Findrwirz Guide Wire System

K080364510(k) TraditionalCleared Aug 11, 2008182-day reviewClass II21 CFR 870.1330Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sentreheart, Inc. · DQX Wire, Guide, Catheter

Cleared on Aug 11, 2008 as substantially equivalent to K914138 (Lake Region Mfg., Inc., 1991) and 2 other predicates after a 182-day review. Its family tree reaches back 4 generations to 1994. 2 of the 23 devices in its family, counting itself, have a Class I recall. It has been recalled once, including a Class I recall.

Decided
Aug 11, 2008
Review
182 days
Class
Class II
Product code
DQX
Recalls
1 (1 Class I)
Predicates
3

Its longest line of descent runs back 4 generations to K941895 (1994). 22 ancestors in all, the oldest K901502 (1990). No later device cites it as a predicate yet. 2 of the 23 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2007: 1 other ancestor1991: 2 other ancestors2002: 1 other ancestor1998: 1 other ancestor1997: 2 other ancestors, 1 with a Class I recall11996: 2 other ancestors1995: 6 other ancestors1994: 1 other ancestor1993: 1 other ancestor1990: 1 other ancestorK941895 · Scimed 0.014 Ptca Guide Wire with Ice Hydrophilic Coating · Scimed Life Systems, Inc. · 1994K970244 · Scimed Choice Family of Ptca Guide Wires · Scimed Life Systems, Inc. · 1997K021363 · Endovascular Guide Wire · Stereotaxis, Inc. · 2002 · recalledK042854 · Stereotaxis Endovascular Guide Wire · Stereotaxis, Inc. · 200519901995200020052010
Other ancestors, by generation 1 generation back: 2 other ancestors2 generations back: 1 other ancestor3 generations back: 3 other ancestors, 1 with a Class I recall14 generations back: 12 other ancestors K042854 · Stereotaxis Endovascular Guide Wire · Stereotaxis, Inc. · 2005K021363 · Endovascular Guide Wire · Stereotaxis, Inc. · 2002 · recalledK970244 · Scimed Choice Family of Ptca Guide Wires · Scimed Life Systems, Inc. · 1997K941895 · Scimed 0.014 Ptca Guide Wire with Ice Hydrophilic Coating · Scimed Life Systems, Inc. · 19940
  1. Root · 4 generations backK941895 Scimed 0.014 Ptca Guide Wire with Ice Hydrophilic CoatingScimed Life Systems, Inc. · 1994
  2. 3 generations backK970244 Scimed Choice Family of Ptca Guide WiresScimed Life Systems, Inc. · 1997
  3. 2 generations backK021363 Endovascular Guide WireStereotaxis, Inc. · 2002 · Recalled
  4. 1 generation backK042854 Stereotaxis Endovascular Guide WireStereotaxis, Inc. · 2005
  5. This deviceK080364 Findrwirz Guide Wire SystemSentreheart, Inc. · 2008

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K941895Scimed 0.014 Ptca Guide Wire with Ice Hydrophilic CoatingScimed Life Systems, Inc.1994—
−3K970244Scimed Choice Family of Ptca Guide WiresScimed Life Systems, Inc.1997—
−2K021363Endovascular Guide WireStereotaxis, Inc.20026
−1K042854Stereotaxis Endovascular Guide WireStereotaxis, Inc.2005—
0 (this device)K080364Findrwirz Guide Wire SystemSentreheart, Inc.2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199120
199310
199410
199560
199620
199721
199810
200210
200710
FIG. 1 — Lineage of K080364, 1994–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQX device

