510(k) Report

510(k) · Cleared

Cleared

Codman Sterilization Containers

K092437510(k) TraditionalCleared Mar 25, 2010230-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Codman & Shurtleff, Inc. · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Mar 25, 2010 as substantially equivalent to K072563 (Miltex, Inc., 2008) after a 230-day review. Its family tree reaches back 3 generations to 1987. 2 later devices descend from it.

Decided
Mar 25, 2010
Review
230 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K871202 (1987). 6 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1997: 3 other ancestors2012: 1 descendant cleared2016: 1 descendant clearedK871202 · Flashpak Container System · Performance Instruments Corp. · 1987K022978 · Steritite Container with Flashtite Valve Plate (S) & Meditray Products · Case Medical, Inc. · 2003K072563 · Miltex Rigid Container Systems · Miltex, Inc. · 20081990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 3 other ancestors 1 generation forward: 2 descendants K072563 · Miltex Rigid Container Systems · Miltex, Inc. · 2008K022978 · Steritite Container with Flashtite Valve Plate (S) & Meditray Products · Case Medical, Inc. · 2003K871202 · Flashpak Container System · Performance Instruments Corp. · 19870
  1. Root · 3 generations backK871202 Flashpak Container SystemPerformance Instruments Corp. · 1987
  2. 2 generations backK022978 Steritite Container with Flashtite Valve Plate (S) & Meditray ProductsCase Medical, Inc. · 2003
  3. 1 generation backK072563 Miltex Rigid Container SystemsMiltex, Inc. · 2008
  4. This deviceK092437 Codman Sterilization ContainersCodman & Shurtleff, Inc. · 2010
  5. Descendant · cites it directly · 1 of its ownK120117 Codman Quad-Lock Sterilization Container SystemCodman & Shurtleff, Inc. · 2012 · Recalled
  6. Descendant · cites it directly · 0 of its ownK152111 Symmetry Surgical Quad-Lock Container SystemSymmetry Surgical, Inc. · 2016

Most influential descendants

Showing 2 of 2 · See all
  • K120117Codman Quad-Lock Sterilization Container SystemCodman & Shurtleff, Inc. · 2012 · cites it directly1descendant
  • K152111Symmetry Surgical Quad-Lock Container SystemSymmetry Surgical, Inc. · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K871202Flashpak Container SystemPerformance Instruments Corp.1987—
−2K022978Steritite Container with Flashtite Valve Plate (S) & Meditray ProductsCase Medical, Inc.2003—
−1K072563Miltex Rigid Container SystemsMiltex, Inc.2008—
0 (this device)K092437Codman Sterilization ContainersCodman & Shurtleff, Inc.2010—
+1K120117Codman Quad-Lock Sterilization Container SystemCodman & Shurtleff, Inc.20121
+1K152111Symmetry Surgical Quad-Lock Container SystemSymmetry Surgical, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199730

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201610
FIG. 1 — Lineage of K092437, 1987–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072563Miltex Rigid Container Systems
Summary page 1
…Class: Class 11per 21 CFR § 580.6850 Sterilization wrap Predicate Device(s): Miltex Rigid Sterilization Container System K072563 Device Description: The Codman Sterile Container System is designed and manufactured on…
Miltex, Inc.2008—

Also cleared against K072563: 1 other device →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120117Codman Quad-Lock Sterilization Container SystemCodman & Shurtleff, Inc.20121
K152111Symmetry Surgical Quad-Lock Container SystemSymmetry Surgical, Inc.20160

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 4.8 years on median.
  • 1 of the 6 ancestors has a recall of any class.
  • Of its descendants, K120117 (Codman & Shurtleff, Inc., 2012) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K072563.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The CODMAN Sterilization Container System is indicated for use by hospitals and by health care facilities to: organize and protect stainless steel, aluminum, and titanium general surgical instruments that will be sterilized allow sterilization of the contained instruments by prevacuum steam sterilization (validated, parameters shown in Table 1) maintain the sterility of the contents for up to 180 days during storage and transport within the health care…

Read the full statement (232 words)

The CODMAN Sterilization Container System is indicated for use by hospitals and by health care facilities to:

organize and protect stainless steel, aluminum, and titanium general surgical instruments that will be sterilized

allow sterilization of the contained instruments by prevacuum steam sterilization (validated, parameters shown in Table 1)

maintain the sterility of the contents for up to 180 days during storage and transport within the health care facility, as long as the integrity of the container has not been compromised

The system has not been tested for maintenance of sterility after transportation outside the health care facility.

The system is intended for use with stainless steel, aluminum, and titanium general surgical instruments.

The system is intended for use with lumen instruments as follows:

Lumen inner diameter 3 mm or larger

Lumen length 400 mm or less (medium and large size containers)

250 mm or less (small size containers)

Quantity Up to 4 lumen instruments per container

The system is not intended for the sterilization of endoscopes.

For effective sterilization and drying of any size CODMAN Sterilization Container, the recommended maximum combined weight of the single container, lid, basket and basket contents is 25 lb. (11.3 kg). Please refer to Appendix A for a table of validated configurations of lids, bottoms, safety covers, and accessories.

Table 1

Sterilization Parameters

Cycle Type: Prevacuum; Temperature: 132 °C; Exposure Time: 4 minutes; Dry Time: 30 minutes

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K092437: Codman Sterilization Containers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K092437

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026