Elecsys Troponin I and Troponin I Stat Test Systems

K082699510(k) TraditionalCleared Aug 19, 2009338-day reviewClass II21 CFR 862.1215Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics · MMI Immunoassay Method, Troponin Subunit

Cleared on Aug 19, 2009 as substantially equivalent to K021814 (Beckman Coulter, Inc., 2002) and 1 other predicate after a 338-day review. Its family tree reaches back 6 generations to 1993. 9 later devices descend from it. It has been recalled 4 times.

Decided
Aug 19, 2009
Review
338 days
Class
Class II
Product code
MMI
Recalls
4
Predicates
2

Its longest line of descent runs back 6 generations to K931172 (1993). 13 ancestors in all. 9 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 2 other ancestors2005: 1 other ancestor2001: 1 other ancestor2003: 1 other ancestor2000: 1 other ancestor1996: 1 other ancestor2017: 1 descendant cleared2018: 2 descendants cleared2021: 1 descendant cleared2019: 3 descendants cleared2022: 1 descendant cleared2023: 1 descendant clearedK931172 · Imx (R) Stat CK-MB · Abbott Laboratories · 1993×2: 2 generations cleared close together, K961501 (1996), K962575 (1996) K983492 · Boehringer Mannheim Elecsys Presicontrol Cardiac · Boehringer Mannheim Corp. · 1998K032089 · Roche Diagnostics Elecsys Precicontrol Cardiac · Roche Diagnostics Corp. · 2003K072437 · Elecsys Probnp II, Precicontrol Cardiac II, Probnp II Calset · Roche Diagnostics · 20081990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 2 other ancestors5 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 6 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant K072437 · Elecsys Probnp II, Precicontrol Cardiac II, Probnp II Calset · Roche Diagnostics · 2008K032089 · Roche Diagnostics Elecsys Precicontrol Cardiac · Roche Diagnostics Corp. · 2003K983492 · Boehringer Mannheim Elecsys Presicontrol Cardiac · Boehringer Mannheim Corp. · 1998K962575 · Elecsys Precicontrol Cardiac · Boehringer Mannheim Corp. · 1996K961501 · Elecsys CK-MB · Boehringer Mannheim Corp. · 1996K931172 · Imx (R) Stat CK-MB · Abbott Laboratories · 1993−50
  1. Root · 6 generations backK931172 Imx (R) Stat CK-MBAbbott Laboratories · 1993
  2. 5 generations backK961501 Elecsys CK-MBBoehringer Mannheim Corp. · 1996
  3. 4 generations backK962575 Elecsys Precicontrol CardiacBoehringer Mannheim Corp. · 1996
  4. 3 generations backK983492 Boehringer Mannheim Elecsys Presicontrol CardiacBoehringer Mannheim Corp. · 1998
  5. 2 generations backK032089 Roche Diagnostics Elecsys Precicontrol CardiacRoche Diagnostics Corp. · 2003
  6. 1 generation backK072437 Elecsys Probnp II, Precicontrol Cardiac II, Probnp II CalsetRoche Diagnostics · 2008
  7. This deviceK082699 Elecsys Troponin I and Troponin I Stat Test SystemsRoche Diagnostics · 2009
  8. Descendant · cites it directly · 8 of its ownK162895 Elecsys Troponin T Gen 5 STAT Assay, Elecsys Troponin T Gen 5 STAT…Roche Diagnostics · 2017
  9. Descendant · 2 generations on · 2 of its ownK191595 ARCHITECT STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division · 2019
  10. Descendant · 3 generations on · 1 of its ownK202525 Alinity i STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division · 2022
  11. Descendant · 4 generations on · 0 of its ownK230994 Alinity i STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division · 2023
  12. Descendant · 2 generations on · 0 of its ownK201441 Elecsys Troponin T Gen 5Roche Diagnostics · 2021

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2017: 12018: 22019: 32021: 12022: 12023: 1
2017high 3 (2019) · low 0 (2020)2026

Most influential descendants

Showing 5 of 9 · See all
  • K162895Elecsys Troponin T Gen 5 STAT Assay, Elecsys Troponin T Gen 5 STAT CalSet, Elecsys…Roche Diagnostics · 2017 · cites it directly8descendants
  • K191595ARCHITECT STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division · 2019 · 2 generations on2descendants
  • K202525Alinity i STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division · 2022 · 3 generations on1descendant
  • K230994Alinity i STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division · 2023 · 4 generations on0descendants
  • K201441Elecsys Troponin T Gen 5Roche Diagnostics · 2021 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K931172Imx (R) Stat CK-MBAbbott Laboratories1993—
−5K961501Elecsys CK-MBBoehringer Mannheim Corp.1996—
−4K962575Elecsys Precicontrol CardiacBoehringer Mannheim Corp.1996—
−3K983492Boehringer Mannheim Elecsys Presicontrol CardiacBoehringer Mannheim Corp.1998—
−2K032089Roche Diagnostics Elecsys Precicontrol CardiacRoche Diagnostics Corp.2003—
−1K072437Elecsys Probnp II, Precicontrol Cardiac II, Probnp II CalsetRoche Diagnostics2008—
0 (this device)K082699Elecsys Troponin I and Troponin I Stat Test SystemsRoche Diagnostics2009—
+1K162895Elecsys Troponin T Gen 5 STAT Assay, Elecsys Troponin T Gen 5 STAT CalSet…Roche Diagnostics2017—
+2K191595ARCHITECT STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division2019—
+3K202525Alinity i STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division2022—
+4K230994Alinity i STAT High Sensitivity Troponin-IAbbott Laboratories Diagnostics Division2023—
+2K201441Elecsys Troponin T Gen 5Roche Diagnostics2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200010
200110
200220
200310
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
201820
201930
202110
202210
202310
FIG. 1 — Lineage of K082699, 1993–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MMI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MMI device