  1. #1 K213854 SavvyWirescore 15.7 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K914138Ontrac
Summary page 2
…ntional devices and Inferior Vena Cava (IVC) filters. The FindrWIRZ system is also contraindicated for use in MRI procedures. Predicate Devices: OnTrac, Lake Region (K914138) Cronus Wire, Stereotaxis (K042854) Radius Snare Device, Radius Medical Technologies (K071457) Safety and Performance: Perform…
Lake Region Mfg., Inc.1991—
*K042854Stereotaxis Endovascular Guide Wire
Summary page 2
…va (IVC) filters. The FindrWIRZ system is also contraindicated for use in MRI procedures. Predicate Devices: OnTrac, Lake Region (K914138) Cronus Wire, Stereotaxis (K042854) Radius Snare Device, Radius Medical Technologies (K071457) Safety and Performance: Performance Functional testing was conducte…
Stereotaxis, Inc.2005—
*K071457Modification to Radius Snare
Summary page 2
…ted for use in MRI procedures. Predicate Devices: OnTrac, Lake Region (K914138) Cronus Wire, Stereotaxis (K042854) Radius Snare Device, Radius Medical Technologies (K071457) Safety and Performance: Performance Functional testing was conducted to support the claim of substantial equivalence and to dem…
Radius Medical Technologies, Inc.2007—

Also cleared against K914138: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 29, 2016Class IZ-0328-2017Under Investigation by firmA complaint was received on September 27th, 2016 at a medical center in Germany for delamination and flaking of the PTFE lubricious coating on three FindrWIRZ Guidewire System .025 guidewires removed from their packaging hoops.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K970466Radius Ptca GuidewireRadius Medical Technologies, Inc. · 1997 · 3 generations back · ◆ in the 1997 ancestors bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 3.5 years on median.
  • 5 of the 22 ancestors have a recall of any class; 1 has a Class I recall.
  • 1 other device also cites its predicate K914138.

How does its category compare?

All DQX devices

DQX Wire, Guide, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · DQX

10in 2025

2007: 142008: 192009: 182010: 212011: 182012: 202013: 182014: 182015: 152016: 142017: 192018: 192019: 132020: 142021: 62022: 152023: 102024: 112025: 102026 (partial year): 12
2007high 21 (2010) · low 6 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQX

177in 2025

2007: 85 days median2008: 87 days median2009: 87 days median2010: 63 days median2011: 43 days median2012: 62 days median2013: 119 days median2014: 103 days median2015: 112 days median2016: 179 days median2017: 85 days median2018: 114 days median2019: 173 days median2020: 106 days median2021: 161 days median2022: 212 days median2023: 136 days median2024: 134 days median2025: 177 days median2026 (partial year): 145 days median
2007high 212 (2022) · low 43 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQX

3in 2024

2007: 122008: 72009: 492010: 342011: 32012: 52013: 7, 4 Class I2015: 12016: 2, 2 Class I2017: 172018: 92019: 22020: 1, 1 Class I2024: 32026 (partial year): 4, 1 Class I
2007high 49 (2009) · low 0 (2014)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · DQX

2,178in 2025

2007: 7602008: 7152009: 6872010: 9872011: 1,3112012: 1,3372013: 1,4742014: 1,3232015: 1,3232016: 1,4662017: 1,5242018: 1,7372019: 1,6792020: 1,5362021: 1,7132022: 1,6472023: 2,0682024: 2,2102025: 2,178
2007high 2,210 (2024) · low 687 (2009)2025

MAUDE reports naming product code DQX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007148512760
200819877715
2009188749687
2010216334987
2011184331,311
2012206251,337
2013181197 (4)1,474
20141810301,323
20151511211,323
2016141792 (2)1,466
20171985171,524
20181911491,737
20191317321,679
2020141061 (1)1,536
2021616101,713
20221521201,647
20231013602,068
20241113432,210
20251017702,178
2026121454 (1)—

What is it indicated for?

The FindrWIRZ System is intended for use in the cardiovascular system for introduction and positioning of over-the-wire catheters and therapeutic devices during interventional procedures. A FindrWIRZ may also be used to manipulate and or reposition another FindrWIRZ.

The FindrWIRZ System is not intended for use in the coronary or cerebral vasculature.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080364: Findrwirz Guide Wire System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K080364

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026