510(k) Report

510(k) · Cleared

Cleared

Bionexia Opticup Multi-Strip Doa Dipstick Test Integrated Cup

K082977510(k) SpecialCleared Aug 20, 2009318-day reviewClass II21 CFR 862.3870

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Mk Bio, Inc. · LDJ Enzyme Immunoassay, Cannabinoids

Cleared on Aug 20, 2009 as substantially equivalent to K061005 (Applied Dna Technologies, Inc., 2006) after a 318-day review. Its family tree reaches back 4 generations to 1995.

Decided
Aug 20, 2009
Review
318 days
Class
Class II
Product code
LDJ
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K945610 (1995). 19 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK945610 Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM)…Princeton BioMeditech Corp. · 1995
  2. 3 generations backK011673 Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step…ACON Laboratories, Inc. · 2001
  3. 2 generations backK020313 Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step…Acon Laboratories Co. · 2002
  4. 1 generation backK061005 Accustep Doa Single and Multi-Strip Screen PanelsApplied Dna Technologies, Inc. · 2006
  5. This deviceK082977 Bionexia Opticup Multi-Strip Doa Dipstick Test Integrated CupMk Bio, Inc. · 2009
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K945610Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-AmpPrinceton BioMeditech Corp.1995—
−3K011673Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test…ACON Laboratories, Inc.2001—
−2K020313Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug…Acon Laboratories Co.2002—
−1K061005Accustep Doa Single and Multi-Strip Screen PanelsApplied Dna Technologies, Inc.2006—
0 (this device)K082977Bionexia Opticup Multi-Strip Doa Dipstick Test Integrated CupMk Bio, Inc.2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199620
199810
199910
200010
200170
200220
200310
FIG. 1 — Lineage of K082977, 1995–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LDJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LDJ device

  1. #1 K232732 Rapid Marijuana (THC) Test Strip 20, Rapid…score 9.4 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061005Accustep Doa Single and Multi-Strip Screen Panels
Summary page 3
…e indicated. Description: Bionexia OptiCup is derived from a previously FDA-cleared device, ADT's ACCUSTEP Single and Multi-Strip Cassette/Dipstick DOA Screen Panels (K061005). The unmodified test is made of various number of single drug-of-abuse (DOA) test strips (from 1 to 10 strips) placed in an ope…
Applied Dna Technologies, Inc.2006—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 5 companies; a new generation was cleared every 4.4 years on median.
  • None of the 19 ancestors has a recall.

How does its category compare?

All LDJ devices

LDJ Enzyme Immunoassay, Cannabinoids: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · LDJ

2in 2024

2008: 22009: 12011: 42012: 42013: 32014: 22015: 32016: 12019: 12023: 12024: 2
2007high 4 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LDJ

48in 2015

2011: 219 days median2012: 193 days median2013: 162 days median2015: 48 days median
2007high 219 (2011) · low 48 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LDJ

1in 2023

2008: 42023: 1
2007high 4 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LDJ

1in 2015

2015: 1
2007high 1 (2015) · low 0 (2007)2025

MAUDE reports naming product code LDJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—40
20091—00
20100—00
2011421900
2012419300
2013316200
20142—00
201534801
20161—00
20170—00
20180—00
20191—00
20200—00
20210—00
20220—00
20231—10
20242—00
20250—00
20260—0—

What is it indicated for?

The Bionexia OptiCup is a rapid chromatographic immuneassay for the qualitative and simultaneous detection of elevated levels of Marijuana, Cocaine, Phencyclidine, Morphine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline in human urine at or above the following concentrations: Marijuana 50 ng/mL Cocaine 300 ng/mL Phencyclidine 25 ng/mL Morphine 2,000 ng/mL Methamphetamine 1,000 ng/mL Methadone 300 ng/mL…

Read the full statement (92 words)

The Bionexia OptiCup is a rapid chromatographic immuneassay for the qualitative and simultaneous detection of elevated levels of Marijuana, Cocaine, Phencyclidine, Morphine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline in human urine at or above the following concentrations:

Marijuana 50 ng/mL

Cocaine 300 ng/mL

Phencyclidine 25 ng/mL

Morphine 2,000 ng/mL

Methamphetamine 1,000 ng/mL

Methadone 300 ng/mL

Amphetamine 1,000 ng/mL

Barbiturates 300 ng/mL

Benzodiazepines 300 ng/mL

Nortriptyline 1,000 ng/mL

The test strips with an integrated cup is intended for healthcare professionals only. It is not intended for over the counter sale to non-professionals.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082977: Bionexia Opticup Multi-Strip Doa Dipstick Test Integrated Cup.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082977

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026