510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Acl Top 700 Las

K091980510(k) SpecialCleared Sep 22, 200983-day reviewClass II21 CFR 864.5400Recalls 24

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Instrumentation Laboratory CO · GKP Instrument, Coagulation, Automated

Cleared on Sep 22, 2009 as substantially equivalent to K073377 (Instrumentation Laboratory CO, 2008) after an 83-day review. Its family tree reaches back 4 generations to 2000. 7 later devices descend from it. It has been recalled 24 times.

Decided
Sep 22, 2009
Review
83 days
Class
Class II
Product code
GKP
Recalls
24
Predicates
1

Its longest line of descent runs back 4 generations to K002400 (2000). 7 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2015: 1 descendant cleared2016: 1 descendant cleared2023: 2 descendants cleared2024: 1 descendant cleared2019: 1 descendant cleared2018: 1 descendant clearedK002400 · Modification of Acl Advance · Instrumentation Laboratory CO · 2000 · recalledK033414 · Acl Top Model, Top · Instrumentation Laboratory CO · 2004 · recalledK063679 · Acl Top · Instrumentation Laboratory CO · 2007 · recalledK073377 · Acl Top (with System Software V3.0.0) · Instrumentation Laboratory CO · 2008200020052010201520202025
  1. Root · 4 generations backK002400 Modification of Acl AdvanceInstrumentation Laboratory CO · 2000 · Recalled
  2. 3 generations backK033414 Acl Top Model, TopInstrumentation Laboratory CO · 2004 · Recalled
  3. 2 generations backK063679 Acl TopInstrumentation Laboratory CO · 2007 · Recalled
  4. 1 generation backK073377 Acl Top (with System Software V3.0.0)Instrumentation Laboratory CO · 2008
  5. This deviceK091980 Acl Top 700 LasInstrumentation Laboratory CO · 2009
  6. Descendant · cites it directly · 3 of its ownK150877 Acl Top Family 50 Series: Acl Top 350 CTS; Acl Top 550 CTS; Acl Top…Instrumentation Laboratory CO · 2015 · Recalled
  7. Descendant · cites it directly · 2 of its ownK160276 Acl Top (Acl Top 700, Acl Top 700 CTS, Acl Top 700 Las, Acl Top 500…Instrumentation Laboratory CO · 2016 · Recalled
  8. Descendant · 2 generations on · 0 of its ownK242127 ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750…Instrumentation Laboratory (IL) Co. · 2024
  9. Descendant · 2 generations on · 0 of its ownK233790 Acl Top 970 CLInstrumentation Laboratory · 2023
  10. Descendant · 2 generations on · 0 of its ownK231031 ACL TOP Family 70 SeriesInstrumentation Laboratory Company · 2023

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2015: 12016: 12018: 12019: 12023: 22024: 1
2015high 2 (2023) · low 0 (2017)2026

Most influential descendants

Showing 5 of 7 · See all
  • K150877Acl Top Family 50 Series: Acl Top 350 CTS; Acl Top 550 CTS; Acl Top 750; Acl Top 750…Instrumentation Laboratory CO · 2015 · cites it directly3descendants
  • K160276Acl Top (Acl Top 700, Acl Top 700 CTS, Acl Top 700 Las, Acl Top 500 CTS, Acl Top 300 CTS)Instrumentation Laboratory CO · 2016 · cites it directly2descendants
  • K242127ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550…Instrumentation Laboratory (IL) Co. · 2024 · 2 generations on0descendants
  • K233790Acl Top 970 CLInstrumentation Laboratory · 2023 · 2 generations on0descendants
  • K231031ACL TOP Family 70 SeriesInstrumentation Laboratory Company · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K002400Modification of Acl AdvanceInstrumentation Laboratory CO20002
−3K033414Acl Top Model, TopInstrumentation Laboratory CO20041
−2K063679Acl TopInstrumentation Laboratory CO20071
−1K073377Acl Top (with System Software V3.0.0)Instrumentation Laboratory CO2008—
0 (this device)K091980Acl Top 700 LasInstrumentation Laboratory CO2009—
+1K150877Acl Top Family 50 Series: Acl Top 350 CTS; Acl Top 550 CTS; Acl Top 750; Acl…Instrumentation Laboratory CO20151
+1K160276Acl Top (Acl Top 700, Acl Top 700 CTS, Acl Top 700 Las, Acl Top 500 CTS, Acl…Instrumentation Laboratory CO20161
+2K242127ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL…Instrumentation Laboratory (IL) Co.2024—
+2K233790Acl Top 970 CLInstrumentation Laboratory2023—
+2K231031ACL TOP Family 70 SeriesInstrumentation Laboratory Company2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201610
201810
201910
202320
202410
FIG. 1 — Lineage of K091980, 2000–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKP device

