510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Syngo.X

K092519510(k) TraditionalCleared Aug 27, 20099-day reviewClass II21 CFR 892.2050Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens AG Medical Solutions · LLZ System, Image Processing, Radiological

Cleared on Aug 27, 2009 as substantially equivalent to K081734 (Siemens AG Medical Solutions, 2008) and 1 other predicate after a 9-day review. Its family tree reaches back 6 generations to 1997. 1 of the 192 devices in its family, counting itself, has a Class I recall. 176 later devices descend from it. It has been recalled 4 times.

Decided
Aug 27, 2009
Review
9 days
Class
Class II
Product code
LLZ
Recalls
4
Predicates
2

Its longest line of descent runs back 6 generations to K971684 (1997). 15 ancestors in all, the oldest K970852 (1997). 176 later devices descend from it, 11 citing it directly. 1 of the 192 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2007: 1 other ancestor2005: 3 other ancestors2001: 1 other ancestor2000: 1 other ancestor1999: 1 other ancestor1997: 1 other ancestor2010: 3 descendants cleared2011: 5 descendants cleared2012: 6 descendants cleared2016: 7 descendants cleared2014: 4 descendants cleared2015: 8 descendants cleared2019: 11 descendants cleared2018: 12 descendants cleared2013: 6 descendants cleared2020: 22 descendants cleared2017: 6 descendants cleared2023: 17 descendants cleared2021: 8 descendants cleared2025: 15 descendants cleared, 1 with a Class I recall12022: 16 descendants cleared2026: 15 descendants cleared2024: 15 descendants clearedK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK052423 · Magnetom Systems with Syngo Expert - I Option · Siemens Medi Cal Solutions, Inc. · 2006K061964 · Syngo Multimodality Workplace · Siemens AG Medical Solutions · 2006K072728 · Syngo Multimodality Workplace · Siemens AG Medical Solutions · 20081995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 1 other ancestor4 generations back: 6 other ancestors 1 generation forward: 11 descendants2 generations forward: 19 descendants3 generations forward: 35 descendants4 generations forward: 33 descendants5 generations forward: 29 descendants, 1 with a Class I recall16 generations forward: 21 descendants7 generations forward: 11 descendants8 generations forward: 5 descendants9 generations forward: 3 descendants10 generations forward: 5 descendants11 generations forward: 2 descendants12 generations forward: 1 descendant13 generations forward: 1 descendant K072728 · Syngo Multimodality Workplace · Siemens AG Medical Solutions · 2008K061964 · Syngo Multimodality Workplace · Siemens AG Medical Solutions · 2006K052423 · Magnetom Systems with Syngo Expert - I Option · Siemens Medi Cal Solutions, Inc. · 2006K041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalled−50+5+10
  1. Root · 6 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 5 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  3. 4 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 3 generations backK052423 Magnetom Systems with Syngo Expert - I OptionSiemens Medi Cal Solutions, Inc. · 2006
  5. 2 generations backK061964 Syngo Multimodality WorkplaceSiemens AG Medical Solutions · 2006
  6. 1 generation backK072728 Syngo Multimodality WorkplaceSiemens AG Medical Solutions · 2008
  7. This deviceK092519 Syngo.XSiemens AG Medical Solutions · 2009
  8. Descendant · cites it directly · 99 of its ownK112020 Syngo.CT Vascular AnalysisSiemens Medi Cal Solutions, Inc. · 2011
  9. Descendant · 2 generations on · 63 of its ownK123920 Syngo.ViaSiemens AG Medical Solutions · 2013
  10. Descendant · cites it directly · 54 of its ownK093621 Syngo.Pet&CT OncologySiemens Medical Solutions USA, Inc. · 2010
  11. Descendant · 3 generations on · 46 of its ownK150843 syngo®.via (version VB10A)Siemens AG · 2015
  12. Descendant · cites it directly · 25 of its ownK110366 Syngo.CT Cardiac FunctionSiemens AG · 2011 · Recalled

