510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Vitros Chemistry Product NA+ Slides; Vitros Chemistry Products Calibrator Kit 2

K081411510(k) SpecialCleared Jun 30, 200841-day reviewClass II21 CFR 862.1665Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho-Clinical Diagnostics, Inc. · JGS Electrode, Ion Specific, Sodium

Cleared on Jun 30, 2008 as substantially equivalent to K961099 (Johnson & Johnson Clinical Diagnostics, Inc., 1996) and 1 other predicate after a 41-day review. Its family tree reaches back 4 generations to 1999. 2 later devices descend from it. It has been recalled 4 times.

Decided
Jun 30, 2008
Review
41 days
Class
Class II
Product code
JGS
Recalls
4
Predicates
2

Its longest line of descent runs back 4 generations to K984303 (1999). 7 ancestors in all, the oldest K800054 (1980). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 2 other ancestors1980: 1 other ancestor2009: 1 descendant cleared2010: 1 descendant clearedK984303 · Vitros Chemistry Products Magnetic HDL-Cholesterol Reagent · Ortho-Clinical Diagnostics, Inc. · 1999K012593 · Vitros Chemistry Products Magnetic HDL-Cholesterol Reagent, Vitros Chemistry Products Calibrator Kit 2 · Ortho-Clinical Diagnostics, Inc. · 2001 · recalled×2: 2 generations cleared close together, K071801 (2007), K073157 (2007) 19801990200020102020
  1. Root · 4 generations backK984303 Vitros Chemistry Products Magnetic HDL-Cholesterol ReagentOrtho-Clinical Diagnostics, Inc. · 1999
  2. 3 generations backK012593 Vitros Chemistry Products Magnetic HDL-Cholesterol Reagent, Vitros…Ortho-Clinical Diagnostics, Inc. · 2001 · Recalled
  3. 2 generations backK071801 Vitros Chemistry Products CI-Slides, Calibrator Kit 2, Model (S)…Ortho-Clinical Diagnostics, Inc. · 2007 · Recalled
  4. 1 generation backK073157 Vitros Chemistry Products K+ Slides and Calibrator Kit 2Ortho-Clinical Diagnostics, Inc. · 2007 · Recalled
  5. This deviceK081411 Vitros Chemistry Product NA+ Slides; Vitros Chemistry Products…Ortho-Clinical Diagnostics, Inc. · 2008
  6. Descendant · cites it directly · 1 of its ownK091861 Vitros Chemistry Products NA + DT Slides, Model 153 2290 and DT…Ortho-Clinical Diagnostics, Inc. · 2009
  7. Descendant · 2 generations on · 0 of its ownK093611 Vitros Chemistry Products CO2 DT Slides; Vitros Chemmistry Products…Ortho-Clinical Diagnostics, Inc. · 2010

Most influential descendants

Showing 2 of 2 · See all
  • K091861Vitros Chemistry Products NA + DT Slides, Model 153 2290 and DT Calibrator Kit, Model…Ortho-Clinical Diagnostics, Inc. · 2009 · cites it directly1descendant
  • K093611Vitros Chemistry Products CO2 DT Slides; Vitros Chemmistry Products DT Calibrator Kit…Ortho-Clinical Diagnostics, Inc. · 2010 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K984303Vitros Chemistry Products Magnetic HDL-Cholesterol ReagentOrtho-Clinical Diagnostics, Inc.1999—
−3K012593Vitros Chemistry Products Magnetic HDL-Cholesterol Reagent, Vitros Chemistry…Ortho-Clinical Diagnostics, Inc.20011
−2K071801Vitros Chemistry Products CI-Slides, Calibrator Kit 2, Model (S) 844-5207…Ortho-Clinical Diagnostics, Inc.20074
−1K073157Vitros Chemistry Products K+ Slides and Calibrator Kit 2Ortho-Clinical Diagnostics, Inc.20071
0 (this device)K081411Vitros Chemistry Product NA+ Slides; Vitros Chemistry Products Calibrator Kit 2Ortho-Clinical Diagnostics, Inc.2008—
+1K091861Vitros Chemistry Products NA + DT Slides, Model 153 2290 and DT Calibrator…Ortho-Clinical Diagnostics, Inc.2009—
+2K093611Vitros Chemistry Products CO2 DT Slides; Vitros Chemmistry Products DT…Ortho-Clinical Diagnostics, Inc.2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
199620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201010
FIG. 1 — Lineage of K081411, 1999–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JGS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JGS device

