510(k) Report

510(k) · Cleared

Cleared

Smith & Nephew Instrument Trays

K091627510(k) SpecialCleared Jul 2, 200929-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smith & Nephew, Inc. · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Jul 2, 2009 as substantially equivalent to K073551 (Smith & Nephew Inc., Endoscopy Div., 2008) and 1 other predicate after a 29-day review. Its family tree reaches back 6 generations to 1984. 4 later devices descend from it.

Decided
Jul 2, 2009
Review
29 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
2

Its longest line of descent runs back 6 generations to K844002 (1984). 9 ancestors in all. 4 later devices descend from it, 3 citing it directly.

  1. Root · 6 generations backK844002 HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc. · 1984
  2. 5 generations backK980065 C/T Med-Systems Cassette SystemC/T Med System, Ltd. · 1998
  3. 4 generations backK990761 Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd. · 1999
  4. 3 generations backK012105 Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc. · 2002
  5. 2 generations backK073551 Smith & Nephew Instrument TraysSmith & Nephew Inc., Endoscopy Div. · 2008
  6. 1 generation backK090562 Smith & Nephew Instrument TraysSmith & Nephew, Inc. · 2009
  7. This deviceK091627 Smith & Nephew Instrument TraysSmith & Nephew, Inc. · 2009
  8. Descendant · cites it directly · 1 of its ownK102122 Smith & Nephew Milti-Purpose Sterilization Tray Model 72202428Smith & Nephew, Inc. · 2010
  9. Descendant · 2 generations on · 0 of its ownK203177 Sterisystem Dry-Base Instrument TrayO&M Halyard, Inc. · 2022
  10. Descendant · cites it directly · 0 of its ownK102544 Smith & Nephew Hip Arthroscopy Repair Instrument TraySmith & Nephew Inc., Endoscopy Div. · 2011 · Recalled
  11. Descendant · cites it directly · 0 of its ownK092551 Smith & Nephew Instrument TraysSmith & Nephew Inc., Endoscopy Div. · 2010

Most influential descendants

Showing 4 of 4 · See all
  • K102122Smith & Nephew Milti-Purpose Sterilization Tray Model 72202428Smith & Nephew, Inc. · 2010 · cites it directly1descendant
  • K203177Sterisystem Dry-Base Instrument TrayO&M Halyard, Inc. · 2022 · 2 generations on0descendants
  • K102544Smith & Nephew Hip Arthroscopy Repair Instrument TraySmith & Nephew Inc., Endoscopy Div. · 2011 · cites it directly0descendants
  • K092551Smith & Nephew Instrument TraysSmith & Nephew Inc., Endoscopy Div. · 2010 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K844002HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc.1984—
−5K980065C/T Med-Systems Cassette SystemC/T Med System, Ltd.1998—
−4K990761Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd.1999—
−3K012105Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc.2002—
−2K073551Smith & Nephew Instrument TraysSmith & Nephew Inc., Endoscopy Div.2008—
−1K090562Smith & Nephew Instrument TraysSmith & Nephew, Inc.2009—
0 (this device)K091627Smith & Nephew Instrument TraysSmith & Nephew, Inc.2009—
+1K102122Smith & Nephew Milti-Purpose Sterilization Tray Model 72202428Smith & Nephew, Inc.2010—
+2K203177Sterisystem Dry-Base Instrument TrayO&M Halyard, Inc.2022—
+1K102544Smith & Nephew Hip Arthroscopy Repair Instrument TraySmith & Nephew Inc., Endoscopy Div.20111
+1K092551Smith & Nephew Instrument TraysSmith & Nephew Inc., Endoscopy Div.2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201020
201110
202210
FIG. 1 — Lineage of K091627, 1984–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K073551Smith & Nephew Instrument Trays
Summary page 1
…ded Use and Fundamental Scientific Technology to the following legally marketed device in commercial distribution: Smith & Nephew Instrument Tray cleared under K073551 and Smith & Nephew Instrument Trays cleared under K090562. E. Description of Device The Smith & Nephew instrument tr…
Smith & Nephew Inc., Endoscopy Div.2008—
*K090562Smith & Nephew Instrument Trays
Summary page 1
…wing legally marketed device in commercial distribution: Smith & Nephew Instrument Tray cleared under K073551 and Smith & Nephew Instrument Trays cleared under K090562. E. Description of Device The Smith & Nephew instrument trays are designed to contain and protect reusable surgical…
Smith & Nephew, Inc.2009—

Also cleared against K073551: 4 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102122Smith & Nephew Milti-Purpose Sterilization Tray Model 72202428Smith & Nephew, Inc.20101
K102544Smith & Nephew Hip Arthroscopy Repair Instrument TraySmith & Nephew Inc., Endoscopy Div.20110
K092551Smith & Nephew Instrument TraysSmith & Nephew Inc., Endoscopy Div.20100

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 6 companies; a new generation was cleared every 3.1 years on median.
  • None of the 9 ancestors has a recall.
  • Of its descendants, K102122 (Smith & Nephew, Inc., 2010) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K073551.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

Smith & Nephew instrument trays are intended to contain Smith & Nephew reusable surgical instruments for convenient organized storage, sterilization and transport between usages. The subject instrument trays are suitable for use in prevacuum steam and high temperature gravity steam sterilization methods. The subject instrument trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated sterilization wrap in…

Read the full statement (101 words)

Smith & Nephew instrument trays are intended to contain Smith & Nephew reusable surgical instruments for convenient organized storage, sterilization and transport between usages. The subject instrument trays are suitable for use in prevacuum steam and high temperature gravity steam sterilization methods. The subject instrument trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated sterilization wrap in order to maintain sterility of the enclosed devices.

Validated Sterilization Parameters:

Method Temperature Exposure Time Drying Time

Pre-vacuum steam 132C (270F) 4 minutes 30 minutes

High Temperature Gravity Steam 132C (270F) 15 minutes 70 minutes

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091627: Smith & Nephew Instrument Trays.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K091627

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026