510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Drive Solstice Oxygen Concentrator

K080391510(k) TraditionalCleared Mar 18, 200834-day reviewClass II21 CFR 868.5440Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medical Depot · CAW Generator, Oxygen, Portable

Cleared on Mar 18, 2008 as substantially equivalent to K071608 (Zhongshan A&J Medical Equipment Co., Ltd., 2008) after a 34-day review. Its family tree reaches back 3 generations to 1988. 2 of the 10 devices in its family, counting itself, have a Class I recall. 5 later devices descend from it. It has been recalled once.

Decided
Mar 18, 2008
Review
34 days
Class
Class II
Product code
CAW
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K883591 (1988). 4 ancestors in all. 5 later devices descend from it, 1 citing it directly. 2 of the 10 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1989: 1 other ancestor2009: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12026: 2 descendants cleared2025: 1 descendant clearedReference device: K020386 · Invacare Platinum, Models Irc 5LX and Irc 5LX02 · Invacare Corp. · 2002 · recalledK883591 · Mark-5 Concentrator · Nidek Medical Products, Inc. · 1988K032509 · Mark 5 Nuvo (M5C5) · Nidek Medical Products, Inc. · 2004 · Class I recallK071608 · A&J POCA01 Oxygen Concentrator · Zhongshan A&J Medical Equipment Co., Ltd. · 20081990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant, 1 with a Class I recall13 generations forward: 3 descendants K071608 · A&J POCA01 Oxygen Concentrator · Zhongshan A&J Medical Equipment Co., Ltd. · 2008K032509 · Mark 5 Nuvo (M5C5) · Nidek Medical Products, Inc. · 2004 · Class I recallK883591 · Mark-5 Concentrator · Nidek Medical Products, Inc. · 19880
  1. Root · 3 generations backK883591 Mark-5 ConcentratorNidek Medical Products, Inc. · 1988
  2. 2 generations backK032509 Mark 5 Nuvo (M5C5)Nidek Medical Products, Inc. · 2004 · Recalled (Class I)
  3. 1 generation backK071608 A&J POCA01 Oxygen ConcentratorZhongshan A&J Medical Equipment Co., Ltd. · 2008
  4. This deviceK080391 Drive Solstice Oxygen ConcentratorMedical Depot · 2008
  5. Descendant · cites it directly · 4 of its ownK090007 Jumao Oxygen Concentrator, Model JM-07000HI, JM-07000I, JM-07000Danyang Jumao Healthcare Equipment Co., Ltd. · 2009
  6. Descendant · 2 generations on · 3 of its ownK230969 Oxygen ConcentratorJiangsu Jumao X-Care Medical Equipment Co., Ltd. · 2023 · Recalled (Class I)
  7. Descendant · 3 generations on · 0 of its ownK252965 Oxygen Concentrator (OX-5A, OX-5C)Oxytek Medical Technology Co., Ltd. · 2026
  8. Descendant · 3 generations on · 0 of its ownK253549 DeVilbiss 5 Liter Oxygen Concentrator (555)Devilbiss Healthcare, LLC · 2026
  9. Descendant · 3 generations on · 0 of its ownK250805 DeVilbiss PulmO2 10-Liter Oxygen Concentrator (1060AW)Devilbiss Healthcare, LLC · 2025

Class I recalls in this family

Most influential descendants

Showing 5 of 5 · See all
  • K090007Jumao Oxygen Concentrator, Model JM-07000HI, JM-07000I, JM-07000Danyang Jumao Healthcare Equipment Co., Ltd. · 2009 · cites it directly4descendants
  • K230969Oxygen ConcentratorJiangsu Jumao X-Care Medical Equipment Co., Ltd. · 2023 · 2 generations on3descendants
  • K252965Oxygen Concentrator (OX-5A, OX-5C)Oxytek Medical Technology Co., Ltd. · 2026 · 3 generations on0descendants
  • K253549DeVilbiss 5 Liter Oxygen Concentrator (555)Devilbiss Healthcare, LLC · 2026 · 3 generations on0descendants
  • K250805DeVilbiss PulmO2 10-Liter Oxygen Concentrator (1060AW)Devilbiss Healthcare, LLC · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K883591Mark-5 ConcentratorNidek Medical Products, Inc.1988—
−2K032509Mark 5 Nuvo (M5C5)Nidek Medical Products, Inc.2004Class I
−1K071608A&J POCA01 Oxygen ConcentratorZhongshan A&J Medical Equipment Co., Ltd.2008—
0 (this device)K080391Drive Solstice Oxygen ConcentratorMedical Depot2008—
+1K090007Jumao Oxygen Concentrator, Model JM-07000HI, JM-07000I, JM-07000Danyang Jumao Healthcare Equipment Co., Ltd.2009—
+2K230969Oxygen ConcentratorJiangsu Jumao X-Care Medical Equipment Co., Ltd.2023Class I
+3K252965Oxygen Concentrator (OX-5A, OX-5C)Oxytek Medical Technology Co., Ltd.2026—
+3K253549DeVilbiss 5 Liter Oxygen Concentrator (555)Devilbiss Healthcare, LLC2026—
+3K250805DeVilbiss PulmO2 10-Liter Oxygen Concentrator (1060AW)Devilbiss Healthcare, LLC2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
202311
202510
202620
FIG. 1 — Lineage of K080391, 1988–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CAW device

