510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

X-Coated Terumo and Sarns Dual Stage Venous Return Catheters, Sarns Malleable Venous and Dual Stage Return Catheters

K083301510(k) SpecialCleared Dec 18, 200838-day reviewClass II21 CFR 870.4210Recalls 37

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Terumo Cardiovascular Systems Corp. · DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Cleared on Dec 18, 2008 as substantially equivalent to K920902 (3M Health Care, Sarns, 1993) and 2 other predicates after a 38-day review. Its family tree reaches back 1 generation to 1977. 3 later devices descend from it. It has been recalled 37 times.

Decided
Dec 18, 2008
Review
38 days
Class
Class II
Product code
DWF
Recalls
37
Predicates
3

Its longest line of descent runs back 1 generation to K770429 (1977). 3 ancestors in all. 3 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1993: 1 other ancestor1991: 1 other ancestor2014: 3 descendants clearedK770429 · Cannulae, Aortic, Arch, High Flow · 3M Health Care, Sarns · 1977 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 3 descendants K770429 · Cannulae, Aortic, Arch, High Flow · 3M Health Care, Sarns · 1977 · recalled0
  1. Root · 1 generation backK770429 Cannulae, Aortic, Arch, High Flow3M Health Care, Sarns · 1977 · Recalled
  2. This deviceK083301 X-Coated Terumo and Sarns Dual Stage Venous Return Catheters, Sarns…Terumo Cardiovascular Systems Corp. · 2008
  3. Descendant · cites it directly · 0 of its ownK132811 Sarns Soft-Flow Extended Aortic CannulaTerumo Cardiovascular Systems Corporation · 2014
  4. Descendant · cites it directly · 0 of its ownK132451 Sarns Soft-Flow Aortic CannulaTerumo Cardiovascular Systems Corp. · 2014
  5. Descendant · cites it directly · 0 of its ownK133151 Sarns High Flow Aortic Arch CannulaTerumo Cardiovascular Systems Corp. · 2014

Most influential descendants

Showing 3 of 3 · See all
  • K132811Sarns Soft-Flow Extended Aortic CannulaTerumo Cardiovascular Systems Corporation · 2014 · cites it directly0descendants
  • K132451Sarns Soft-Flow Aortic CannulaTerumo Cardiovascular Systems Corp. · 2014 · cites it directly0descendants
  • K133151Sarns High Flow Aortic Arch CannulaTerumo Cardiovascular Systems Corp. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K770429Cannulae, Aortic, Arch, High Flow3M Health Care, Sarns197712
0 (this device)K083301X-Coated Terumo and Sarns Dual Stage Venous Return Catheters, Sarns Malleable…Terumo Cardiovascular Systems Corp.2008—
+1K132811Sarns Soft-Flow Extended Aortic CannulaTerumo Cardiovascular Systems Corporation2014—
+1K132451Sarns Soft-Flow Aortic CannulaTerumo Cardiovascular Systems Corp.2014—
+1K133151Sarns High Flow Aortic Arch CannulaTerumo Cardiovascular Systems Corp.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201430
FIG. 1 — Lineage of K083301, 1977–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWF device

  1. #1 K251258 MC2™ Two-Stage Venous Cannula; MC2X™…score 14.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K920902Modification of Sarns Two Stage Venous Cannula
Summary page 2
…Return Cannula SarnsTM Dual Stage X-coated SarnsTM Venous Return K920902 1993-05-11 Dual Stage Venous Catheters…
3M Health Care, Sarns199310
*K905224Cardiopulmonary Bypass Vascular Catheter
Summary page 2
…Catheters SarnsTM Malleable X-coated SarnsTM Venous Return K905224 1991-02-05 Malleable Venous Catheters…
3M Health Care, Sarns19912
*K770429Cannulae, Aortic, Arch, High Flow
Summary page 2
…Return Catheters SarnsTM High Flow X-coated SarnsTM Aortic Arch Cannula K770429 1977-03-16 High Flow Aortic…
3M Health Care, Sarns197712
*—Terumo® Dual-Stage Venous Return Cannula
Named without a submission number
—
*—SarnsTM Malleable Dual Stage Venous Return Catheters
Named without a submission number
—
*—SarnsTM Soft-FlowTM Extended Aortic Cannula
Named without a submission number
—