Showing 5 of 176 descendants. See all descendants

Descendants cleared per year

2010: 32011: 52012: 62013: 62014: 42015: 82016: 72017: 62018: 122019: 112020: 222021: 82022: 162023: 172024: 152025: 152026: 15
2010high 22 (2020) · low 3 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 176 · See all
  • K112020Syngo.CT Vascular AnalysisSiemens Medi Cal Solutions, Inc. · 2011 · cites it directly99descendants
  • K123920Syngo.ViaSiemens AG Medical Solutions · 2013 · 2 generations on63descendants
  • K093621Syngo.Pet&CT OncologySiemens Medical Solutions USA, Inc. · 2010 · cites it directly54descendants
  • K150843syngo®.via (version VB10A)Siemens AG · 2015 · 3 generations on46descendants
  • K110366Syngo.CT Cardiac FunctionSiemens AG · 2011 · cites it directly25descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K971684Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc.19974
−5K032428Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc.20034
−4K041112Magnetom Espree SystemSiemens Medical Solutions USA, Inc.20044
−3K052423Magnetom Systems with Syngo Expert - I OptionSiemens Medi Cal Solutions, Inc.2006—
−2K061964Syngo Multimodality WorkplaceSiemens AG Medical Solutions2006—
−1K072728Syngo Multimodality WorkplaceSiemens AG Medical Solutions2008—
0 (this device)K092519Syngo.XSiemens AG Medical Solutions2009—
+1K112020Syngo.CT Vascular AnalysisSiemens Medi Cal Solutions, Inc.2011—
+2K123920Syngo.ViaSiemens AG Medical Solutions2013—
+1K093621Syngo.Pet&CT OncologySiemens Medical Solutions USA, Inc.2010—
+3K150843syngo®.via (version VB10A)Siemens AG2015—
+1K110366Syngo.CT Cardiac FunctionSiemens AG20111

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199910
200010
200110
200530
200710
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201030
201150
201260
201360
201440
201580
201670
201760
2018120
2019110
2020220
202180
2022160
2023170
2024150
2025151
2026150
FIG. 1 — Lineage of K092519, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K081734Syngo Imaging (Version V31)
Summary page 4
…mCer FD Number K081734 Siemens syngo Imaging Siemens syngo MultiModality Workplace K072728 T…
Siemens AG Medical Solutions20085
*K072728Syngo Multimodality Workplace
Summary page 2
…. syngo.x allows to use a variety of advanced applications (clinical applica- tions) designed for syngo.x just as the predicate device syngo MultiMo- dality Workplace (K072728, cleared on April 22, 2007). These applica- tions are medical devices on their own rights and filed separately. They are not part…
Siemens AG Medical Solutions2008—

What descends from it? (176)

See all 176

11 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K112020Syngo.CT Vascular AnalysisSiemens Medi Cal Solutions, Inc.201199
K093621Syngo.Pet&CT OncologySiemens Medical Solutions USA, Inc.201054
K110366Syngo.CT Cardiac FunctionSiemens AG201125
K120315Syngo.MR Spectro Engine, Syngo.MR Spectro SVS, Syngo.MR Spectro Csi, Syngo.MR Spectro…Siemens Medical Solutions USA, Inc.201223
K113029Syngo Via MI WorkflowsSiemens Medical Solutions USA, Inc.201120
K101666Syngo. Plaza VA20ASiemens AG Medical Solutions20109
K121459Syngo.MR NeurologySiemens Medical Solutions USA, Inc.20123
K100637Syngo.CT Coronary AnalysisSiemens AG Medical Solutions20102
K111079Syngo.Via Web ViewerSiemens AG Medical Solutions20111
K120363Syngo RT InterfaceSiemens Medical Solutions USA, Inc.20120

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 6, 2017Class IIZ-1688-2017Software designSoftware changes now available to address several issues
Aug 5, 2016Class IIZ-2852-2016Software designSoftware error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high.
Jun 23, 2016Class IIZ-2245-2016Software designIncorrect values for the volume calculation. Software update VB30B via Update Instructions SY018/16/P to resolve software errors.
May 15, 2015Class IIZ-1705-2015Software designmeasurements drawn on the 2nd and subsequent images of the series are not visible on printouts when the series is sent to print.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K250443MAGNETOM Avanto Fit; MAGNETOM Skyra Fit; MAGNETOM Sola Fit; MAGNETOM Viato.MobileSiemens Healthcare GmbH · 2025 · 5 generations on · ◆ in the 2025 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 1.5 years on median.
  • 10 of the 15 ancestors have a recall of any class.
  • Of its descendants, K112020 (Siemens Medi Cal Solutions, Inc., 2011) has gone furthest, with 99 descendants of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

syngo.x is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images.

It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo.x based software options.

syngo.x supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments.

The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K092519: Syngo.X.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K092519

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026