  1. #1 K230052 Inogen Rove 6score 29.9 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071608A&J POCA01 Oxygen Concentrator
Summary page 2
…e 90% oxygen to the patient. Predicate Device: Invacare Platinum 5 Oxygen Concentrator - K020386 - Invacare Corporation A&J-POCAO1 Oxygen Concentrator - K071608 - Zhongshan A&J Medical Equipment Co., Ltd Substantial Equivalence: Medical Depot claims the proposed device to be substantiall…
Zhongshan A&J Medical Equipment Co., Ltd.2008—
†K020386Invacare Platinum, Models Irc 5LX and Irc 5LX02Invacare Corp.20022

Also cleared against K071608: 3 other devices →

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090007Jumao Oxygen Concentrator, Model JM-07000HI, JM-07000I, JM-07000Danyang Jumao Healthcare Equipment Co., Ltd.20094

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 9, 2017Class IIZ-0407-2018Under Investigation by firmPost-market surveillance for Pure Stationary Oxygen Concentrators model CHSOOO and CHSOOOS demonstrate an aggregate complaint/service rate greater than 35% for low oxygen purity (output oxygen below the published specification).

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K032509Mark 5 Nuvo (M5C5)Nidek Medical Products, Inc. · 2004 · 2 generations back · ◆ on the line of descent in FIG. 17 Class I recalls
  • K230969Oxygen ConcentratorJiangsu Jumao X-Care Medical Equipment Co., Ltd. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 20 years and passes through 3 companies; a new generation was cleared every 3.8 years on median.
  • 1 of the 4 ancestors has a recall of any class; 1 has a Class I recall.
  • Of its descendants, K090007 (Danyang Jumao Healthcare Equipment Co., Ltd., 2009) has gone furthest, with 4 descendants of its own.
  • 3 other devices also cite its predicate K071608.

How does its category compare?

All CAW devices

CAW Generator, Oxygen, Portable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · CAW

12in 2025

2007: 62008: 112009: 82010: 52011: 62012: 52013: 22014: 62015: 12016: 12017: 12018: 32019: 22020: 52021: 22022: 42023: 82024: 22025: 122026 (partial year): 6
2007high 12 (2025) · low 1 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CAW

178in 2025

2007: 143 days median2008: 90 days median2009: 193 days median2010: 133 days median2011: 174 days median2012: 148 days median2014: 277 days median2018: 203 days median2020: 224 days median2022: 458 days median2023: 261 days median2025: 178 days median2026 (partial year): 260 days median
2007high 458 (2022) · low 90 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CAW

1in 2025

2007: 12008: 32009: 22011: 12013: 22014: 22016: 12017: 12019: 22020: 12022: 22024: 5, 1 Class I2025: 1
2007high 5 (2024) · low 0 (2010)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · CAW

611in 2025

2007: 232008: 252009: 432010: 482011: 1672012: 3952013: 2,3092014: 24,5382015: 18,9732016: 3,1672017: 1072018: 1162019: 1202020: 1252021: 1942022: 3052023: 3572024: 4502025: 611
2007high 24,538 (2014) · low 23 (2007)2025

MAUDE reports naming product code CAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076143123
20081190325
20098193243
20105133048
201161741167
201251480395
20132—22,309
20146277224,538
20151—018,973
20161—13,167
20171—1107
201832030116
20192—2120
202052241125
20212—0194
202244582305
202382610357
20242—5 (1)450
2025121781611
202662600—

What is it indicated for?

The intended function and use of the Drive Solistice Oxygen Concentrator (models 18050 and 18055) is to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve. It is not intended to sustain or support life.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080391: Drive Solstice Oxygen Concentrator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K080391

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026