K083558510(k) TraditionalCleared Mar 6, 200994-day reviewClass II21 CFR 888.3353Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medacta International S.A. · LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Cleared on Mar 6, 2009 as substantially equivalent to K072857 (Medacta International S.A., 2008) and 2 other predicates after a 94-day review. Its family tree reaches back 4 generations to 1994. 9 later devices descend from it. It has been recalled once.

Decided
Mar 6, 2009
Review
94 days
Class
Class II
Product code
LZO
Recalls
1
Predicates
3

Its longest line of descent runs back 4 generations to K932481 (1994). 8 ancestors in all. 9 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor1999: 1 other ancestor2007: 1 other ancestor2005: 1 other ancestor2020: 4 descendants cleared2010: 1 descendant cleared2022: 1 descendant cleared2021: 1 descendant cleared2025: 2 descendants clearedK932481 · SL-Plus and SLR-Plus · Encore Orthopedics, Inc. · 1994K001942 · SL-Plus and SLR-Plus Stems · Plus Orthopedics · 2000K021178 · SL-Plus Lateralized Stem · Plus Orthopedics · 2002K072857 · Total Hip Prosthesis System Quadra S + Cocrmo Femoral Ball Head + Apricot · Medacta International S.A. · 2008 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 6 descendants K072857 · Total Hip Prosthesis System Quadra S + Cocrmo Femoral Ball Head + Apricot · Medacta International S.A. · 2008 · recalledK021178 · SL-Plus Lateralized Stem · Plus Orthopedics · 2002K001942 · SL-Plus and SLR-Plus Stems · Plus Orthopedics · 2000K932481 · SL-Plus and SLR-Plus · Encore Orthopedics, Inc. · 19940
  1. Root · 4 generations backK932481 SL-Plus and SLR-PlusEncore Orthopedics, Inc. · 1994
  2. 3 generations backK001942 SL-Plus and SLR-Plus StemsPlus Orthopedics · 2000
  3. 2 generations backK021178 SL-Plus Lateralized StemPlus Orthopedics · 2002
  4. 1 generation backK072857 Total Hip Prosthesis System Quadra S + Cocrmo Femoral Ball Head +…Medacta International S.A. · 2008 · Recalled
  5. This deviceK083558 Medacta Total Hip Prosthesis System - Quadra CMedacta International S.A. · 2009
  6. Descendant · cites it directly · 6 of its ownK192827 Quadra-PMedacta International S.A. · 2020
  7. Descendant · 2 generations on · 1 of its ownK202730 Quadra-PMedacta International S.A. · 2020
  8. Descendant · 2 generations on · 1 of its ownK203041 SMS Cementless StemMedacta International S.A. · 2020
  9. Descendant · cites it directly · 1 of its ownK103189 Medacta Total Hip Prosthesis System - Amistem C & Quadra C Short NeckMedacta International S.A. · 2010 · Recalled
  10. Descendant · 2 generations on · 0 of its ownK250450 Coated hip implantsMedacta International S.A. · 2025

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2010: 12020: 42021: 12022: 12025: 2
2010high 4 (2020) · low 0 (2011)2026

Most influential descendants

Showing 5 of 9 · See all
  • K192827Quadra-PMedacta International S.A. · 2020 · cites it directly6descendants
  • K202730Quadra-PMedacta International S.A. · 2020 · 2 generations on1descendant
  • K203041SMS Cementless StemMedacta International S.A. · 2020 · 2 generations on1descendant
  • K103189Medacta Total Hip Prosthesis System - Amistem C & Quadra C Short NeckMedacta International S.A. · 2010 · cites it directly1descendant
  • K250450Coated hip implantsMedacta International S.A. · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K932481SL-Plus and SLR-PlusEncore Orthopedics, Inc.1994—
−3K001942SL-Plus and SLR-Plus StemsPlus Orthopedics2000—
−2K021178SL-Plus Lateralized StemPlus Orthopedics2002—
−1K072857Total Hip Prosthesis System Quadra S + Cocrmo Femoral Ball Head + ApricotMedacta International S.A.20081
0 (this device)K083558Medacta Total Hip Prosthesis System - Quadra CMedacta International S.A.2009—
+1K192827Quadra-PMedacta International S.A.2020—
+2K202730Quadra-PMedacta International S.A.2020—
+2K203041SMS Cementless StemMedacta International S.A.2020—
+1K103189Medacta Total Hip Prosthesis System - Amistem C & Quadra C Short NeckMedacta International S.A.20101
+2K250450Coated hip implantsMedacta International S.A.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199910
200210
200510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
202040
202110
202210
202520
FIG. 1 — Lineage of K083558, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZO device