  1. #1 K242127 ACL TOP Family 50 Series (ACL TOP 750; ACL TOP…score 9.4 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K073377Acl Top (with System Software V3.0.0)
Summary page 1
…Section: Coagulation Instrument (864.5400) Classification: Class II Product Code: GKP Panel: Hematology Predicate Device: K073377 ACL TOP Indications for Use: The ACL TOP is a bench top, fully automated, random acces's analyzer designed spe…
Instrumentation Laboratory CO2008—

Also cleared against K073377: 3 other devices →

What descends from it? (7)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K150877Acl Top Family 50 Series: Acl Top 350 CTS; Acl Top 550 CTS; Acl Top 750; Acl Top 750…Instrumentation Laboratory CO20153
K160276Acl Top (Acl Top 700, Acl Top 700 CTS, Acl Top 700 Las, Acl Top 500 CTS, Acl Top 300 CTS)Instrumentation Laboratory CO20162

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 3, 2014Class IIZ-2061-2014Device DesignPotential for sample misidentification.
Jun 3, 2014Class IIZ-2066-2014Device DesignPotential for sample misidentification.
Jun 3, 2014Class IIZ-2064-2014Device DesignPotential for sample misidentification.
Jun 3, 2014Class IIZ-2060-2014Device DesignPotential for sample misidentification.
Jun 3, 2014Class IIZ-2063-2014Device DesignPotential for sample misidentification.
Jun 3, 2014Class IIZ-2062-2014Device DesignPotential for sample misidentification.
Jun 3, 2014Class IIZ-2065-2014Device DesignPotential for sample misidentification.
Jul 19, 2013Class IIZ-2248-2013Device DesignPotential carryover issue that can can cause shortened APTT clotting times on the instrument.
Jul 19, 2013Class IIZ-2244-2013Device DesignPotential carryover issue that can can cause shortened APTT clotting times on the instrument.
Jul 19, 2013Class IIZ-2245-2013Device DesignPotential carryover issue that can can cause shortened APTT clotting times on the instrument.
Jul 19, 2013Class IIZ-2246-2013Device DesignPotential carryover issue that can can cause shortened APTT clotting times on the instrument.
Jul 19, 2013Class IIZ-2250-2013Device DesignPotential carryover issue that can can cause shortened APTT clotting times on the instrument.
Jul 19, 2013Class IIZ-2247-2013Device DesignPotential carryover issue that can can cause shortened APTT clotting times on the instrument.
Jul 19, 2013Class IIZ-2249-2013Device DesignPotential carryover issue that can can cause shortened APTT clotting times on the instrument.
Aug 3, 2012Class IIZ-2387-2012Software designLow Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low
Aug 3, 2012Class IIZ-2388-2012Software designLow Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low
Aug 3, 2012Class IIZ-2392-2012Software designLow Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low
Aug 3, 2012Class IIZ-2390-2012Software designLow Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low
Aug 3, 2012Class IIZ-2389-2012Software designLow Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low
Aug 3, 2012Class IIZ-2391-2012Software designLow Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low
Aug 13, 2010Not classifiedZ-0135-2011Software designSoftware: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves
Aug 13, 2010Not classifiedZ-0137-2011Software designSoftware: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves
Aug 13, 2010Not classifiedZ-0136-2011Software designSoftware: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves
Aug 13, 2010Not classifiedZ-0138-2011Software designSoftware: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 1 company; a new generation was cleared every 3.0 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K150877 (Instrumentation Laboratory CO, 2015) has gone furthest, with 3 descendants of its own.
  • 3 other devices also cite its predicate K073377.

How does its category compare?

All GKP devices

GKP Instrument, Coagulation, Automated: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · GKP

1in 2024

2007: 12008: 12009: 12015: 12016: 12023: 12024: 1
2007high 1 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKP

Too few GKP 510(k)s in any year since 2007 to give a median.

Recalls per year · GKP

1in 2022

2007: 32010: 42022: 1
2007high 4 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GKP

5in 2024

2007: 32008: 52009: 1902010: 742011: 452012: 352013: 422014: 262015: 242016: 132017: 182018: 192019: 382020: 142021: 172022: 42023: 42024: 5
2007high 190 (2009) · low 0 (2025)2025

MAUDE reports naming product code GKP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—33
20081—05
20091—0190
20100—474
20110—045
20120—035
20130—042
20140—026
20151—024
20161—013
20170—018
20180—019
20190—038
20200—014
20210—017
20220—14
20231—04
20241—05
20250—00
20260—0—

What is it indicated for?

The ACL TOP 700 LAS is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis.

The system provides results for both direct hemostasis measurements and calculated parameters.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091980: Acl Top 700 Las.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K091980

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026