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132811Sarns Soft-Flow Extended Aortic CannulaTerumo Cardiovascular Systems Corporation20140
K132451Sarns Soft-Flow Aortic CannulaTerumo Cardiovascular Systems Corp.20140
K133151Sarns High Flow Aortic Arch CannulaTerumo Cardiovascular Systems Corp.20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 23, 2014Class IIZ-2097-2014Process controlDuring an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications
May 23, 2014Class IIZ-2096-2014Process controlDuring an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications
Jan 7, 2013Class IIZ-0667-2013Process controlFlexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in…
Jan 7, 2013Class IIZ-0668-2013Process controlFlexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in…
Jan 7, 2013Class IIZ-0670-2013Process controlFlexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in…
Jan 7, 2013Class IIZ-0671-2013Process controlFlexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in…
Jan 7, 2013Class IIZ-0666-2013Process controlFlexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in…
Jan 7, 2013Class IIZ-0669-2013Process controlFlexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in…
May 9, 2012Class IIZ-1834-2012Nonconforming Material/ComponentDuring production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found…
Feb 23, 2012Not classifiedZ-1294-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1297-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1313-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1302-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1310-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1299-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1309-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1300-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1305-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1301-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1307-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1293-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1298-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1292-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1306-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1296-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1287-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1291-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1288-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1304-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1290-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1311-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1312-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1295-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1303-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1314-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1308-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
Feb 23, 2012Not classifiedZ-1289-2012Nonconforming Material/ComponentDuring in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K132811 (Terumo Cardiovascular Systems Corporation, 2014) has gone furthest, with 0 descendants of its own.

How does its category compare?

All DWF devices

DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · DWF

4in 2025

2007: 22008: 62009: 82010: 32011: 52012: 102013: 92014: 142015: 92016: 32017: 52018: 52019: 32020: 32021: 52022: 22023: 22024: 32025: 42026 (partial year): 8
2007high 14 (2014) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWF

89in 2025

2008: 95 days median2009: 148 days median2010: 62 days median2011: 85 days median2012: 110 days median2013: 110 days median2014: 110 days median2015: 46 days median2016: 32 days median2017: 127 days median2018: 30 days median2019: 120 days median2020: 30 days median2021: 358 days median2024: 29 days median2025: 89 days median2026 (partial year): 81 days median
2007high 358 (2021) · low 29 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DWF

4in 2024

2007: 252008: 252009: 212010: 262011: 132012: 372013: 14, 1 Class I2014: 62015: 3, 1 Class I2016: 32017: 2, 1 Class I2018: 102019: 22021: 32024: 42026 (partial year): 20
2007high 37 (2012) · low 0 (2020)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DWF

1,020in 2025

2007: 1212008: 1262009: 992010: 1972011: 2122012: 2332013: 2382014: 3252015: 2182016: 1442017: 1012018: 1182019: 1792020: 1422021: 1622022: 2762023: 4702024: 4012025: 1,020
2007high 1,020 (2025) · low 99 (2009)2025

MAUDE reports naming product code DWF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—25121
200869525126
200981482199
201036226197
201158513212
20121011037233
2013911014 (1)238
2014141106325
20159463 (1)218
20163323144
201751272 (1)101
201853010118
201931202179
20203300142
202153583162
20222—0276
20232—0470
20243294401
202548901,020
202688120—

What is it indicated for?

The X-coating is intended to reduce the adhesion of platelets to the blood contacting surfaces of the device during cardiopulmonary bypass procedures. The X-coated Terumo® Dual-Stage Venous Return Cannula is indicated for use for adult single tube blood drainage from the right atrium and inferior vena cava (IVC) and delivery of the blood to the extracorporeal circuit during cardiopulmonary bypass. The X-coated SarnsTM Dual-Stage Venous Return Catheters…

Read the full statement (210 words)

The X-coating is intended to reduce the adhesion of platelets to the blood contacting surfaces of the device during cardiopulmonary bypass procedures.

The X-coated Terumo® Dual-Stage Venous Return Cannula is indicated for use for adult single tube blood drainage from the right atrium and inferior vena cava (IVC) and delivery of the blood to the extracorporeal circuit during cardiopulmonary bypass.

The X-coated SarnsTM Dual-Stage Venous Return Catheters are indicated for single tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery.

The SarnsTM Malleable Venous Return Catheters are indicated for venous drainage during cardiopulmonary bypass surgery: the 24, 28, 30, 32, 34, 36 and 40 Fr. are for dual cannulation of the superior and inferior vena cava; the 40 Fr. is-for single cannulation of the right atrium.

The X-coated SarnsTM Malleable Dual-Stage Venous Return Catheters are indicated for single-tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery.

The X-coated SarnsTM High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.

The X-coated SarnsTM Soft-FlowTM Extended Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.

These devices are indicated for up to 6 hours of use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K083301: X-Coated Terumo and Sarns Dual Stage Venous Return Catheters, Sarns Malleable Venous and Dual Stage Return Catheters.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K083301

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026