Heart Sync, Models C-100 and T-100

K080421510(k) TraditionalCleared Feb 29, 200814-day reviewClass III21 CFR 870.5310Recalls 11 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Heart Sync, LLC · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Feb 29, 2008 as substantially equivalent to K020203 (Padpro, LLC, 2002) and 1 other predicate after a 14-day review. Its family tree reaches back 2 generations to 1998. 2 of the 7 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 11 times, including 5 Class I recalls.

Decided
Feb 29, 2008
Review
14 days
Class
Class III
Product code
MKJ
Recalls
11 (5 Class I)
Predicates
2

Its longest line of descent runs back 2 generations to K981737 (1998). 4 later devices descend from it, 4 citing it directly. 2 of the 7 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2011: 3 descendants cleared2013: 1 descendant clearedK981737 · K-Defib/Pace Adult Electrode Model Number KDP-60 · Katecho, Inc. · 1998K020203 · Padpro Model #2516 · Padpro, LLC · 2002 · Class I recall1995200020052010201520202025
Descendants, by generation 1 generation forward: 4 descendants K020203 · Padpro Model #2516 · Padpro, LLC · 2002 · Class I recallK981737 · K-Defib/Pace Adult Electrode Model Number KDP-60 · Katecho, Inc. · 19980
  1. Root · 2 generations backK981737 K-Defib/Pace Adult Electrode Model Number KDP-60Katecho, Inc. · 1998
  2. 1 generation backK020203 Padpro Model #2516Padpro, LLC · 2002 · Recalled (Class I)
  3. This deviceK080421 Heart Sync, Models C-100 and T-100Heart Sync, LLC · 2008
  4. Descendant · cites it directly · 0 of its ownK131550 Heart SyncHeart Sync, Inc. · 2013
  5. Descendant · cites it directly · 0 of its ownK103783 Euro Defi Pads Adult; Euro Defi Pads Adult Radiotransparent: Euro…Fiab Spa · 2011
  6. Descendant · cites it directly · 0 of its ownK103334 Cardio MedCardio Medical Products, Inc. · 2011
  7. Descendant · cites it directly · 0 of its ownK103159 Cardio Med Model A102, H104, O101, P103, and Z100Cardio Medical Products, Inc. · 2011 · Recalled

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K131550Heart SyncHeart Sync, Inc. · 2013 · cites it directly0descendants
  • K103783Euro Defi Pads Adult; Euro Defi Pads Adult Radiotransparent: Euro Defi Pads PediatricFiab Spa · 2011 · cites it directly0descendants
  • K103334Cardio MedCardio Medical Products, Inc. · 2011 · cites it directly0descendants
  • K103159Cardio Med Model A102, H104, O101, P103, and Z100Cardio Medical Products, Inc. · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K981737K-Defib/Pace Adult Electrode Model Number KDP-60Katecho, Inc.1998—
−1K020203Padpro Model #2516Padpro, LLC2002Class I
0 (this device)K080421Heart Sync, Models C-100 and T-100Heart Sync, LLC2008—
+1K131550Heart SyncHeart Sync, Inc.2013—
+1K103783Euro Defi Pads Adult; Euro Defi Pads Adult Radiotransparent: Euro Defi Pads…Fiab Spa2011—
+1K103334Cardio MedCardio Medical Products, Inc.2011—
+1K103159Cardio Med Model A102, H104, O101, P103, and Z100Cardio Medical Products, Inc.20111

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201130
201310
FIG. 1 — Lineage of K080421, 1998–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020203Padpro Model #2516
Summary page 1
…rocardiograph, Multi-Function; MLN Common/Usual Name: Defibrillator Electrode 2. Equivalent legally marketed devices: This is identical in function to the PadPro (K020203), and Katecho (K981737) and identical in design. 3. Indications for Use: The Heart Sync Adult Radio Transparent and/or Transluce…
Padpro, LLC2002Class I
*K981737K-Defib/Pace Adult Electrode Model Number KDP-60
Summary page 1
…nction; MLN Common/Usual Name: Defibrillator Electrode 2. Equivalent legally marketed devices: This is identical in function to the PadPro (K020203), and Katecho (K981737) and identical in design. 3. Indications for Use: The Heart Sync Adult Radio Transparent and/or Translucent Multi Function El…
Katecho, Inc.1998—

Also cleared against K020203: 2 other devices →

What descends from it? (4)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131550Heart SyncHeart Sync, Inc.20130
K103783Euro Defi Pads Adult; Euro Defi Pads Adult Radiotransparent: Euro Defi Pads PediatricFiab Spa20110
K103334Cardio MedCardio Medical Products, Inc.20110
K103159Cardio Med Model A102, H104, O101, P103, and Z100Cardio Medical Products, Inc.20110

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 11, 2014Class IZ-0584-2015Device DesignHeart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads…
Nov 11, 2014Class IZ-0585-2015Device DesignHeart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads…
Nov 11, 2014Class IZ-0583-2015Device DesignHeart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads…
Nov 11, 2014Class IZ-0586-2015Device DesignHeart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads…
Nov 11, 2014Class IZ-0587-2015Device DesignHeart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads…
Apr 26, 2013Class IIZ-1434-2013Process designOn 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012.…
Apr 26, 2013Class IIZ-1435-2013Process designOn 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012.…
Dec 17, 2010Not classifiedZ-1372-2011PackagingThe seal on the electrode pouch may be open, exposing the electrode to the outside environment. The package does have a caution statement that states, " Do not open package until immediately prior to using electrodes. and Electrodes may…
Dec 17, 2010Not classifiedZ-1371-2011PackagingThe seal on the electrode pouch may be open, exposing the electrode to the outside environment. The package does have a caution statement that states, " Do not open package until immediately prior to using electrodes. and Electrodes may…
Dec 5, 2008Not classifiedZ-1950-2011Equipment maintenanceThe "leads out" connector had bent pins which would not allow connection to the Medtronic Physio Control Defibrillator cable during a product demonstration.
Dec 5, 2008Not classifiedZ-1951-2011Equipment maintenanceThe "leads out" connector had bent pins which would not allow connection to the Medtronic Physio Control Defibrillator cable during a product demonstration.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K020203Padpro Model #2516Padpro, LLC · 2002 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 5.8 years on median.
  • 1 of the 2 ancestors has a recall of any class; 1 has a Class I recall.
  • Of its descendants, K131550 (Heart Sync, Inc., 2013) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K020203.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The Heart Sync Adult Radio Transparent Multi Function Electrodes and Heart Sync Adult Radiotranslucent Multifunction Electrodes are indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single patient use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patient's skin. The electrode…

Read the full statement (125 words)

The Heart Sync Adult Radio Transparent Multi Function Electrodes and Heart Sync Adult Radiotranslucent Multifunction Electrodes are indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single patient use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patient's skin. The electrode is intended for use on adult patients. When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (360 joule maximum).

Compatible with Medtronic Physio Control, Zoll, Hewlett-Packard, Philips Medical, and Welch-Allyn models of monophasic and bi-phasic defibrillators, external pacemakers.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080421: Heart Sync, Models C-100 and T-100.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K080421

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026