510(k) Report

510(k) · Cleared

Cleared

PAM 3000

K082626510(k) TraditionalCleared Jan 16, 2009129-day reviewClass II21 CFR 870.2300

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Wireless 2000 RF & Uwb Technologies, Ltd. · DRT Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm)

Cleared on Jan 16, 2009 as substantially equivalent to K070375 (Earlysense, Ltd., 2007) and 1 other predicate after a 129-day review. Its family tree reaches back 3 generations to 1983. 21 later devices descend from it.

Decided
Jan 16, 2009
Review
129 days
Class
Class II
Product code
DRT
Recalls
None
Predicates
2

Its longest line of descent runs back 3 generations to K830295 (1983). 17 ancestors in all. 21 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2003: 1 other ancestor2002: 1 other ancestor1999: 2 other ancestors2000: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor1995: 2 other ancestors1994: 1 other ancestor1992: 1 other ancestor1991: 1 other ancestor1986: 1 other ancestor2016: 1 descendant cleared2010: 1 descendant cleared2021: 2 descendants cleared2012: 1 descendant cleared2018: 2 descendants cleared2026: 5 descendants cleared2024: 2 descendants cleared2022: 2 descendants cleared2023: 1 descendant cleared2013: 1 descendant cleared2014: 1 descendant cleared2025: 1 descendant cleared2019: 1 descendant clearedK830295 · Medilog 9000 System · Oxford Medilog, Inc. · 1983K971813 · EMBLA · Ferguson Medical · 1998K070375 · Earlysense ES-16 System · Earlysense, Ltd. · 200719801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 10 other ancestors 1 generation forward: 3 descendants2 generations forward: 7 descendants3 generations forward: 7 descendants4 generations forward: 4 descendants K070375 · Earlysense ES-16 System · Earlysense, Ltd. · 2007K971813 · EMBLA · Ferguson Medical · 1998K830295 · Medilog 9000 System · Oxford Medilog, Inc. · 19830
  1. Root · 3 generations backK830295 Medilog 9000 SystemOxford Medilog, Inc. · 1983
  2. 2 generations backK971813 EMBLAFerguson Medical · 1998
  3. 1 generation backK070375 Earlysense ES-16 SystemEarlysense, Ltd. · 2007
  4. This deviceK082626 PAM 3000Wireless 2000 RF & Uwb Technologies, Ltd. · 2009
  5. Descendant · cites it directly · 10 of its ownK092062 Everon 1.0Earlysense, Ltd. · 2010
  6. Descendant · cites it directly · 9 of its ownK202464 Vital Sign Monitoring Sensor (Model :XK300)Xandar Kardian, Inc. · 2021
  7. Descendant · 2 generations on · 9 of its ownK120465 Earlysense (Everon) SystemEarlysense, Ltd. · 2012
  8. Descendant · 3 generations on · 7 of its ownK131379 EarlysenseEarlysense, Ltd. · 2013
  9. Descendant · 3 generations on · 6 of its ownK133661 Chair Sensing Unit for Earlysense SystemEarlysense, Ltd. · 2014

Showing 5 of 21 descendants. See all descendants

Descendants cleared per year

2010: 12012: 12013: 12014: 12016: 12018: 22019: 12021: 22022: 22023: 12024: 22025: 12026: 5
2010high 5 (2026) · low 0 (2011)2026

