510(k) · Cleared
Recalled · 2011Acuson S2000 Diagnostic Ultrasound System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medical Solutions USA, Inc., Ultrasound DI · IYN System, Imaging, Pulsed Doppler, Ultrasonic
Cleared on Nov 13, 2008 after a 106-day review. The FDA summary names its predicate without a number. It has been recalled 8 times.
- Decided
- Nov 13, 2008
- Review
- 106 days
- Class
- Class II
- Product code
- IYN
- Recalls
- 8
- Predicates
- 0
The FDA summary names its predicate without a number.
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in IYN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IYN device
- #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Acuson Antares Ultrasound System Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Mar 15, 2011 | Not classified | Z-2051-2011 | Software design | The Firm discovered a software error on the Accuson S2000 ultrasound systems affecting the measurement results in the patient report. |
| Aug 17, 2010 | Not classified | Z-1837-2011 | Software change control | There is an error in the estimation of mean pressure gradient in the Cardiac measurement package. The currently implemented calculation will result in an underestimation of the mean pressure gradient (PGmean) when estimated from a trace… |
| Oct 2, 2009 | Not classified | Z-1007-2011 | Nonconforming Material/Component | ECG waveform does not accurately align with 2D, M-Mode or Spectral Images. |
| Oct 2, 2009 | Not classified | Z-1005-2011 | Nonconforming Material/Component | ECG waveform does not accurately align with 2D, M-Mode or Spectral Images. |
| Oct 2, 2009 | Not classified | Z-1006-2011 | Nonconforming Material/Component | ECG waveform does not accurately align with 2D, M-Mode or Spectral Images. |
| Aug 28, 2009 | Not classified | Z-1392-2011 | Software design | If the transducer is not disconnected from the system after an automatic shutdown due to thermal limits being exceeded, then the next time it is used, the thermal safety guards do not take effect. Normally, the thermal limit before… |
| Apr 7, 2009 | Not classified | Z-1884-2011 | Software design | Device function failure--When performing a TEE exam with the transducer, CW (Continuous Wave) mode is not functional. Due to work flow, this may not be apparent until the transducer is inserted into the patient's esophagus. |
| Jan 14, 2009 | Not classified | Z-1900-2011 | Software change control | The product has a software problem in which previous patient measurement data gets associated with another patient's image. |
The 5 latest of 8 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
How does its category compare?
All IYN devicesIYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · IYN
57in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IYN
124in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IYN
10in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · IYN
610in 2025
MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 13 | 26 | 16 | 3 |
| 2008 | 25 | 18 | 12 | 2 |
| 2009 | 18 | 29 | 8 | 9 |
| 2010 | 29 | 67 | 10 | 17 |
| 2011 | 39 | 46 | 8 | 18 |
| 2012 | 43 | 37 | 3 | 20 |
| 2013 | 51 | 60 | 8 | 18 |
| 2014 | 56 | 39 | 2 | 48 |
| 2015 | 54 | 55 | 3 | 116 |
| 2016 | 60 | 56 | 2 | 145 |
| 2017 | 64 | 83 | 0 | 146 |
| 2018 | 59 | 85 | 7 | 85 |
| 2019 | 47 | 79 | 0 | 59 |
| 2020 | 57 | 72 | 4 | 67 |
| 2021 | 46 | 105 | 8 | 213 |
| 2022 | 53 | 105 | 10 | 186 |
| 2023 | 44 | 115 | 4 | 229 |
| 2024 | 50 | 119 | 1 | 266 |
| 2025 | 57 | 124 | 10 | 610 |
| 2026 | 44 | 107 | 2 (1) | — |
What is it indicated for?
The S2000TM ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ…
Read the full statement (95 words)
The S2000TM ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transrectal, transvaginal, peripheral vessel, musculo-skeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K082142 · printed Nov 4, 2026
Cite this page
510(k) Report. “K082142: Acuson S2000 Diagnostic Ultrasound System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K082142
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)