510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Atomlab 400 and Atomlab 500 Dose Calibrators, Models 086-330, -331, -332, -335 and -336

K090296510(k) TraditionalCleared Feb 19, 200913-day reviewClass II21 CFR 892.1360Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biodex Medical Systems, Inc. · KPT Calibrator, Dose, Radionuclide

Cleared on Feb 19, 2009 as substantially equivalent to K884312 (Atomic Products Corp., 1989) and 1 other predicate after a 13-day review. Its family tree reaches back 1 generation to 1989. It has been recalled 2 times.

Decided
Feb 19, 2009
Review
13 days
Class
Class II
Product code
KPT
Recalls
2
Predicates
2

Its longest line of descent runs back 1 generation to K884312 (1989). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2007: 1 other ancestorK884312 · Atomlab 100, Model 086-250 & 200 Model 086-270 · Atomic Products Corp. · 1989198519901995200020052010
Other ancestors, by generation 1 generation back: 1 other ancestor K884312 · Atomlab 100, Model 086-250 & 200 Model 086-270 · Atomic Products Corp. · 19890
  1. Root · 1 generation backK884312 Atomlab 100, Model 086-250 & 200 Model 086-270Atomic Products Corp. · 1989
  2. This deviceK090296 Atomlab 400 and Atomlab 500 Dose Calibrators, Models 086-330, -331…Biodex Medical Systems, Inc. · 2009
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K884312Atomlab 100, Model 086-250 & 200 Model 086-270Atomic Products Corp.1989—
0 (this device)K090296Atomlab 400 and Atomlab 500 Dose Calibrators, Models 086-330, -331, -332, -335…Biodex Medical Systems, Inc.2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200710
FIG. 1 — Lineage of K090296, 1989–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K884312Atomlab 100, Model 086-250 & 200 Model 086-270
Summary page 1
…ces. Classification #: 90 KPT, 21 CFR #892.1360 The Atonilab dose calibrators comply with Performance Standard IEC 1303 and IEC 61145. Predicate Devices: Atomlab 1.00 (K884312/A), Atomlab 200 (K884312/A), Capintec 25 (K071396). The basic detection, measurement process, design concepts, functionality, ca…
Atomic Products Corp.1989—
*K071396CRC 25 Series (CRC 25R, CRC 25PET, CRC 25W) Dose Calibrators
Summary page 1
…he Atonilab dose calibrators comply with Performance Standard IEC 1303 and IEC 61145. Predicate Devices: Atomlab 1.00 (K884312/A), Atomlab 200 (K884312/A), Capintec 25 (K071396). The basic detection, measurement process, design concepts, functionality, calculations and algorithms remain the same as the p…
Capintec, Inc.20071

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 16, 2021Class IIZ-2580-2021Software designWhen deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose…
Aug 16, 2021Class IIZ-2581-2021Software designWhen deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.

How does its category compare?

All KPT devices

KPT Calibrator, Dose, Radionuclide: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · KPT

1in 2019

2007: 12009: 12010: 12014: 22017: 12019: 1
2007high 2 (2014) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPT

Too few KPT 510(k)s in any year since 2007 to give a median.

Recalls per year · KPT

1in 2011

2010: 12011: 1
2007high 1 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPT

No MAUDE report names KPT since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20080—00
20091—00
20101—10
20110—10
20120—00
20130—00
20142—00
20150—00
20160—00
20171—00
20180—00
20191—00
20200—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The Atomlab dose calibrators are used in Nuclear Medicine Departments, clinics and nuclear pharmacies for calibrating radioactive doses for administration to patients.

The instrument is used to verify the radioactivity of a radionuclide before administration to the patient.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K090296: Atomlab 400 and Atomlab 500 Dose Calibrators, Models 086-330, -331, -332, -335 and -336.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K090296

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026