Most influential descendants

Showing 5 of 21 · See all
  • K092062Everon 1.0Earlysense, Ltd. · 2010 · cites it directly10descendants
  • K202464Vital Sign Monitoring Sensor (Model :XK300)Xandar Kardian, Inc. · 2021 · cites it directly9descendants
  • K120465Earlysense (Everon) SystemEarlysense, Ltd. · 2012 · 2 generations on9descendants
  • K131379EarlysenseEarlysense, Ltd. · 2013 · 3 generations on7descendants
  • K133661Chair Sensing Unit for Earlysense SystemEarlysense, Ltd. · 2014 · 3 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K830295Medilog 9000 SystemOxford Medilog, Inc.1983—
−2K971813EMBLAFerguson Medical1998—
−1K070375Earlysense ES-16 SystemEarlysense, Ltd.2007—
0 (this device)K082626PAM 3000Wireless 2000 RF & Uwb Technologies, Ltd.2009—
+1K092062Everon 1.0Earlysense, Ltd.2010—
+1K202464Vital Sign Monitoring Sensor (Model :XK300)Xandar Kardian, Inc.2021—
+2K120465Earlysense (Everon) SystemEarlysense, Ltd.2012—
+3K131379EarlysenseEarlysense, Ltd.2013—
+3K133661Chair Sensing Unit for Earlysense SystemEarlysense, Ltd.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199110
199210
199410
199520
199610
199710
199920
200010
200210
200310
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201210
201310
201410
201610
201820
201910
202120
202220
202310
202420
202510
202650
FIG. 1 — Lineage of K082626, 1983–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DRT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DRT device

  1. #1 K252676 The Circadia C300 System (C300)score 15.3 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K070375Earlysense ES-16 System
Summary page 1
…l. Predicate Device The Wireless 2000 PAMTM 3000 is substantially equivalent to the EarlySense ES-16 System from EarlySense Ltd (K070375) and LG1Tm Intelligent Medical Vigilance System from Hoana Medical Inc (K052446) IV. Intended Use…
Earlysense, Ltd.2007—
*K052446LG1 Intelligent Medical Vigilance System
Summary page 1
…ent to the EarlySense ES-16 System from EarlySense Ltd (K070375) and LG1Tm Intelligent Medical Vigilance System from Hoana Medical Inc (K052446) IV. Intended Use of the Device The PAMm3000 system is intended to measure heart rate and respira…
Hoana Medical, Inc.2006—

Also cleared against K070375: 1 other device →

What descends from it? (21)

See all 21

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K092062Everon 1.0Earlysense, Ltd.201010
K202464Vital Sign Monitoring Sensor (Model :XK300)Xandar Kardian, Inc.20219
K152911EarlySense Insight SytemEarlysense, Ltd.20165

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 4 companies; a new generation was cleared every 9.8 years on median.
  • 1 of the 17 ancestors has a recall of any class.
  • Of its descendants, K092062 (Earlysense, Ltd., 2010) has gone furthest, with 10 descendants of its own.
  • 1 other device also cites its predicate K070375.

How does its category compare?

All DRT devices

DRT Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm): FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DRT

1in 2025

2007: 12008: 12009: 62010: 32011: 32012: 12013: 12015: 12016: 12017: 12018: 12019: 12021: 22022: 12023: 12024: 32025: 12026 (partial year): 5
2007high 6 (2009) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DRT

228in 2024

2009: 70 days median2010: 44 days median2011: 149 days median2024: 228 days median2026 (partial year): 120 days median
2007high 228 (2024) · low 44 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DRT

1in 2024

2011: 12021: 12024: 1
2007high 1 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DRT

40in 2025

2007: 1132008: 992009: 1182010: 632011: 362012: 472013: 162014: 262015: 482016: 552017: 1592018: 2242019: 4472020: 502021: 852022: 352023: 472024: 292025: 40
2007high 447 (2019) · low 16 (2013)2025

MAUDE reports naming product code DRT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—0113
20081—099
20096700118
2010344063
20113149136
20121—047
20131—016
20140—026
20151—048
20161—055
20171—0159
20181—0224
20191—0447
20200—050
20212—185
20221—035
20231—047
20243228129
20251—040
202651200—

What is it indicated for?

The PAAvTm3000 system is intended to measure heart rate and respiration rate m adult patients, m a general care hospital environment including, nursing homes The system momtors presence or absence of a patient in bed (bed exit) The system also times the length of continuous patient motion or absence of patient motion

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082626: PAM 3000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082626

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026