  1. #1 K253051 i-STAT hs-TnI cartridge with the i-STAT…score 13.0 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K021814Accutni and Accutni Calibrators on the Access Immunoassay Systems, Model #33340, 33345
Summary page 4
…1.) Elecsys Troponin I and Troponin I STAT Immunoassays are equivalent to the Beckman Coulter Access AccuTnI Immunoassay (K021814). 2.) Elecsys PreciControl Troponi.n is equivalent to Elecsys PreciControl Cardiac II (K07…
Beckman Coulter, Inc.20023
*K072437Elecsys Probnp II, Precicontrol Cardiac II, Probnp II Calset
Summary page 4
…Immunoassay (K021814). 2.) Elecsys PreciControl Troponi.n is equivalent to Elecsys PreciControl Cardiac II (K072437). 3.) Elecsys Troponin I CalSet and Troponin I STAT CalSet are equivalent to Elecsys proBN…
Roche Diagnostics2008—

Also cleared against K021814: 2 other devices →

What descends from it? (9)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162895Elecsys Troponin T Gen 5 STAT Assay, Elecsys Troponin T Gen 5 STAT CalSet, Elecsys…Roche Diagnostics20178

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 7, 2021Class IIZ-1596-2021Device DesignRecall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When…
Apr 7, 2021Class IIZ-1595-2021Device DesignRecall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When…
Mar 12, 2012Not classifiedZ-1388-2012Nonconforming Material/ComponentAn investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of the serum…
Mar 12, 2012Not classifiedZ-1387-2012Nonconforming Material/ComponentAn investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of the serum…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 3.0 years on median.
  • 2 of the 13 ancestors have a recall of any class.
  • Of its descendants, K162895 (Roche Diagnostics, 2017) has gone furthest, with 8 descendants of its own.
  • 2 other devices also cite its predicate K021814.

How does its category compare?

All MMI devices

MMI Immunoassay Method, Troponin Subunit: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · MMI

4in 2025

2007: 32008: 12009: 12010: 22011: 12013: 22017: 12018: 42019: 32021: 12022: 12023: 32024: 22025: 42026 (partial year): 1
2007high 4 (2018) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MMI

266in 2025

2007: 190 days median2018: 344 days median2019: 88 days median2023: 270 days median2025: 266 days median
2007high 344 (2018) · low 88 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MMI

1in 2024

2007: 62008: 22009: 32010: 52011: 12012: 62014: 12015: 12019: 12020: 12021: 12024: 1
2007high 6 (2007) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MMI

805in 2025

2007: 3522008: 8372009: 1,0102010: 1,3352011: 2,4062012: 1,8412013: 1,3942014: 1,1502015: 8312016: 4872017: 6182018: 4872019: 5212020: 3872021: 3562022: 5072023: 7482024: 9212025: 805
2007high 2,406 (2011) · low 352 (2007)2025

MAUDE reports naming product code MMI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731906352
20081—2837
20091—31,010
20102—51,335
20111—12,406
20120—61,841
20132—01,394
20140—11,150
20150—1831
20160—0487
20171—0618
201843440487
20193881521
20200—1387
20211—1356
20221—0507
202332700748
20242—1921
202542660805
20261—0—

What is it indicated for?

Immunoassay for the in vitro quantitative determination of cardiac troponin I in human serum and plasma. The Elecsys Troponin i assay is intended to aid in the diagnosis of myocardial infarction. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Elecsys and MODULAR Analytics E170 immunoassay analyzers. Immunoassay for the in vitro quantitative determination of cardiac troponin I in human serum and plasma. The Elecsys Troponin I…

Read the full statement (161 words)

Immunoassay for the in vitro quantitative determination of cardiac troponin I in human serum and plasma. The Elecsys Troponin i assay is intended to aid in the diagnosis of myocardial infarction.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Elecsys and MODULAR Analytics E170 immunoassay analyzers.

Immunoassay for the in vitro quantitative determination of cardiac troponin I in human serum and plasma. The Elecsys Troponin I STAT assay is intended to aid in the diagnosis of myocardial infarction.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Elecsys analyzers.

Elecsys PreciControl Troponin is used for quality control of the Elecsys Troponin-I and Elecsys Troponin I STAT immunoassays on the Elecsys and MODULAR Analytics El170 Analyzers.

Elecsys Troponin I CalSet is used for calibrating the quantitative Elecsys Troponin I assay on the Elecsys and MODULAR Analytics El 70 analyzers.

Elecsys Troponin I STAT CalSet is used for calibrating the quantitative Elecsys Troponin I STAT assay on the Elecsys analyzers.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K082699: Elecsys Troponin I and Troponin I Stat Test Systems.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082699

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026