  1. #1 K252475 ABL90 FLEX PLUS Systemscore 12.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K961099Ektachem Clinical Chemistry Slide Sodium (NA+)
Summary page 1
…ent slide) and VITROS Chemistry Products Calibrator Kit 2. The FDA cleared the VITROS Chemistry Products Na+ Slides on May 7, 1996 (K96 1099) under the product name EKTACHEM Clinical Chemistry Slide Sodium (Nat). With the purchase of KODAK Clinical Pr…
Johnson & Johnson Clinical Diagnostics, Inc.1996—
*K073157Vitros Chemistry Products K+ Slides and Calibrator Kit 2
Summary page 1
…emistry Products Na+ Slides. The most recent FDA clearance for the VITROS Chemistry Products Calibrator Kit 2 was December 7, 2007 (K073 157). Ortho Clinical Diagnostics, Inc. Special 510(k) VITROS Chemistry Products Na± Slides VITROS Chemistry Products Calibrator K…
Ortho-Clinical Diagnostics, Inc.20071

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K091861Vitros Chemistry Products NA + DT Slides, Model 153 2290 and DT Calibrator Kit, Model…Ortho-Clinical Diagnostics, Inc.20091

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 22, 2019Class IIZ-0992-2019Under Investigation by firmImprecision in the affected lot may produce negatively or positively biased patient sample and quality control fluid results.
Dec 20, 2016Class IIZ-1117-2017Under Investigation by firmThere is a potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ Slide cartridges warmed at room temperature and loaded onto the analyzer after using the minimum warm up requirements…
Oct 17, 2013Class IIZ-1088-2014Under Investigation by firmOrtho Clinical Diagnostics has recalled VITROS Chemistry Products Na+ Slides and the Calibrator Kit 2 due to false positive serum and plasma sample results.
Oct 17, 2013Class IIZ-1087-2014Under Investigation by firmOrtho Clinical Diagnostics has recalled VITROS Chemistry Products Na+ Slides and the Calibrator Kit 2 due to false positive serum and plasma sample results.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 1 company; a new generation was cleared every 2.6 years on median.
  • 3 of the 7 ancestors have a recall of any class.
  • Of its descendants, K091861 (Ortho-Clinical Diagnostics, Inc., 2009) has gone furthest, with 1 descendant of its own.

How does its category compare?

All JGS devices

JGS Electrode, Ion Specific, Sodium: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · JGS

3in 2023

2008: 22009: 22010: 12011: 32012: 12013: 22014: 12015: 12016: 42017: 12018: 32019: 32020: 22022: 12023: 32026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JGS

347in 2023

2011: 191 days median2016: 198 days median2018: 180 days median2019: 58 days median2023: 347 days median
2007high 347 (2023) · low 58 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JGS

4in 2023

2007: 72008: 12010: 12011: 12012: 22013: 52015: 12019: 52020: 12021: 12023: 4
2007high 7 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JGS

64in 2025

2007: 332008: 32009: 572010: 3612011: 9122012: 4122013: 5102014: 2512015: 1582016: 842017: 2922018: 1582019: 1052020: 892021: 912022: 602023: 982024: 962025: 64
2007high 912 (2011) · low 3 (2008)2025

MAUDE reports naming product code JGS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—733
20082—13
20092—057
20101—1361
201131911912
20121—2412
20132—5510
20141—0251
20151—1158
20164198084
20171—0292
201831800158
20193585105
20202—189
20210—191
20221—060
20233347498
20240—096
20250—064
20261—0—

What is it indicated for?

For in vitro diagnostic use only. VITROS Na + Slides quantitatively measure sodium (Na +) concentration in serum, plasma, and urine. Measurements used by this device are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic hormone secretion, or other disease involving electrolyte imbalance.

For in vitro diagnostic use only. VITROS Calibrator Kit 2 is used to calibrate the VITROS Chemistry Systems for the quantitative measurement of CHOL, CY, ECO2, HDLC, K +, Na +, and TRIG.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K081411: Vitros Chemistry Product NA+ Slides; Vitros Chemistry Products Calibrator Kit 2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K081411

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026