Recalled · Class I · 2022Sensititre 18-24 Hour Susceptibility Plates
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Trek Diagnostic Systems, Inc. · JWY Manual Antimicrobial Susceptibility Test Systems
Cleared on Jan 22, 2008 after a 48-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. It has a Class I recall. It has been recalled 9 times, including 9 Class I recalls.
- Decided
- Jan 22, 2008
- Review
- 48 days
- Class
- Class II
- Product code
- JWY
- Recalls
- 9 (9 Class I)
- Predicates
- 0
FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JWY device
- #1 K260447 ETEST Gepotidacin (GEP) (0.016-256 µg/mL)score 12.7 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 20, 2022 | Class I | Z-0327-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0329-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0334-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0330-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0328-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0331-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0332-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0333-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0335-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
The 5 latest of 9 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
How does its category compare?
All JWY devicesJWY Manual Antimicrobial Susceptibility Test Systems: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · JWY
14in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JWY
106in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JWY
31in 2022
Class I (13 since 2007) Class II/III not classified
Adverse event reports per year · JWY
1in 2024
MAUDE reports naming product code JWY, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 6 | 85 | 1 | 1 |
| 2009 | 2 | — | 0 | 1 |
| 2010 | 4 | 73 | 0 | 0 |
| 2011 | 4 | 125 | 3 | 0 |
| 2012 | 1 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 6 | 106 | 5 | 4 |
| 2015 | 9 | 90 | 6 (4) | 11 |
| 2016 | 4 | 88 | 0 | 20 |
| 2017 | 14 | 61 | 2 | 9 |
| 2018 | 12 | 51 | 0 | 13 |
| 2019 | 18 | 60 | 0 | 10 |
| 2020 | 5 | 66 | 0 | 20 |
| 2021 | 5 | 266 | 0 | 41 |
| 2022 | 0 | — | 31 (9) | 11 |
| 2023 | 7 | 85 | 0 | 9 |
| 2024 | 6 | 90 | 0 | 1 |
| 2025 | 14 | 106 | 0 | 0 |
| 2026 | 2 | — | 0 | — |
https://510kreport.com/device/K073424 · printed Nov 5, 2026
Cite this page
510(k) Report. “K073424: Sensititre 18-24 Hour Susceptibility Plates.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K073424
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Page last updated Nov 3, 2026