510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Tissu Trans Filtron, Models 3-TT-Filtron 2000-500

K092482510(k) TraditionalCleared Dec 4, 2009113-day reviewClass II21 CFR 878.5040Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shippert Medical Technologies Corp. · MUU System, Suction, Lipoplasty

Cleared on Dec 4, 2009 as substantially equivalent to K072587 (Cytori Therapeutics, Inc., 2007) and 1 other predicate after a 113-day review. Its family tree reaches back 6 generations to 1991. 17 later devices descend from it. It has been recalled 8 times.

Decided
Dec 4, 2009
Review
113 days
Class
Class II
Product code
MUU
Recalls
8
Predicates
2

Its longest line of descent runs back 6 generations to K910696 (1991). 26 ancestors in all, the oldest K861153 (1986). 17 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor1988: 2 other ancestors1986: 2 other ancestors1998: 3 other ancestors2001: 1 other ancestor1999: 3 other ancestors1997: 2 other ancestors1996: 2 other ancestors1992: 1 other ancestor1995: 2 other ancestors1994: 1 other ancestor2012: 2 descendants cleared2010: 1 descendant cleared2017: 5 descendants cleared2015: 1 descendant cleared2014: 1 descendant cleared2020: 2 descendants cleared2019: 2 descendants cleared2016: 1 descendant cleared2022: 1 descendant cleared2021: 1 descendant clearedK910696 · Cusa (R) System 200C/200T/200M/200H Ultra Aspirator · Valleylab, Inc. · 1991K970471 · Mentor Ultrasound Assisted Tissue Removal System (TRS) · Mentor Corp. · 1997K983065 · Mentor Ultrasound-Assisted Tissue Removal System · Mentor Corp. · 1998K022051 · Soundvaser System · Sound Surgical Technologies, LLC · 2002 · recalledK042261 · Macropore Puricel Lipoplasty System · Macropore Biosurgery, Inc. · 2004K072587 · Cytori Aft System · Cytori Therapeutics, Inc. · 20071990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors3 generations back: 1 other ancestor4 generations back: 4 other ancestors5 generations back: 8 other ancestors6 generations back: 4 other ancestors 1 generation forward: 4 descendants2 generations forward: 6 descendants3 generations forward: 2 descendants4 generations forward: 3 descendants5 generations forward: 1 descendant6 generations forward: 1 descendant K072587 · Cytori Aft System · Cytori Therapeutics, Inc. · 2007K042261 · Macropore Puricel Lipoplasty System · Macropore Biosurgery, Inc. · 2004K022051 · Soundvaser System · Sound Surgical Technologies, LLC · 2002 · recalledK983065 · Mentor Ultrasound-Assisted Tissue Removal System · Mentor Corp. · 1998K970471 · Mentor Ultrasound Assisted Tissue Removal System (TRS) · Mentor Corp. · 1997K910696 · Cusa (R) System 200C/200T/200M/200H Ultra Aspirator · Valleylab, Inc. · 1991−50+5
  1. Root · 6 generations backK910696 Cusa (R) System 200C/200T/200M/200H Ultra AspiratorValleylab, Inc. · 1991
  2. 5 generations backK970471 Mentor Ultrasound Assisted Tissue Removal System (TRS)Mentor Corp. · 1997
  3. 4 generations backK983065 Mentor Ultrasound-Assisted Tissue Removal SystemMentor Corp. · 1998
  4. 3 generations backK022051 Soundvaser SystemSound Surgical Technologies, LLC · 2002 · Recalled
  5. 2 generations backK042261 Macropore Puricel Lipoplasty SystemMacropore Biosurgery, Inc. · 2004
  6. 1 generation backK072587 Cytori Aft SystemCytori Therapeutics, Inc. · 2007
  7. This deviceK092482 Tissu Trans Filtron, Models 3-TT-Filtron 2000-500Shippert Medical Technologies Corp. · 2009
  8. Descendant · cites it directly · 11 of its ownK120902 GID 700The Gid Group, Inc. · 2012 · Recalled
  9. Descendant · 2 generations on · 6 of its ownK142682 The Lipogems SystemLipogems International Spa · 2014
  10. Descendant · 3 generations on · 5 of its ownK161636 Lipogems SystemLipogems International Spa · 2016
  11. Descendant · 4 generations on · 2 of its ownK171135 Lipogems SystemLipogems International Spa · 2017
  12. Descendant · cites it directly · 2 of its ownK120328 Origins Lipoharvesting SystemSound Surgical Technologies, LLC · 2012

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2010: 12012: 22014: 12015: 12016: 12017: 52019: 22020: 22021: 12022: 1
2010high 5 (2017) · low 0 (2011)2026

