# K091146: Modification to Cyberknife Robotic Radiosurgery System

- Pathway: 510(k) (Special)
- Applicant: Accuray, Inc.
- Decided: May 1, 2009 after 11 days
- Product code: MUJ (System, Planning, Radiation Therapy Treatment)
- Device class: 2
- Regulation: 21 CFR 892.5050
- Recalls of this device: 3 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. The FDA summary names no predicate by number.

## Recalls of this device
- Jul 6, 2010 · Not classified · Z-0890-2011: Latches responsible for securing the cover to the device may come lose if not properly secured and the cover may unexpectedly drop off. A design change is planned.
- Jan 15, 2010 · Not classified · Z-1023-2011: When a user modified the align center of an existing plan during the treatment planning process the effective depths of each beam in the plan were not updated by the system. These depths were incorrec
- Dec 29, 2009 · Not classified · Z-1008-2011: If electron density values are left empty, the calculation of radiation dose in a patient will be modeled as air-like density material rather than correct density. A plan may be created and saved, thu

## Indications for use
The CyberKnife® Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.