510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

GE Logiq P6

K073297510(k) SpecialCleared Dec 18, 200725-day reviewClass II21 CFR 892.1550Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

General Electric Co. · IYN System, Imaging, Pulsed Doppler, Ultrasonic

Cleared on Dec 18, 2007 as substantially equivalent to K060993 (General Electric Co., 2006) after a 25-day review. Its family tree reaches back 4 generations to 2001. 224 later devices descend from it. It has been recalled 2 times.

Decided
Dec 18, 2007
Review
25 days
Class
Class II
Product code
IYN
Recalls
2
Predicates
1

Its longest line of descent runs back 4 generations to K010329 (2001). 224 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2010: 2 descendants cleared2014: 5 descendants cleared2012: 1 descendant cleared2011: 1 descendant cleared2013: 2 descendants cleared2016: 10 descendants cleared2015: 6 descendants cleared2017: 16 descendants cleared2018: 20 descendants cleared2025: 17 descendants cleared2023: 18 descendants cleared2019: 18 descendants cleared2020: 29 descendants cleared2021: 18 descendants cleared2022: 25 descendants cleared2024: 19 descendants cleared2026: 17 descendants clearedK010329 · LOGIQ 500 · General Electric Co. · 2001K014097 · GE Logiq 5 · GE Medical Systems · 2002 · recalledK032974 · GE Logiq 5 Pro or Expert · General Electric Co. · 2003K060993 · GE Logiq P5 and A5 · General Electric Co. · 2006200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 6 descendants4 generations forward: 10 descendants5 generations forward: 25 descendants6 generations forward: 33 descendants7 generations forward: 37 descendants8 generations forward: 38 descendants9 generations forward: 29 descendants10 generations forward: 22 descendants11 generations forward: 9 descendants12 generations forward: 5 descendants13 generations forward: 4 descendants14 generations forward: 1 descendant K060993 · GE Logiq P5 and A5 · General Electric Co. · 2006K032974 · GE Logiq 5 Pro or Expert · General Electric Co. · 2003K014097 · GE Logiq 5 · GE Medical Systems · 2002 · recalledK010329 · LOGIQ 500 · General Electric Co. · 20010+5+10
  1. Root · 4 generations backK010329 LOGIQ 500General Electric Co. · 2001
  2. 3 generations backK014097 GE Logiq 5GE Medical Systems · 2002 · Recalled
  3. 2 generations backK032974 GE Logiq 5 Pro or ExpertGeneral Electric Co. · 2003
  4. 1 generation backK060993 GE Logiq P5 and A5General Electric Co. · 2006
  5. This deviceK073297 GE Logiq P6General Electric Co. · 2007
  6. Descendant · cites it directly · 221 of its ownK101874 GE Logiq P6/P6 ProGE Healthcare · 2010
  7. Descendant · 2 generations on · 220 of its ownK122114 Logiq S7 Expert, Logiq S7 ProGE Healthcare · 2012 · Recalled
  8. Descendant · 3 generations on · 219 of its ownK131527 Logiq S8GE Healthcare · 2013
  9. Descendant · 4 generations on · 155 of its ownK133034 Logiq F SeriesGE Healthcare · 2014
  10. Descendant · 4 generations on · 149 of its ownK133533 Logiq EGE Healthcare · 2014

Showing 5 of 224 descendants. See all descendants

Descendants cleared per year

2010: 22011: 12012: 12013: 22014: 52015: 62016: 102017: 162018: 202019: 182020: 292021: 182022: 252023: 182024: 192025: 172026: 17
2010high 29 (2020) · low 1 (2011)2026

