510(k) Report

510(k) · Cleared

Recalled · Class I · 2016

Modification to Avea Ventilator

K073069510(k) TraditionalCleared Jan 25, 200887-day reviewClass II21 CFR 868.5895Recalls 14 (13 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Viasys Respiratory Care, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Jan 25, 2008 as substantially equivalent to K982499 (Novametrix Medical Systems, Inc., 1998) and 2 other predicates after an 87-day review. Its family tree reaches back 3 generations to 1997. 2 of the 15 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 14 times, including 13 Class I recalls.

Decided
Jan 25, 2008
Review
87 days
Class
Class II
Product code
CBK
Recalls
14 (13 Class I)
Predicates
3

Its longest line of descent runs back 3 generations to K962517 (1997). 11 ancestors in all, the oldest K963380 (1997). 3 later devices descend from it, 1 citing it directly. 2 of the 15 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1998: 2 other ancestors1997: 1 other ancestor2001: 3 other ancestors2000: 2 other ancestors, 1 with a Class I recall12010: 1 descendant cleared2011: 1 descendant cleared2021: 1 descendant clearedK962517 · Model 7700 Quantum Pressure Support Ventilator PSV · Healthdyne Technologies, Inc. · 1997K984407 · Bipap Harmony S/T, Model 1001445 · Respironics, Inc. · 1999K022132 · Servo Ventilator System, Model 64 87 800 E407E · Siemens Medical Solutions USA, Inc. · 20021995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 6 other ancestors, 1 with a Class I recall1 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K022132 · Servo Ventilator System, Model 64 87 800 E407E · Siemens Medical Solutions USA, Inc. · 2002K984407 · Bipap Harmony S/T, Model 1001445 · Respironics, Inc. · 1999K962517 · Model 7700 Quantum Pressure Support Ventilator PSV · Healthdyne Technologies, Inc. · 19970
  1. Root · 3 generations backK962517 Model 7700 Quantum Pressure Support Ventilator PSVHealthdyne Technologies, Inc. · 1997
  2. 2 generations backK984407 Bipap Harmony S/T, Model 1001445Respironics, Inc. · 1999
  3. 1 generation backK022132 Servo Ventilator System, Model 64 87 800 E407ESiemens Medical Solutions USA, Inc. · 2002
  4. This deviceK073069 Modification to Avea VentilatorViasys Respiratory Care, Inc. · 2008
  5. Descendant · cites it directly · 2 of its ownK093094 Vela VentilatorCardinal Health 207, Inc. · 2010
  6. Descendant · 2 generations on · 1 of its ownK101643 Model LTV 1100 VentilatorCare Fusion · 2011
  7. Descendant · 3 generations on · 0 of its ownK201082 LTV2 Series VentilatorsVyaire Medical, Inc. · 2021

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K093094Vela VentilatorCardinal Health 207, Inc. · 2010 · cites it directly2descendants
  • K101643Model LTV 1100 VentilatorCare Fusion · 2011 · 2 generations on1descendant
  • K201082LTV2 Series VentilatorsVyaire Medical, Inc. · 2021 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K962517Model 7700 Quantum Pressure Support Ventilator PSVHealthdyne Technologies, Inc.1997—
−2K984407Bipap Harmony S/T, Model 1001445Respironics, Inc.1999—
−1K022132Servo Ventilator System, Model 64 87 800 E407ESiemens Medical Solutions USA, Inc.2002—
0 (this device)K073069Modification to Avea VentilatorViasys Respiratory Care, Inc.2008—
+1K093094Vela VentilatorCardinal Health 207, Inc.2010—
+2K101643Model LTV 1100 VentilatorCare Fusion2011—
+3K201082LTV2 Series VentilatorsVyaire Medical, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199820
200021
200130

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
202110
FIG. 1 — Lineage of K073069, 1997–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982499Cosmo Plus with Nico, Model 8200
Summary page 1
…Device Class: Class I1 Classification Anesthesiology Panel: Predicate Device: The predicate devices are: • Respironics NICO K982499 * Maquet Servo-I K022132 * Cosmo Plus 8000 K963380 Device The…
Novametrix Medical Systems, Inc.19981
*K022132Servo Ventilator System, Model 64 87 800 E407E
Summary page 1
…Anesthesiology Panel: Predicate Device: The predicate devices are: • Respironics NICO K982499 * Maquet Servo-I K022132 * Cosmo Plus 8000 K963380 Device The AVEA is a servo-controlled, software-driven ventila…
Siemens Medical Solutions USA, Inc.2002—
*K963380CO 2 Smo Plus
Summary page 1
…edicate devices are: • Respironics NICO K982499 * Maquet Servo-I K022132 * Cosmo Plus 8000 K963380 Device The AVEA is a servo-controlled, software-driven ventilator. It has a dynamic Description: range of…
Novametrix Medical Systems, Inc.19973

Also cleared against K982499: 2 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093094Vela VentilatorCardinal Health 207, Inc.20102

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 17, 2016Class IZ-1888-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1895-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1896-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1899-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1892-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1897-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1898-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1894-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1889-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1891-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1890-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
May 17, 2016Class IZ-1893-2016Under Investigation by firmCareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface…
Sep 5, 2013Class IZ-2273-2013Component design/selectionCareFusion has identified a potential risk associated with AVEA¿ ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is voluntarily initiating a…
Sep 8, 2011Not classifiedZ-0510-2012Nonconforming Material/ComponentThe recall was initiated because Carefusion has identified a potential risk associated with certain AVEA ventilators and affected replacement parts manufactured between March 1, 2009 and June 30, 2011. Carefusion is voluntarily initiating…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 2 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 11 years and passes through 4 companies; a new generation was cleared every 3.3 years on median.
  • 5 of the 11 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K093094 (Cardinal Health 207, Inc., 2010) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K982499.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on adult, pediatric, and neonatal patients. It should only be operated by properly trained clinical personnel, under the direction of a physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073069: Modification to Avea Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K073069

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.