  1. #1 K192827 Quadra-Pscore 19.3 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072857Total Hip Prosthesis System Quadra S + Cocrmo Femoral Ball Head + Apricot
Summary page 1
…s 11 Device Product Code: LZO, JDI Predicate Devices: Medata Total Hip Prosthesis - Quadra S femoral stems, K072857, cleared on Feb. 4, 2008 Sulzer Orthopedics MS-30 Femoral Stem, K993043, cleared…
Medacta International S.A.20081
*K993043MS-30 Femoral Stem
Summary page 1
…is - Quadra S femoral stems, K072857, cleared on Feb. 4, 2008 Sulzer Orthopedics MS-30 Femoral Stem, K993043, cleared on Dec. 2, 1999 Sulzer Orthopedics MS-30 Lateral Femoral Stems, K020713,…
Sulzer Orthopedics, Inc.1999—
*K020713MS-30 Lateral Femoral Stem
Summary page 1
…MS-30 Femoral Stem, K993043, cleared on Dec. 2, 1999 Sulzer Orthopedics MS-30 Lateral Femoral Stems, K020713, cleared on May 14, 2002 Product Description: The Medacta Total Hip Prosthesis System - Quadra C femoral stem is…
Sulzer Orthopedics, Inc.20021

Also cleared against K072857: 5 other devices →

What descends from it? (9)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K192827Quadra-PMedacta International S.A.20206
K103189Medacta Total Hip Prosthesis System - Amistem C & Quadra C Short NeckMedacta International S.A.20101
K220405Amis K LongMedacta International S.A.20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 17, 2012Class IIZ-0791-2013Device DesignThe tip of the Quadra Trial Broach size 0, broke during a surgery.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 5.7 years on median.
  • 2 of the 8 ancestors have a recall of any class.
  • Of its descendants, K192827 (Medacta International S.A., 2020) has gone furthest, with 6 descendants of its own.
  • 5 other devices also cite its predicate K072857.

How does its category compare?

All LZO devices

LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · LZO

14in 2025

2007: 92008: 112009: 72010: 152011: 132012: 142013: 262014: 132015: 122016: 102017: 162018: 182019: 282020: 182021: 122022: 112023: 102024: 112025: 142026 (partial year): 9
2007high 28 (2019) · low 7 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZO

119in 2025

2007: 87 days median2008: 63 days median2009: 83 days median2010: 78 days median2011: 87 days median2012: 98 days median2013: 100 days median2014: 85 days median2015: 115 days median2016: 177 days median2017: 257 days median2018: 163 days median2019: 109 days median2020: 112 days median2021: 170 days median2022: 279 days median2023: 122 days median2024: 111 days median2025: 119 days median2026 (partial year): 102 days median
2007high 279 (2022) · low 63 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZO

1in 2025

2007: 22008: 132009: 422011: 102012: 142014: 12015: 12016: 92017: 52018: 22019: 42020: 32022: 42023: 22024: 12025: 1
2007high 42 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZO

3,244in 2025

2007: 1382008: 1252009: 1892010: 3942011: 5442012: 9632013: 1,3702014: 1,4772015: 1,3482016: 1,7282017: 2,1652018: 3,1332019: 3,0662020: 1,9632021: 2,1992022: 1,9852023: 2,0662024: 2,5862025: 3,244
2007high 3,244 (2025) · low 125 (2008)2025

MAUDE reports naming product code LZO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20079872138
2008116313125
200978342189
201015780394
2011138710544
2012149814963
20132610001,370
2014138511,477
20151211511,348
20161017791,728
20171625752,165
20181816323,133
20192810943,066
20201811231,963
20211217002,199
20221127941,985
20231012222,066
20241111112,586
20251411913,244
202691020—

What is it indicated for?

The Medacta Total Hip Prosthesis System - Quadra C Femoral Stems are intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery. Hip replacement is indicated in the following cases: Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia Avascular necrosis of the femoral head Acute traumatic fracture of the femoral head or neck…

Read the full statement (104 words)

The Medacta Total Hip Prosthesis System - Quadra C Femoral Stems are intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery.

Hip replacement is indicated in the following cases:

Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia

Avascular necrosis of the femoral head

Acute traumatic fracture of the femoral head or neck

Failure of previous hip surgery, joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

Size 0 implants should not be implanted in patients with a mass of 65 kg or greater.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K083558: Medacta Total Hip Prosthesis System - Quadra C.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K083558

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026