not classified · Z-0909-2011: The product has the potential for the extension coupling of the Synchrony extension mounting hardware to become loose over time. The assembly weighs approximately 17 pounds; an unexpected contact with
- Mar 16, 2009 · Not classified · Z-0908-2011: The product has the potential for the extension coupling of the Synchrony extension mounting hardware to become loose over time. The assembly weighs approximately 17 pounds; an unexpected contact with
- Nov 14, 2008 · Not classified · Z-0629-2009: System may use random incorrect data to calculate dose. Resulting dose calculation error can exceed 100% of correct dose which may lead to serious patient injury.
- May 1, 2008 · Not classified · Z-1862-2008: Couch may move unexpectedly, which may result in patient impacting the linear accelerator.
- Apr 9, 2008 · Not classified · Z-2056-2008: Sample beam data (which should not be used to treat patients) may differ from actual radiation output of an installed product, and which may be used by users.

## Indications for use
The CyberKnife® Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.