510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

3.25 Spline Twist Implant, HA Coated

K093164510(k) TraditionalCleared Dec 15, 200970-day reviewClass II21 CFR 872.3640Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer Dental, Inc. · DZE Implant, Endosseous, Root-Form

Cleared on Dec 15, 2009 as substantially equivalent to K944327 (Calcitek, Inc., 1995) and 1 other predicate after a 70-day review. Its family tree reaches back 2 generations to 1995. It has been recalled once.

Decided
Dec 15, 2009
Review
70 days
Class
Class II
Product code
DZE
Recalls
1
Predicates
2

Its longest line of descent runs back 2 generations to K944327 (1995). 4 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by generation 2 generations back: 2 other ancestors K012055 · 3.25MM Spline Twist Implant · Sulzer Dental, Inc. · 2001K944327 · Integral VI Biointegrated Dental Implant System · Calcitek, Inc. · 19950
  1. Root · 2 generations backK944327 Integral VI Biointegrated Dental Implant SystemCalcitek, Inc. · 1995
  2. 1 generation backK012055 3.25MM Spline Twist ImplantSulzer Dental, Inc. · 2001
  3. This deviceK093164 3.25 Spline Twist Implant, HA CoatedZimmer Dental, Inc. · 2009
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K944327Integral VI Biointegrated Dental Implant SystemCalcitek, Inc.1995—
−1K0120553.25MM Spline Twist ImplantSulzer Dental, Inc.2001—
0 (this device)K0931643.25 Spline Twist Implant, HA CoatedZimmer Dental, Inc.2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199610
FIG. 1 — Lineage of K093164, 1995–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DZE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DZE device

  1. #1 K211090 ZENEX Implant Systemscore 25.4 · 15 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K944327Integral VI Biointegrated Dental Implant System
Summary page 1
…Integral VI Biointegrated Dental Implant System 510(k) Number: K944327 Classification: DZE Classification Name: Endosseous Dental Imp…
Calcitek, Inc.1995—
*K0120553.25MM Spline Twist Implant
Summary page 1
…Predicate Device #2 Trade Name: Small Diameter Spline Twist Implant 510(k) Number: K012055 Classification: DZE 3.25mm Spline®g Twist1" 4 Implant, HA Coated 51…
Sulzer Dental, Inc.2001—
†K960021Calcitek HA-Coated Endosseous Dental Implant SystemsCalcitek, Inc.1996—
†K953101Paragon Implant SystemCore-Vent Bio-Engineering, Inc.1995—
†K011028Screw-Vent Dental Implant SystemSulzer Dental, Inc.20012
†K013227Screw Vent Implant; Tapered Screw Vent ImplantSulzer Dental, Inc.20012
†K061410Screw-Vent Implant System, SV;Tapered Screw -Vent Implant System, TSV;Advent Implant…Zimmer Dental, Inc.20061
†K072589Tapered Screw-Vent Implant, 4.1MM, Models: TSV4B8, B10, B11, B13 and B16, TSV4H8, H11…Zimmer Dental, Inc.20073

Also cleared against K944327: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 1, 2014Class IIZ-0153-2015Labeling Change ControlZimmer Dental is recalling Tapered Screw-Vent Implants because the incorrect diameter size was printed on the inner vial.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 8.3 years on median.
  • None of the 4 ancestors has a recall.
  • 1 other device also cites its predicate K944327.

How does its category compare?

All DZE devices

DZE Implant, Endosseous, Root-Form: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DZE

31in 2025

2007: 432008: 612009: 412010: 342011: 262012: 392013: 422014: 362015: 362016: 402017: 482018: 412019: 312020: 412021: 342022: 342023: 352024: 402025: 312026 (partial year): 15
2007high 61 (2008) · low 26 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DZE

184in 2025

2007: 107 days median2008: 118 days median2009: 110 days median2010: 162 days median2011: 194 days median2012: 174 days median2013: 195 days median2014: 265 days median2015: 215 days median2016: 219 days median2017: 223 days median2018: 200 days median2019: 231 days median2020: 230 days median2021: 182 days median2022: 243 days median2023: 248 days median2024: 222 days median2025: 184 days median2026 (partial year): 194 days median
2007high 265 (2014) · low 107 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DZE

1in 2025

2007: 112008: 42009: 32010: 32012: 112013: 12014: 72015: 32016: 152017: 42018: 42019: 82021: 62024: 12025: 12026 (partial year): 2
2007high 15 (2016) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DZE

465,788in 2025

2007: 2252008: 3932009: 1642010: 3602011: 3622012: 3582013: 3912014: 1,1242015: 2,8942016: 1,1162017: 9,7192018: 12,8922019: 167,0162020: 255,5242021: 479,6182022: 483,2242023: 537,5852024: 473,6292025: 465,788
2007high 537,585 (2023) · low 164 (2009)2025

MAUDE reports naming product code DZE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074310711225
2008611184393
2009411103164
2010341623360
2011261940362
20123917411358
2013421951391
20143626571,124
20153621532,894
201640219151,116
20174822349,719
201841200412,892
2019312318167,016
2020412300255,524
2021341826479,618
2022342430483,224
2023352480537,585
2024402221473,629
2025311841465,788
2026151942—

What is it indicated for?

Zimmer Dental Spline implant systems are designed for use in edentulous mandibles or maxillae for attachment of complete denture prostheses for immediate or conventional loading, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a free standing single tooth replacement. The threaded implants should only be immediately loaded when good primary stability is achieved and the functional load is appropriate.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093164: 3.25 Spline Twist Implant, HA Coated.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093164

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026