# K080347: Uniscan -Doa System

- Pathway: 510(k) (Traditional)
- Applicant: Taiwan Unison Biotech, Inc.
- Decided: Oct 9, 2008 after 244 days
- Product code: DJC (Thin Layer Chromatography, Methamphetamine)
- Device class: 2
- Regulation: 21 CFR 862.3610
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K012745 (2002). 2 ancestors in all.
No later device cites it as a predicate yet.
0 of 3 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K043242: Triage Tox Drug Screen, Biosite Incorporated (2005) Summary page 2.

## Indications for use
TUBI's UNISCAN-DoA is a system intended for use in Drugs of Abuse Screening Tests. This mAMP/Opi/THC panel test is a prescription assay intended for use with UNISCAN-DoA scanner in laboratory by professional personnel. The mAMP/Opi/THC assays were calibrated with d-methamphetamine/morphine/ 11 -nor-A 9 -THC-9-COOH, respectively. It provides qualitative screening results for Methamphetamine/ Opiate/cannabinoids in human urine at a cutoff concentration of 1000/300/50 ng/ml. For In Vitro Diagnostic Use. This assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any Drug of Abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) is the recommended confirmatory method.

Source page: https://510kreport.com/device/K080347