# K072676: Acuson X300 Ultrasound System

- Pathway: 510(k) (Special)
- Applicant: Siemens Medical Solutions USA, Inc.
- Decided: Dec 18, 2007 after 88 days
- Product code: IYO (System, Imaging, Pulsed Echo, Ultrasonic)
- Device class: 2
- Regulation: 21 CFR 892.1560
- Recalls of this device: 5 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 10 generations back to K935595 (1995). 40 ancestors in all.
279 later devices descend from it; 3 cite it directly.
0 of 320 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K071036: Modification to Acuson X300 Ultrasound System, Siemens Medical Solutions USA, Inc., Ultrasound DI (2007) Summary page 6.
- K061946: Acuson X300 Diagnostic Ultrasound System, Siemens Medical Solutions USA, Inc. (2006) Summary page 1.

## Most influential descendants
- K080760: Acuson X300 Ultrasound System (2008), 278 descendants of its own
- K090276: Acuson X300 Diagnostic Ultrasound System (2009), 277 descendants of its own
- K092691: DC-7 Diagnostic Ultrasound System (2009), 176 descendants of its own
- K100830: M7/M7T Diagnostic Ultrasound System (2010), 138 descendants of its own
- K101041: DC-7 Diagnostic Ultrasound System (2010), 130 descendants of its own

## Recalls of this device
- Aug 17, 2010 · Not classified · Z-1838-2011: There is an error in the estimation of mean pressure gradient in the Cardiac measurement package. The currently implemented calculation will result in an underestimation of the mean pressure gradient 
- Jul 30, 2009 · Not classified · Z-2065-2011: Software Error with potential for misinterpretation of data-- The previous patient's last capture clip image can appear in the next patient's study.
- Jul 24, 2009 · Not classified · Z-2061-2011: Product Failure-- When the 64-Channel System V5M is used in the 128-Channel mode, an over current fault occurs which shuts the system down.
- Jul 24, 2009 · Not classified · Z-2147-2011: Mistreatment risk-- when a user defined formula is used to calculate a value in a vascular exam, data is switched laterally when it is internally transferred and displayed on the summary report within
- Aug 19, 2008 · Not classified · Z-0111-2009: Thermal Index cranial (TIC) is not displayed for the Neo-Head exam type with the C8-5 transducer. This is a required display to support the alara principle when imaging cranial structure.

## Indications for use
The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.