510(k) Report

510(k) · Cleared

Recalled · Class I · 2022

Dialog with Adimea Option

K083460510(k) TraditionalCleared Feb 18, 2010454-day reviewClass II21 CFR 876.5860Recalls 8 (6 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

B. Braun Avitum AG · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Feb 18, 2010 after a 454-day review. The FDA summary names its predicate without a number. 1 of the 4 devices in its family, counting itself, has a Class I recall. 3 later devices descend from it. It has been recalled 8 times, including 6 Class I recalls.

Decided
Feb 18, 2010
Review
454 days
Class
Class II
Product code
KDI
Recalls
8 (6 Class I)
Predicates
0

3 later devices descend from it, 2 citing it directly. 1 of the 4 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2016: 1 descendant cleared2019: 1 descendant cleared2025: 1 descendant cleared2010201520202025
  1. This deviceK083460 Dialog with Adimea OptionB. Braun Avitum AG · 2010
  2. Descendant · cites it directly · 1 of its ownK152938 DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD. · 2016 · Recalled
  3. Descendant · 2 generations on · 0 of its ownK242155 DBB-06 PRO Hemodialysis Delivery System (DBB-06 PRO)Nikkiso CO., LTD. · 2025
  4. Descendant · cites it directly · 0 of its ownK182940 Surdial DX Hemodialysis SystemNipro Medical Corporation · 2019 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K152938DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD. · 2016 · cites it directly1descendant
  • K242155DBB-06 PRO Hemodialysis Delivery System (DBB-06 PRO)Nikkiso CO., LTD. · 2025 · 2 generations on0descendants
  • K182940Surdial DX Hemodialysis SystemNipro Medical Corporation · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K083460Dialog with Adimea OptionB. Braun Avitum AG2010—
+1K152938DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD.20161
+2K242155DBB-06 PRO Hemodialysis Delivery System (DBB-06 PRO)Nikkiso CO., LTD.2025—
+1K182940Surdial DX Hemodialysis SystemNipro Medical Corporation20191

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201910
202510
FIG. 1 — Lineage of K083460, 2010–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Dialog+® Hemodialysis System
Named without a submission number
—

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K152938DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD.20161
K182940Surdial DX Hemodialysis SystemNipro Medical Corporation20190

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 28, 2022Class IIZ-0873-2022Process controlFirm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on…
Jan 28, 2022Class IIZ-0872-2022Process controlFirm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on…
Apr 1, 2016Class IZ-1531-2016Process controlPotential leakage of conductivity sensors of the Dialog Dialysis machines.
Apr 1, 2016Class IZ-1534-2016Process controlPotential leakage of conductivity sensors of the Dialog Dialysis machines.
Apr 1, 2016Class IZ-1532-2016Process controlPotential leakage of conductivity sensors of the Dialog Dialysis machines.
Apr 1, 2016Class IZ-1535-2016Process controlPotential leakage of conductivity sensors of the Dialog Dialysis machines.
Apr 1, 2016Class IZ-1533-2016Process controlPotential leakage of conductivity sensors of the Dialog Dialysis machines.
Apr 1, 2016Class IZ-1536-2016Process controlPotential leakage of conductivity sensors of the Dialog Dialysis machines.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K152938 (Nikkiso CO., LTD., 2016) has gone furthest, with 1 descendant of its own.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

This dialysis machine can be used for implementing and monitoring hemodialysis treatments for patients with acute or chronic kidney failure. The system can be used in hospital, health center and outpatient dialysis center settings when prescribed by a physician.

The following types ‘of renal replacement therapy can be carried out with the Dialog+ System:

Hemodialysis (HD) with or without phases of pure ultrafiltration —

* High flux hemodialysis

* Low flux hemodialysis

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K083460: Dialog with Adimea Option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K083460

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026