Most influential descendants

Showing 5 of 224 · See all
  • K101874GE Logiq P6/P6 ProGE Healthcare · 2010 · cites it directly221descendants
  • K122114Logiq S7 Expert, Logiq S7 ProGE Healthcare · 2012 · 2 generations on220descendants
  • K131527Logiq S8GE Healthcare · 2013 · 3 generations on219descendants
  • K133034Logiq F SeriesGE Healthcare · 2014 · 4 generations on155descendants
  • K133533Logiq EGE Healthcare · 2014 · 4 generations on149descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K010329LOGIQ 500General Electric Co.2001—
−3K014097GE Logiq 5GE Medical Systems20021
−2K032974GE Logiq 5 Pro or ExpertGeneral Electric Co.2003—
−1K060993GE Logiq P5 and A5General Electric Co.2006—
0 (this device)K073297GE Logiq P6General Electric Co.2007—
+1K101874GE Logiq P6/P6 ProGE Healthcare2010—
+2K122114Logiq S7 Expert, Logiq S7 ProGE Healthcare20121
+3K131527Logiq S8GE Healthcare2013—
+4K133034Logiq F SeriesGE Healthcare2014—
+4K133533Logiq EGE Healthcare2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201020
201110
201210
201320
201450
201560
2016100
2017160
2018200
2019180
2020290
2021180
2022250
2023180
2024190
2025170
2026170
FIG. 1 — Lineage of K073297, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYN device

  1. #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060993GE Logiq P5 and A5
Summary page 1
…Ultrasound System Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN 3. Marketed Device: GE LOGIQ P5 Diagnostic Ultrasound System, K060993 currently in commercial distribution. 4. Device Description: The GE LOGIQ P6 is full-featured, general-pu…
General Electric Co.2006—

Also cleared against K060993: 11 other devices →

What descends from it? (224)

See all 224

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K101874GE Logiq P6/P6 ProGE Healthcare2010221
K102104GE Vivid P3 Diagnostic Ultrasound Imaging SystemWipro GE Healthcare Private, Ltd.20101

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 28, 2018Class IIZ-0788-2019Device DesignIncreased temperature in the probe-head surface with LOGIQ P6 Model BT07 and BT09 scanners when used in specific types of scans and with specific probes.
Dec 23, 2011Not classifiedZ-0834-2012Software designGE Healthcare has recently become aware of a safety issue associated with the Probe of your LOGIC P6 system that may impact patient safety. GE Logic P6 probe head has a possibility to cause burn injuries to patients or operators.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 2 companies; a new generation was cleared every 1.8 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K101874 (GE Healthcare, 2010) has gone furthest, with 221 descendants of its own.
  • 11 other devices also cite its predicate K060993.

How does its category compare?

All IYN devices

IYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · IYN

57in 2025

2007: 132008: 252009: 182010: 292011: 392012: 432013: 512014: 562015: 542016: 602017: 642018: 592019: 472020: 572021: 462022: 532023: 442024: 502025: 572026 (partial year): 44
2007high 64 (2017) · low 13 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYN

124in 2025

2007: 26 days median2008: 18 days median2009: 29 days median2010: 67 days median2011: 46 days median2012: 37 days median2013: 60 days median2014: 39 days median2015: 55 days median2016: 56 days median2017: 83 days median2018: 85 days median2019: 79 days median2020: 72 days median2021: 105 days median2022: 105 days median2023: 115 days median2024: 119 days median2025: 124 days median2026 (partial year): 107 days median
2007high 124 (2025) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYN

10in 2025

2007: 162008: 122009: 82010: 102011: 82012: 32013: 82014: 22015: 32016: 22018: 72020: 42021: 82022: 102023: 42024: 12025: 102026 (partial year): 2, 1 Class I
2007high 16 (2007) · low 0 (2017)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · IYN

610in 2025

2007: 32008: 22009: 92010: 172011: 182012: 202013: 182014: 482015: 1162016: 1452017: 1462018: 852019: 592020: 672021: 2132022: 1862023: 2292024: 2662025: 610
2007high 610 (2025) · low 2 (2008)2025

MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071326163
20082518122
2009182989
201029671017
20113946818
20124337320
20135160818
20145639248
201554553116
201660562145
201764830146
20185985785
20194779059
20205772467
2021461058213
20225310510186
2023441154229
2024501191266
20255712410610
2026441072 (1)—

What is it indicated for?

The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications of: Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal (TE); Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, vascular and neurological).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073297: GE Logiq P6.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K073297

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026