The Anspach Effort, Inc. · HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
Cleared on Apr 30, 2008 as substantially equivalent to K011444 (The Anspach Effort, Inc., 2001) after a 40-day review. Its family tree reaches back 1 generation to 2001. 14 later devices descend from it. It has been recalled 35 times.
- Decided
- Apr 30, 2008
- Review
- 40 days
- Class
- Class II
- Product code
- HBE
- Recalls
- 35
- Predicates
- 1
Its longest line of descent runs back 1 generation to K011444 (2001). 14 later devices descend from it, 2 citing it directly.
- Root · 1 generation backK011444 Anspach Emax Drill SystemThe Anspach Effort, Inc. · 2001 · Recalled
- This deviceK080802 Emax 2 Plus System, Model Emax 2 Plus Motor, Emax 2 Console>Firmware…The Anspach Effort, Inc. · 2008
- Descendant · cites it directly · 13 of its ownK113476 Anspach Dissection ToolsThe Anspach Effort, Inc. · 2011 · Recalled
- Descendant · 2 generations on · 4 of its ownK130755 Hicut Highspeed InstrumentAdeor Medical AG · 2014
- Descendant · 3 generations on · 3 of its ownK232684 Surgify HaloSurgify Medical OY · 2023
- Descendant · cites it directly · 3 of its ownK173066 OSSEODUO Shaver and Drill SystemBien-Air Surgery SA · 2018
- Descendant · 4 generations on · 2 of its ownK250380 Surgify Halo (54.085.SHD.H1); Surgify Halo (54.140.SHD.H1); Surgify…Surgify Medical OY · 2025 · Recalled
Showing 5 of 14 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 14 · See all- K113476Anspach Dissection ToolsThe Anspach Effort, Inc. · 2011 · cites it directly13descendants
- K130755Hicut Highspeed InstrumentAdeor Medical AG · 2014 · 2 generations on4descendants
- K232684Surgify HaloSurgify Medical OY · 2023 · 3 generations on3descendants
- K173066OSSEODUO Shaver and Drill SystemBien-Air Surgery SA · 2018 · cites it directly3descendants
- K250380Surgify Halo (54.085.SHD.H1); Surgify Halo (54.140.SHD.H1); Surgify Halo…Surgify Medical OY · 2025 · 4 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K011444 | Anspach Emax Drill System | The Anspach Effort, Inc. | 2001 | 70 |
| 0 (this device) | K080802 | Emax 2 Plus System, Model Emax 2 Plus Motor, Emax 2 Console>Firmware Upgrade… | The Anspach Effort, Inc. | 2008 | — |
| +1 | K113476 | Anspach Dissection Tools | The Anspach Effort, Inc. | 2011 | 39 |
| +2 | K130755 | Hicut Highspeed Instrument | Adeor Medical AG | 2014 | — |
| +3 | K232684 | Surgify Halo | Surgify Medical OY | 2023 | — |
| +1 | K173066 | OSSEODUO Shaver and Drill System | Bien-Air Surgery SA | 2018 | — |
| +4 | K250380 | Surgify Halo (54.085.SHD.H1); Surgify Halo (54.140.SHD.H1); Surgify Halo… | Surgify Medical OY | 2025 | 5 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
| 2014 | 1 | 0 |
| 2018 | 2 | 0 |
| 2019 | 3 | 0 |
| 2022 | 1 | 0 |
| 2023 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HBE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HBE device
- #1 K251433 Surgify Halo (54.085.SHD.H1); Surgify Halo…score 13.2 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K011444 | Anspach Emax Drill SystemSummary page 2…lusion The eMax 2 Plus System is substantially equivalent to the currently marketed eMax Drill System cleared by K0 11444 on August 8, 2001. The Anspach Effort, 4500 Riverside Drive, Palm Beach Gard… | 2001 |
What descends from it? (14)
See all 142 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 6, 2016 | Class II | Z-1949-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-2031-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-1948-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-2001-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-1956-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-1952-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-2026-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-2019-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-2015-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-2032-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-1955-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Oct 6, 2016 | Class II | Z-1957-2017 | Under Investigation by firm | Supplied Directions for Use (DFU) did not contain a recommended service interval. |
| Dec 17, 2013 | Class II | Z-0726-2014 | Process design | Some carbide cutting burr fractured during use. |
| Dec 17, 2013 | Class II | Z-0727-2014 | Process design | Some carbide cutting burr fractured during use. |
| Dec 17, 2013 | Class II | Z-0728-2014 | Process design | Some carbide cutting burr fractured during use. |
| Dec 17, 2013 | Class II | Z-0729-2014 | Process design | Some carbide cutting burr fractured during use. |
| Nov 28, 2012 | Class II | Z-1892-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1880-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1893-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1891-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1879-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1889-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1885-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1878-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1886-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1888-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1877-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1887-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1890-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Nov 28, 2012 | Class II | Z-1881-2014 | Process design | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements. |
| Jan 9, 2012 | Class II | Z-0074-2013 | Nonconforming Material/Component | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. |
| Jan 9, 2012 | Class II | Z-0076-2013 | Nonconforming Material/Component | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. |
| Jan 9, 2012 | Class II | Z-0075-2013 | Nonconforming Material/Component | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. |
| May 6, 2011 | Class II | Z-0200-2014 | Labeling Change Control | Anspach Effort, Inc. Palm Beach Gardens, FL initiated a correction of its medical Device labeling for its Pneumatic Systems Operating Manual, Cutting Burr Chart and Product Catalog due to a recent review of their regulations and… |
| Aug 13, 2009 | Not classified | Z-0083-2010 | Device Design | The Anspach Effort, Inc. is recalling the Foot Control/ Foot Pedal for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and accessories. The Foot Control Cables lacks the shield connection from the cable to the… |
The 5 latest of 35 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All HBE devicesHBE Drills, Burrs, Trephines & Accessories (Simple, Powered): FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · HBE
6in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HBE
30in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HBE
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HBE
1,924in 2025
MAUDE reports naming product code HBE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 0 | 14 |
| 2008 | 2 | — | 47 | 14 |
| 2009 | 0 | — | 7 | 242 |
| 2010 | 0 | — | 23 | 492 |
| 2011 | 2 | — | 1 | 1,013 |
| 2012 | 1 | — | 26 | 1,146 |
| 2013 | 1 | — | 4 | 1,583 |
| 2014 | 2 | — | 0 | 559 |
| 2015 | 3 | 155 | 0 | 483 |
| 2016 | 2 | — | 7 | 848 |
| 2017 | 2 | — | 3 | 848 |
| 2018 | 2 | — | 0 | 787 |
| 2019 | 5 | 133 | 0 | 836 |
| 2020 | 0 | — | 0 | 1,025 |
| 2021 | 6 | 193 | 1 | 1,047 |
| 2022 | 2 | — | 0 | 900 |
| 2023 | 2 | — | 3 | 948 |
| 2024 | 0 | — | 1 | 1,153 |
| 2025 | 6 | 30 | 0 | 1,924 |
| 2026 | 1 | — | 19 | — |
What is it indicated for?
The eMax 2 Plus System is intended for Cutting and shaping bone including spine and cranium.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K080802 · printed Nov 3, 2026
Cite this page
510(k) Report. “K080802: Emax 2 Plus System, Model Emax 2 Plus Motor, Emax 2 Console>Firmware Upgrade, Emax 2 FP Foot Pedal.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K080802
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026