510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Loci Cardiac Troponin I Calibrator, Model RC621

K081683510(k) TraditionalCleared Jul 7, 200820-day reviewClass II21 CFR 862.1150Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Healthcare Diagnostics · JIT Calibrator, Secondary

Cleared on Jul 7, 2008 as substantially equivalent to K053577 (Dade Behring, Inc., 2006) after a 20-day review. Its family tree reaches back 2 generations to 2001. It has been recalled 3 times.

Decided
Jul 7, 2008
Review
20 days
Class
Class II
Product code
JIT
Recalls
3
Predicates
1

Its longest line of descent runs back 2 generations to K010313 (2001). 3 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2001: 1 other ancestorK010313 · Dimension Ctni Flex Reagent Cartridge · Dade Behring, Inc. · 2001 · recalledK053577 · Dimension Vista Ctni Flex Flex Reagent Cartridge, Calibrator and Sdil Sample Diluent · Dade Behring, Inc. · 2006200020022004200620082010
  1. Root · 2 generations backK010313 Dimension Ctni Flex Reagent CartridgeDade Behring, Inc. · 2001 · Recalled
  2. 1 generation backK053577 Dimension Vista Ctni Flex Flex Reagent Cartridge, Calibrator and…Dade Behring, Inc. · 2006
  3. This deviceK081683 Loci Cardiac Troponin I Calibrator, Model RC621Siemens Healthcare Diagnostics · 2008
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K010313Dimension Ctni Flex Reagent CartridgeDade Behring, Inc.20012
−1K053577Dimension Vista Ctni Flex Flex Reagent Cartridge, Calibrator and Sdil Sample…Dade Behring, Inc.2006—
0 (this device)K081683Loci Cardiac Troponin I Calibrator, Model RC621Siemens Healthcare Diagnostics2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
FIG. 1 — Lineage of K081683, 2001–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JIT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K053577Dimension Vista Ctni Flex Flex Reagent Cartridge, Calibrator and Sdil Sample Diluent
Summary page 1
…e LOCI® Cardiac Troponin I Calibrator/ 21 CFR 862.1150 3. Identification of the Predicate Device Dimension Vista® Cardiac Troponin I Calibrator, K053577 4. Device Description The LOCI® Cardiac Troponin I Calibrator is a liquid, frozen, human serum based prod…
Dade Behring, Inc.2006—

Also cleared against K053577: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 17, 2013Class IIZ-1926-2014Process controlSiemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shift of 24% in QC and patient results following calibration.
Nov 17, 2011Not classifiedZ-0513-2012Nonconforming Material/ComponentSiemens Healthcare Diagnostics has confirmed positive bias on QC and Patients.
Aug 31, 2011Not classifiedZ-0038-2012OtherInternal testing has confirmed that recovery of levels 2 and 3 of LOCl Cardiac Troponin I Calibrator is lower than expected bottle value. Following calibration with LOCl Cardiac Troponin I Calibrator, lot 1AD045 an upward shift in patient…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 5.0 years on median.
  • 1 of the 3 ancestors has a recall of any class.
  • 1 other device also cites its predicate K053577.

How does its category compare?

All JIT devices

JIT Calibrator, Secondary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · JIT

7in 2017

2007: 52008: 62009: 22010: 62011: 72012: 22013: 52014: 102015: 42016: 42017: 7
2007high 10 (2014) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JIT

28in 2017

2007: 78 days median2008: 46 days median2010: 65 days median2011: 48 days median2013: 140 days median2014: 33 days median2015: 30 days median2016: 33 days median2017: 28 days median
2007high 140 (2013) · low 28 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JIT

1in 2023

2007: 12009: 22011: 32012: 32014: 22015: 12018: 52020: 22023: 12026 (partial year): 1
2007high 5 (2018) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JIT

7in 2025

2010: 32011: 22012: 102013: 432014: 92015: 62016: 32017: 72018: 82019: 2152020: 122021: 32022: 42023: 62024: 172025: 7
2007high 215 (2019) · low 0 (2007)2025

MAUDE reports naming product code JIT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200757810
200864600
20092—20
201066503
201174832
20122—310
20135140043
2014103329
201543016
201643303
201772807
20180—58
20190—0215
20200—212
20210—03
20220—04
20230—16
20240—017
20250—07
20260—1—

What is it indicated for?

The LOCI® TNI CAL is an in vitro diagnostic product for the calibration of the Cardiac Troponin I (TNI) method on the Dimension® EXLTM integrated chemistry system with LOCI® module.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K081683: Loci Cardiac Troponin I Calibrator, Model RC621.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K081683

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026