510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Dimension Enzymatic Creatinine Flex Reagent Cartridge with Model DF270

K073055510(k) TraditionalCleared Jan 3, 200865-day reviewClass II21 CFR 862.1225Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dade Behring, Inc. · JFY Enzymatic Method, Creatinine

Cleared on Jan 3, 2008 as substantially equivalent to K003261 (Roche Diagnostics Corp., 2000) after a 65-day review. Its family tree reaches back 1 generation to 2000. It has been recalled 5 times.

Decided
Jan 3, 2008
Review
65 days
Class
Class II
Product code
JFY
Recalls
5
Predicates
1

Its longest line of descent runs back 1 generation to K003261 (2000). No later device cites it as a predicate yet.

K003261 · Creatinine Plus · Roche Diagnostics Corp. · 2000 · recalled200020022004200620082010
  1. Root · 1 generation backK003261 Creatinine PlusRoche Diagnostics Corp. · 2000 · Recalled
  2. This deviceK073055 Dimension Enzymatic Creatinine Flex Reagent Cartridge with Model DF270Dade Behring, Inc. · 2008
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K003261Creatinine PlusRoche Diagnostics Corp.20001
0 (this device)K073055Dimension Enzymatic Creatinine Flex Reagent Cartridge with Model DF270Dade Behring, Inc.2008—
FIG. 1 — Lineage of K073055, 2000–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JFY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JFY device

  1. #1 K252619 Qscheck Uisacrscore 12.4 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K003261Creatinine Plus
Summary page 1
…edicate Device: The Dimension® ECRE Flex® reagent cartridge is substantially equivalent to Roche's Creatinine Plus Reagent (K00326 1). G. Device Description: The Dimension® ECRE Flex® reagent cartridge is a prepackaged in…
Roche Diagnostics Corp.20001

Also cleared against K003261: 5 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 4, 2020Class IIZ-2284-2020Device DesignThe presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Dec 18, 2019Class IIZ-1232-2020Under Investigation by firmThere is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
Mar 17, 2016Class IIZ-1660-2016Device DesignSiemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples…
Jun 10, 2010Not classifiedZ-2177-2011Device DesignPatient results may be falsely depressed or falsely elevated across the assay range.
Sep 29, 2008Not classifiedZ-0485-2009Component design/selectionReagent may exhibit unflagged inaccurate patient sample results.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 5 other devices also cite its predicate K003261.

How does its category compare?

All JFY devices

JFY Enzymatic Method, Creatinine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · JFY

2in 2021

2007: 82008: 42009: 32011: 22014: 22016: 12017: 12018: 22019: 22021: 22026 (partial year): 1
2007high 8 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JFY

59in 2009

2007: 193 days median2008: 170 days median2009: 59 days median
2007high 193 (2007) · low 59 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JFY

1in 2013

2008: 22009: 32010: 32013: 1
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JFY

61in 2025

2007: 12008: 42009: 222010: 222011: 422012: 242013: 392014: 42015: 72016: 162017: 162018: 342019: 562020: 342021: 572022: 92023: 362024: 342025: 61
2007high 61 (2025) · low 1 (2007)2025

MAUDE reports naming product code JFY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007819301
2008417024
2009359322
20100—322
20112—042
20120—024
20130—139
20142—04
20150—07
20161—016
20171—016
20182—034
20192—056
20200—034
20212—057
20220—09
20230—036
20240—034
20250—061
20261—0—

What is it indicated for?

The ECRE method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension® clinical chemistry system. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for other urine analytes.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073055: Dimension Enzymatic Creatinine Flex Reagent Cartridge with Model DF270.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K073055

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026