510(k) · Cleared
ClearedRapidsense Drugs of Abuse Phencyclidine (PCP) 25 Device with Models 900-0054
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Quantrx Biomedical Corporation · LCM Enzyme Immunoassay, Phencyclidine
Cleared on Jan 8, 2009 as substantially equivalent to K011730 (ACON Laboratories, Inc., 2001) after a 324-day review. Its family tree reaches back 2 generations to 1996.
- Decided
- Jan 8, 2009
- Review
- 324 days
- Class
- Class U
- Product code
- LCM
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 2 generations to K961266 (1996). No later device cites it as a predicate yet.
- Root · 2 generations backK961266 Accusign, Biosign PCPPrinceton BioMeditech Corp. · 1996
- 1 generation backK011730 Acon PCP One-Step Phencyclidine Test Strip, Acon PCP One-Step…ACON Laboratories, Inc. · 2001
- This deviceK080431 Rapidsense Drugs of Abuse Phencyclidine (PCP) 25 Device with Models…Quantrx Biomedical Corporation · 2009
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K961266 | Accusign, Biosign PCP | Princeton BioMeditech Corp. | 1996 | — |
| −1 | K011730 | Acon PCP One-Step Phencyclidine Test Strip, Acon PCP One-Step Phencyclidine… | ACON Laboratories, Inc. | 2001 | — |
| 0 (this device) | K080431 | Rapidsense Drugs of Abuse Phencyclidine (PCP) 25 Device with Models 900-0054 | Quantrx Biomedical Corporation | 2009 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LCM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LCM device
- #1 K212952 Psychemedics Homogeneous Enzyme Immunoassay…score 3.5 · 2 generations · no Class I recall
- #2 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K011730 | Acon PCP One-Step Phencyclidine Test Strip, Acon PCP One-Step Phencyclidine Test DeviceSummary page 2…30 Comparison to Predicate Device Both the RapidSense TM Drugs of Abuse Phencyclidine (PCP) 25 device and the ACON PCP One Step Phencyclidine Test Device, cleared under K011730 have the indications for use and same cut-off of 25 ng/mL. Both assays are lateral flow competitive immunoassays which provide a… | 2001 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 7.4 years on median.
- None of the 2 ancestors has a recall.
- 5 other devices also cite its predicate K011730.
How does its category compare?
All LCM devicesLCM Enzyme Immunoassay, Phencyclidine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.
FDA decisions per year · LCM
1in 2022
2026 to Oct 25, outlined.
Median 510(k) review, days · LCM
Too few LCM 510(k)s in any year since 2007 to give a median.
Recalls per year · LCM
No recall names LCM since 2007.
Adverse event reports per year · LCM
2in 2019
MAUDE reports naming product code LCM, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 1 | — | 0 | 0 |
| 2010 | 1 | — | 0 | 0 |
| 2011 | 2 | — | 0 | 2 |
| 2012 | 1 | — | 0 | 1 |
| 2013 | 2 | — | 0 | 0 |
| 2014 | 1 | — | 0 | 0 |
| 2015 | 1 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 1 | — | 0 | 0 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 1 | — | 0 | 2 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 1 | — | 0 | 0 |
| 2022 | 1 | — | 0 | 0 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 0 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The RapidSense TM Drugs of Abuse Phencyclidine (PCP) 25 Device is a lateral flow competitive immunoassay intended for the qualitative detection for phencyclidine in human urine at a cut-off concentration of 25 ng/mL. The assay is intended for use in professional laboratories by healthcare professionals. For in vitro diagnostic use. This assay provides only'a preliminary result. Clinical consideration and professional judgment should be applied to any…
Read the full statement (128 words)
The RapidSense TM Drugs of Abuse Phencyclidine (PCP) 25 Device is a lateral flow competitive immunoassay intended for the qualitative detection for phencyclidine in human urine at a cut-off concentration of 25 ng/mL. The assay is intended for use in professional laboratories by healthcare professionals. For in vitro diagnostic use.
This assay provides only'a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method. Tests for PCP cannot distinguish between abused drugs and certain prescribed medications. Certain foods or medications may interfere with tests for, PCP and cause false positive results.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K080431 · printed Nov 6, 2026
Cite this page
510(k) Report. “K080431: Rapidsense Drugs of Abuse Phencyclidine (PCP) 25 Device with Models 900-0054.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K080431
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026