Most influential descendants

Showing 5 of 17 · See all
  • K120902GID 700The Gid Group, Inc. · 2012 · cites it directly11descendants
  • K142682The Lipogems SystemLipogems International Spa · 2014 · 2 generations on6descendants
  • K161636Lipogems SystemLipogems International Spa · 2016 · 3 generations on5descendants
  • K171135Lipogems SystemLipogems International Spa · 2017 · 4 generations on2descendants
  • K120328Origins Lipoharvesting SystemSound Surgical Technologies, LLC · 2012 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K910696Cusa (R) System 200C/200T/200M/200H Ultra AspiratorValleylab, Inc.1991—
−5K970471Mentor Ultrasound Assisted Tissue Removal System (TRS)Mentor Corp.1997—
−4K983065Mentor Ultrasound-Assisted Tissue Removal SystemMentor Corp.1998—
−3K022051Soundvaser SystemSound Surgical Technologies, LLC20022
−2K042261Macropore Puricel Lipoplasty SystemMacropore Biosurgery, Inc.2004—
−1K072587Cytori Aft SystemCytori Therapeutics, Inc.2007—
0 (this device)K092482Tissu Trans Filtron, Models 3-TT-Filtron 2000-500Shippert Medical Technologies Corp.2009—
+1K120902GID 700The Gid Group, Inc.20121
+2K142682The Lipogems SystemLipogems International Spa2014—
+3K161636Lipogems SystemLipogems International Spa2016—
+4K171135Lipogems SystemLipogems International Spa2017—
+1K120328Origins Lipoharvesting SystemSound Surgical Technologies, LLC2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198620
198820
199210
199410
199520
199620
199720
199830
199930
200110
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201220
201410
201510
201610
201750
201920
202020
202110
202210
FIG. 1 — Lineage of K092482, 1991–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUU? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUU device

  1. #1 K210528 IntelliFat Disposable Adipose Tissue…score 5.7 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072587Cytori Aft System
Summary page 4
…external sources of vacuum to withdraw, hold, and/or inject fluids/tissues into the body. The Tissu Trans Filtron is substantially equivalent to the Cytori AFT System k072587, The Shippert Medical Tissu Trans k<050797 and the Lipivage k# unknown, predicate devices as they all consist of a polycarbonate…
Cytori Therapeutics, Inc.2007—
*K050797Tissu-Trans
Summary page 3
…inciples with the following predicate devices,, c~aebeen determined byfFPA~to~bgsubstantiaIly equivalent to premar-ket devices: Cytari AFT,Systenr k 7257 l rn k050797, and Genesis Biasystems Lipivage (510(ky,number unknown); Class I and Class II medica!.devlces that were cleared for marketing in…
Shippert Medical Technologies Corp.2005—
*—Genesis Biosystems Lipivage
Named without a submission number
—

Also cleared against K072587: 5 other devices →

What descends from it? (17)

See all 17

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120902GID 700The Gid Group, Inc.201211
K120328Origins Lipoharvesting SystemSound Surgical Technologies, LLC20122
K170449Autoclavable Suction Jar with Luer Lock extension - 250, 500, 1,000, 2,000 and 3,000 ml…Millennium Medical Technologies, Inc.20170
K102117Tissu Trans Filtron Syringe FillShippert Medical Technologies Corp.20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 2, 2018Class IIZ-1330-2019Packaging process controlThe pouch integrity of the sterile product can't be assured without a visual examination due to possibly containing a cut or hole in the pouch.
Feb 2, 2018Class IIZ-1331-2019Packaging process controlThe pouch integrity of the sterile product can't be assured without a visual examination due to possibly containing a cut or hole in the pouch.
Feb 2, 2018Class IIZ-1333-2019Packaging process control
Feb 2, 2018Class IIZ-1332-2019Packaging process controlThe pouch integrity of the sterile product can't be assured without a visual examination due to possibly containing a cut or hole in the pouch.
Feb 2, 2018Class IIZ-1335-2019Packaging process controlThe pouch integrity of the sterile product can't be assured without a visual examination due to possibly containing a cut or hole in the pouch.
Feb 2, 2018Class IIZ-1336-2019Packaging process controlThe pouch integrity of the sterile product can't be assured without a visual examination due to possibly containing a cut or hole in the pouch.
Feb 2, 2018Class IIZ-1334-2019Packaging process control
Oct 20, 2017Class IIZ-0170-2018Packaging process controlPotential breach of sterility of the packaging.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 6 companies; a new generation was cleared every 3.2 years on median.
  • 4 of the 26 ancestors have a recall of any class.
  • Of its descendants, K120902 (The Gid Group, Inc., 2012) has gone furthest, with 11 descendants of its own.
  • 5 other devices also cite its predicate K072587.

How does its category compare?

All MUU devices

MUU System, Suction, Lipoplasty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · MUU

1in 2022

2007: 12008: 12009: 32010: 42012: 52013: 12014: 42015: 32016: 12017: 72018: 12019: 42020: 22021: 22022: 1
2007high 7 (2017) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUU

190in 2019

2009: 113 days median2010: 58 days median2012: 171 days median2014: 114 days median2015: 269 days median2017: 91 days median2019: 190 days median
2007high 269 (2015) · low 58 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUU

2in 2017

2007: 12015: 12016: 12017: 2
2007high 2 (2017) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUU

10in 2025

2007: 22008: 42009: 22010: 42011: 52012: 42013: 212014: 132015: 442016: 202017: 282018: 302019: 352020: 362021: 282022: 502023: 152024: 52025: 10
2007high 50 (2022) · low 2 (2007)2025

MAUDE reports naming product code MUU, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—12
20081—04
2009311302
201045804
20110—05
2012517104
20131—021
20144114013
20153269144
20161—120
2017791228
20181—030
20194190035
20202—036
20212—028
20221—050
20230—015
20240—05
20250—010
20260—0—

What is it indicated for?

The Tissu Trans Filtron is intended to be used with house vacuum and/or cleared pumps, tubing and cannulas, for the collection of aspirated fat, for aesthetic body contouring. If the fat is untreated, it may be reinjected via a cleared injection apparatus.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K092482: Tissu Trans Filtron, Models 3-TT-Filtron 2000-500.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K092482

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026