510(k) · Cleared
Recalled · Class II · 2013Loci Cardiac Troponin I Calibrator, Model RC621
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Healthcare Diagnostics · JIT Calibrator, Secondary
Cleared on Jul 7, 2008 as substantially equivalent to K053577 (Dade Behring, Inc., 2006) after a 20-day review. Its family tree reaches back 2 generations to 2001. It has been recalled 3 times.
- Decided
- Jul 7, 2008
- Review
- 20 days
- Class
- Class II
- Product code
- JIT
- Recalls
- 3
- Predicates
- 1
Its longest line of descent runs back 2 generations to K010313 (2001). 3 ancestors in all. No later device cites it as a predicate yet.
- Root · 2 generations backK010313 Dimension Ctni Flex Reagent CartridgeDade Behring, Inc. · 2001 · Recalled
- 1 generation backK053577 Dimension Vista Ctni Flex Flex Reagent Cartridge, Calibrator and…Dade Behring, Inc. · 2006
- This deviceK081683 Loci Cardiac Troponin I Calibrator, Model RC621Siemens Healthcare Diagnostics · 2008
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K010313 | Dimension Ctni Flex Reagent Cartridge | Dade Behring, Inc. | 2001 | 2 |
| −1 | K053577 | Dimension Vista Ctni Flex Flex Reagent Cartridge, Calibrator and Sdil Sample… | Dade Behring, Inc. | 2006 | — |
| 0 (this device) | K081683 | Loci Cardiac Troponin I Calibrator, Model RC621 | Siemens Healthcare Diagnostics | 2008 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2001 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JIT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K053577 | Dimension Vista Ctni Flex Flex Reagent Cartridge, Calibrator and Sdil Sample DiluentSummary page 1…e LOCI® Cardiac Troponin I Calibrator/ 21 CFR 862.1150 3. Identification of the Predicate Device Dimension Vista® Cardiac Troponin I Calibrator, K053577 4. Device Description The LOCI® Cardiac Troponin I Calibrator is a liquid, frozen, human serum based prod… | 2006 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 17, 2013 | Class II | Z-1926-2014 | Process control | Siemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shift of 24% in QC and patient results following calibration. |
| Nov 17, 2011 | Not classified | Z-0513-2012 | Nonconforming Material/Component | Siemens Healthcare Diagnostics has confirmed positive bias on QC and Patients. |
| Aug 31, 2011 | Not classified | Z-0038-2012 | Other | Internal testing has confirmed that recovery of levels 2 and 3 of LOCl Cardiac Troponin I Calibrator is lower than expected bottle value. Following calibration with LOCl Cardiac Troponin I Calibrator, lot 1AD045 an upward shift in patient… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 5.0 years on median.
- 1 of the 3 ancestors has a recall of any class.
- 1 other device also cites its predicate K053577.
How does its category compare?
All JIT devicesJIT Calibrator, Secondary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · JIT
7in 2017
2026 to Oct 25, outlined.
Median 510(k) review, days · JIT
28in 2017
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JIT
1in 2023
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JIT
7in 2025
MAUDE reports naming product code JIT, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 5 | 78 | 1 | 0 |
| 2008 | 6 | 46 | 0 | 0 |
| 2009 | 2 | — | 2 | 0 |
| 2010 | 6 | 65 | 0 | 3 |
| 2011 | 7 | 48 | 3 | 2 |
| 2012 | 2 | — | 3 | 10 |
| 2013 | 5 | 140 | 0 | 43 |
| 2014 | 10 | 33 | 2 | 9 |
| 2015 | 4 | 30 | 1 | 6 |
| 2016 | 4 | 33 | 0 | 3 |
| 2017 | 7 | 28 | 0 | 7 |
| 2018 | 0 | — | 5 | 8 |
| 2019 | 0 | — | 0 | 215 |
| 2020 | 0 | — | 2 | 12 |
| 2021 | 0 | — | 0 | 3 |
| 2022 | 0 | — | 0 | 4 |
| 2023 | 0 | — | 1 | 6 |
| 2024 | 0 | — | 0 | 17 |
| 2025 | 0 | — | 0 | 7 |
| 2026 | 0 | — | 1 | — |
What is it indicated for?
The LOCI® TNI CAL is an in vitro diagnostic product for the calibration of the Cardiac Troponin I (TNI) method on the Dimension® EXLTM integrated chemistry system with LOCI® module.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K081683 · printed Nov 4, 2026
Cite this page
510(k) Report. “K081683: Loci Cardiac Troponin I Calibrator, Model RC621.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K081683
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026