510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Human Igg Subclass Liquid Reagent Kits (for Use on Hitachi 911/ 912/ 917 and Modular P)

K090920510(k) TraditionalCleared Dec 30, 2009273-day reviewClass II21 CFR 866.5510Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

The Binding Site, Ltd. · CFN Method, Nephelometric, Immunoglobulins (G, A, M)

Cleared on Dec 30, 2009 after a 273-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. It has been recalled once.

Decided
Dec 30, 2009
Review
273 days
Class
Class II
Product code
CFN
Recalls
1
Predicates
0

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

FIG. 1 — Lineage of K090920Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CFN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CFN device

  1. #1 K233663 N Antisera to Human Immunoglobulins (IgG, IgA…score 8.3 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 11, 2014Class IIZ-1070-2017Nonconforming Material/ComponentCustomer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

No ancestor or descendant has a Class I recall.

How does its category compare?

All CFN devices

CFN Method, Nephelometric, Immunoglobulins (G, A, M): FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · CFN

2in 2023

2008: 72009: 42010: 22012: 12013: 42014: 22015: 12017: 12019: 62023: 2
2007high 7 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CFN

28in 2019

2008: 69 days median2009: 160 days median2013: 431 days median2019: 28 days median
2007high 431 (2013) · low 28 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CFN

1in 2023

2014: 52018: 22022: 22023: 1
2007high 5 (2014) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CFN

1in 2024

2010: 32012: 32016: 22017: 42018: 22019: 82022: 12024: 1
2007high 8 (2019) · low 0 (2007)2025

MAUDE reports naming product code CFN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
200876900
2009416000
20102—03
20110—00
20121—03
2013443100
20142—50
20151—00
20160—02
20171—04
20180—22
201962808
20200—00
20210—00
20220—21
20232—10
20240—01
20250—00
20260—0—
Cite this page

510(k) Report. “K090920: Human Igg Subclass Liquid Reagent Kits (for Use on Hitachi 911/ 912/ 917 and Modular P).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K090920

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026