510(k) · Cleared
Recalled · Class II · 2020Shimadzu Mobile X-Ray System Mobiledart Evolution
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Shimadzu Corp. · IZL System, X-Ray, Mobile
Cleared on May 14, 2008 after a 64-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 7 times.
- Decided
- May 14, 2008
- Review
- 64 days
- Class
- Class II
- Product code
- IZL
- Recalls
- 7
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in IZL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IZL device
- #1 K242015 TOPAZ Mobile X-ray System (Models : TOPAZ-32D…score 19.9 · 19 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 9, 2020 | Class II | Z-0534-2021 | Device Design | There is a potential that the adjustable handle option on the X-Ray system may have been installed with improper bolts. This could result in improper driving operation which could lead to harm of individuals around the device. |
| Jul 19, 2019 | Class II | Z-2441-2019 | Software design | The graphics driver of the FDR Go PLUS might cause the appearance of a Blue Screen of Death (BSoD) |
| Jan 15, 2019 | Class II | Z-1040-2019 | Device Design | FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released. |
| Jan 14, 2019 | Class II | Z-1039-2019 | Device Design | In these units, the brake is usually released by gripping the drive handle and applied by letting go of the handle. However, for systems within a specified manufacture period, we have identified that even when letting go of the handle… |
| Nov 10, 2015 | Class II | Z-0451-2016 | Software design | The firm is recalling the Shimadzu Mobile X-ray system because an image may not transfer to image server properly. |
| Mar 18, 2013 | Class II | Z-1369-2013 | Radiation Control for Health and Safety Act | The recall was initiated because Shimadzu Corporation has identified a potential issue in certain manufacturing lots of the Shimadzu MobileDaRt Evolution in combination with CXDI-70C, CXDI-55C, and CXDI-55G. As a result of Shimadzu's… |
| Oct 28, 2011 | Class II | Z-0265-2013 | Nonconforming Material/Component | Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer… |
The 5 latest of 7 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
How does its category compare?
All IZL devicesIZL System, X-Ray, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · IZL
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IZL
153in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IZL
47in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · IZL
27in 2025
MAUDE reports naming product code IZL, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 30 | 7 | 16 |
| 2008 | 4 | 49 | 0 | 27 |
| 2009 | 3 | 125 | 0 | 35 |
| 2010 | 3 | 65 | 4 | 75 |
| 2011 | 6 | 66 | 6 | 118 |
| 2012 | 3 | 83 | 5 | 24 |
| 2013 | 0 | — | 10 | 17 |
| 2014 | 7 | 104 | 4 | 51 |
| 2015 | 4 | 88 | 4 | 30 |
| 2016 | 6 | 47 | 0 | 33 |
| 2017 | 7 | 45 | 0 | 35 |
| 2018 | 8 | 47 | 0 | 27 |
| 2019 | 9 | 33 | 2 | 40 |
| 2020 | 9 | 43 | 1 | 48 |
| 2021 | 13 | 48 | 0 | 47 |
| 2022 | 16 | 89 | 1 | 32 |
| 2023 | 3 | 55 | 6 | 77 |
| 2024 | 9 | 159 | 5 | 39 |
| 2025 | 11 | 153 | 47 | 27 |
| 2026 | 6 | 157 | 5 | — |
What is it indicated for?
This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.
Exposure can be captured on film, CR or Digital Radiography (DR)
Indications for use as stated in the FDA 510(k) statement.
https://510kreport.com/device/K080701 · printed Nov 6, 2026
Cite this page
510(k) Report. “K080701: Shimadzu Mobile X-Ray System Mobiledart Evolution.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K080701
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)