Prelude Short Sheath Introducer

K082063510(k) TraditionalCleared Oct 15, 200886-day reviewClass II21 CFR 870.1340Recalls 19 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Merit Medical Systems, Inc. · DYB Introducer, Catheter

Cleared on Oct 15, 2008 as substantially equivalent to K961780 (Thomas Medical Products, Inc., 1996) after an 86-day review. Its family tree reaches back 1 generation to 1996. 1 of the 3 devices in its family, counting itself, has a Class I recall. 1 later device descends from it. It has been recalled 19 times, including a Class I recall.

Decided
Oct 15, 2008
Review
86 days
Class
Class II
Product code
DYB
Recalls
19 (1 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K961780 (1996). 1 later device descends from it, 1 citing it directly. 1 of the 3 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2012: 1 descendant clearedK961780 · Hemodialysis Introducer Set · Thomas Medical Products, Inc. · 19961995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant K961780 · Hemodialysis Introducer Set · Thomas Medical Products, Inc. · 19960
  1. Root · 1 generation backK961780 Hemodialysis Introducer SetThomas Medical Products, Inc. · 1996
  2. This deviceK082063 Prelude Short Sheath IntroducerMerit Medical Systems, Inc. · 2008
  3. Descendant · cites it directly · 0 of its ownK122190 Prelude 7F Short Sheath IntroducerMerit Medical Systems, Inc. · 2012

Most influential descendants

Showing 1 of 1 · See all
  • K122190Prelude 7F Short Sheath IntroducerMerit Medical Systems, Inc. · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K961780Hemodialysis Introducer SetThomas Medical Products, Inc.1996—
0 (this device)K082063Prelude Short Sheath IntroducerMerit Medical Systems, Inc.2008—
+1K122190Prelude 7F Short Sheath IntroducerMerit Medical Systems, Inc.2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
FIG. 1 — Lineage of K082063, 1996–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DYB device

  1. #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K961780Hemodialysis Introducer Set
Summary page 1
…Cardiovascular PREDICATE DEVICE(S) Hemodialysis Introducer Set (Distributed as DialEaseTM Introducer Sheath), K961780 DEVICE DESCRIPTION The Prelude® Short Sheath Introducer (PSS) consists of a sheath introducer with integral he…
Thomas Medical Products, Inc.1996—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122190Prelude 7F Short Sheath IntroducerMerit Medical Systems, Inc.20120

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 21, 2021Class IIZ-2243-2021Process controlThere is a potential that a package sheath contains an incorrect dilator.
Feb 16, 2017Class IZ-1497-2017Process controlMerit Medical Systems, Inc. announces a voluntary field action for sheaths included with the 7F Prelude¿ Short Sheath Introducers because the sheath marker tips may detach during use.
Dec 17, 2010Not classifiedZ-0941-2011Component change controlIntroducer tips may detach and embolize during use, or the tip material may elicit a thrombogenic response.
Sep 29, 2009Not classifiedZ-0518-2010OtherPotential sterility breach caused by excessive shipment damage.
Sep 29, 2009Not classifiedZ-0517-2010OtherPotential sterility breach caused by excessive shipment damage.
Apr 14, 2009Not classifiedZ-1406-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1404-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1403-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1407-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1408-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1414-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1415-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1413-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1402-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1410-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1409-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1647-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1405-2009Process controlExternal tubing may separate during use, allowing blood loss.
Apr 14, 2009Not classifiedZ-1412-2009Process controlExternal tubing may separate during use, allowing blood loss.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K122190 (Merit Medical Systems, Inc., 2012) has gone furthest, with 0 descendants of its own.

How does its category compare?

All DYB devices

DYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · DYB

16in 2025

2007: 122008: 172009: 142010: 182011: 172012: 222013: 262014: 302015: 172016: 192017: 242018: 192019: 182020: 92021: 122022: 132023: 152024: 182025: 162026 (partial year): 17
2007high 30 (2014) · low 9 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DYB

63in 2025

2007: 95 days median2008: 37 days median2009: 52 days median2010: 123 days median2011: 100 days median2012: 128 days median2013: 118 days median2014: 122 days median2015: 129 days median2016: 95 days median2017: 70 days median2018: 156 days median2019: 47 days median2020: 49 days median2021: 78 days median2022: 143 days median2023: 55 days median2024: 56 days median2025: 63 days median2026 (partial year): 59 days median
2007high 156 (2018) · low 37 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DYB

3in 2025

2007: 152008: 92009: 252010: 182011: 42012: 32013: 5, 1 Class I2014: 22015: 2, 1 Class I2016: 45, 3 Class I2017: 22, 1 Class I2018: 12019: 22020: 21, 21 Class I2021: 62022: 22023: 102024: 22025: 32026 (partial year): 23, 21 Class I
2007high 45 (2016) · low 1 (2018)2026

Class I (48 since 2007) Class II/III not classified

Adverse event reports per year · DYB

4,539in 2025

2007: 3162008: 4232009: 3992010: 4452011: 6282012: 8982013: 7432014: 7802015: 9132016: 1,2342017: 1,6722018: 1,9032019: 2,4502020: 2,4652021: 2,8352022: 4,0842023: 3,7752024: 3,9312025: 4,539
2007high 4,539 (2025) · low 316 (2007)2025

MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007129515316
200817379423
2009145225399
20101812318445
2011171004628
2012221283898
2013261185 (1)743
2014301222780
2015171292 (1)913
2016199545 (3)1,234
2017247022 (1)1,672
20181915611,903
2019184722,450
202094921 (21)2,465
2021127862,835
20221314324,084
20231555103,775
2024185623,931
2025166334,539
2026175923 (21)—

What is it indicated for?

Merit's Prelude® Short Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. This device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis. The device is not indicated for long term vascular or hemodialysis access.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082063: Prelude Short Sheath Introducer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082063

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026