510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

R3 Constrained Liner, Constrained Liner

K083566510(k) TraditionalCleared Mar 3, 200990-day reviewClass II21 CFR 888.3310Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smith & Nephew, Inc. · KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer

Cleared on Mar 3, 2009 as substantially equivalent to P960047 (Howmedica Osteonics Corp., 1997) after a 90-day review. Its family tree reaches back 1 generation to 1997. 10 later devices descend from it. It has been recalled 13 times.

Decided
Mar 3, 2009
Review
90 days
Class
Class II
Product code
KWZ
Recalls
13
Predicates
1

Its longest line of descent runs back 1 generation to P960047 (1997). 10 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2022: 1 descendant cleared2013: 1 descendant cleared2011: 1 descendant cleared2012: 1 descendant cleared2025: 1 descendant cleared2017: 1 descendant cleared2024: 2 descendants cleared2023: 1 descendant cleared2026: 1 descendant clearedReference device: K021803 · Reflection Constrained Liner · Smith & Nephew, Inc. · 2002Reference device: K023743 · Smith & Nephew Global Bipolar System · Smith & Nephew, Inc. · 2003 · recalledP960047 · Osteonics Constrained Acetabular Insert · Howmedica Osteonics Corp. · 19971995200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 6 descendants3 generations forward: 1 descendant P960047 · Osteonics Constrained Acetabular Insert · Howmedica Osteonics Corp. · 19970
  1. Root · 1 generation backP960047 Osteonics Constrained Acetabular InsertHowmedica Osteonics Corp. · 1997
  2. This deviceK083566 R3 Constrained Liner, Constrained LinerSmith & Nephew, Inc. · 2009
  3. Descendant · cites it directly · 4 of its ownK111635 Smith & Nephew RJ Constrained LinersSmith & Nephew, Inc. · 2011
  4. Descendant · cites it directly · 3 of its ownK211176 Smith & Nephew Hip SystemsSmith & Nephew, Inc. · 2022 · Recalled
  5. Descendant · 2 generations on · 1 of its ownK241461 Mpact Constrained LinerMedacta International S.A. · 2025
  6. Descendant · 2 generations on · 1 of its ownK122139 Smith & Nephew R3 Constrained LinersSmith & Nephew, Inc. · 2012
  7. Descendant · 3 generations on · 0 of its ownK260173 M-Vizion Femoral Revision System ExtensionMedacta International S.A. · 2026

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2011: 12012: 12013: 12017: 12022: 12023: 12024: 22025: 12026: 1
2011high 2 (2024) · low 0 (2014)2026

Most influential descendants

Showing 5 of 10 · See all
  • K111635Smith & Nephew RJ Constrained LinersSmith & Nephew, Inc. · 2011 · cites it directly4descendants
  • K211176Smith & Nephew Hip SystemsSmith & Nephew, Inc. · 2022 · cites it directly3descendants
  • K241461Mpact Constrained LinerMedacta International S.A. · 2025 · 2 generations on1descendant
  • K122139Smith & Nephew R3 Constrained LinersSmith & Nephew, Inc. · 2012 · 2 generations on1descendant
  • K260173M-Vizion Femoral Revision System ExtensionMedacta International S.A. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1P960047Osteonics Constrained Acetabular InsertHowmedica Osteonics Corp.1997—
0 (this device)K083566R3 Constrained Liner, Constrained LinerSmith & Nephew, Inc.2009—
+1K111635Smith & Nephew RJ Constrained LinersSmith & Nephew, Inc.2011—
+1K211176Smith & Nephew Hip SystemsSmith & Nephew, Inc.20221
+2K241461Mpact Constrained LinerMedacta International S.A.2025—
+2K122139Smith & Nephew R3 Constrained LinersSmith & Nephew, Inc.2012—
+3K260173M-Vizion Femoral Revision System ExtensionMedacta International S.A.2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201210
201310
201710
202210
202310
202420
202510
202610
FIG. 1 — Lineage of K083566, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWZ device

  1. #1 K241461 Mpact Constrained Linerscore 15.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*P960047Osteonics Constrained Acetabular Insert
Summary page 1
…e',~''7,~~: Osteonics Constrained Osteonics Corporation Osteonics Cnst P960047 6/10/1997 AcetabularInsert Smith & Nephew, Inc. REFLECTION Constrain…
Howmedica Osteonics Corp.1997—
†K021803Reflection Constrained LinerSmith & Nephew, Inc.2002—
†K023743Smith & Nephew Global Bipolar SystemSmith & Nephew, Inc.20033

Also cleared against P960047: 5 other devices →

What descends from it? (10)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111635Smith & Nephew RJ Constrained LinersSmith & Nephew, Inc.20114
K211176Smith & Nephew Hip SystemsSmith & Nephew, Inc.20223
K123598Smith & Nephew, Inc. Hip System InstrumentationSmith & Nephew, Inc.20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 6, 2015Class IIZ-1739-2015Process change controlThe R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
May 6, 2015Class IIZ-1735-2015Process change controlThe R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
May 6, 2015Class IIZ-1738-2015Process change controlThe R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
May 6, 2015Class IIZ-1736-2015Process change controlThe R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
May 6, 2015Class IIZ-1737-2015Process change controlThe R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
Aug 13, 2010Not classifiedZ-0081-2011OtherThe firm received reports of a femoral head popping out of the liner intraoperatively.
Aug 13, 2010Not classifiedZ-0085-2011OtherThe firm received reports of a femoral head popping out of the liner intraoperatively.
Aug 13, 2010Not classifiedZ-0080-2011OtherThe firm received reports of a femoral head popping out of the liner intraoperatively.
Aug 13, 2010Not classifiedZ-0082-2011OtherThe firm received reports of a femoral head popping out of the liner intraoperatively.
Aug 13, 2010Not classifiedZ-0086-2011OtherThe firm received reports of a femoral head popping out of the liner intraoperatively.
Aug 13, 2010Not classifiedZ-0079-2011OtherThe firm received reports of a femoral head popping out of the liner intraoperatively.
Aug 13, 2010Not classifiedZ-0083-2011OtherThe firm received reports of a femoral head popping out of the liner intraoperatively.
Aug 13, 2010Not classifiedZ-0084-2011OtherThe firm received reports of a femoral head popping out of the liner intraoperatively.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K111635 (Smith & Nephew, Inc., 2011) has gone furthest, with 4 descendants of its own.
  • 5 other devices also cite its predicate P960047.

How does its category compare?

All KWZ devices

KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · KWZ

1in 2025

2007: 72008: 12009: 12010: 12011: 12012: 12016: 12017: 22024: 12025: 1
2007high 7 (2007) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWZ

104in 2007

2007: 104 days median
2007high 104 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWZ

1in 2023

2010: 82012: 12023: 12026 (partial year): 1
2007high 8 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWZ

193in 2025

2007: 412008: 442009: 812010: 982011: 1272012: 1682013: 2742014: 2782015: 6362016: 1,1802017: 4772018: 2832019: 5322020: 2882021: 2162022: 1582023: 1642024: 1452025: 193
2007high 1,180 (2016) · low 41 (2007)2025

MAUDE reports naming product code KWZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077104041
20081—044
20091—081
20101—898
20111—0127
20121—1168
20130—0274
20140—0278
20150—0636
20161—01,180
20172—0477
20180—0283
20190—0532
20200—0288
20210—0216
20220—0158
20230—1164
20241—0145
20251—0193
20260—1—

What is it indicated for?

The R3 Constrained Liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained Liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained Liner is intended for single use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K083566: R3 Constrained Liner, Constrained Liner.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K083566

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026