Ringloc + Hybrid Acetabular System

K093235510(k) TraditionalCleared Apr 30, 2010197-day reviewClass III21 CFR 888.3330Recalls 30

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biomet, Inc. · KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)

Cleared on Apr 30, 2010 as substantially equivalent to K002379 (Biomet, Inc., 2000) and 1 other predicate after a 197-day review. Its family tree reaches back 5 generations to 1985. 65 later devices descend from it. It has been recalled 30 times.

Decided
Apr 30, 2010
Review
197 days
Class
Class III
Product code
KWA
Recalls
30
Predicates
2

Its longest line of descent runs back 5 generations to K854046 (1985). 21 ancestors in all. 65 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 2 other ancestors1994: 6 other ancestors1992: 1 other ancestor1999: 2 other ancestors1998: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1986: 1 other ancestor2012: 2 descendants cleared2016: 1 descendant cleared2017: 1 descendant cleared2015: 3 descendants cleared2014: 4 descendants cleared2019: 9 descendants cleared2018: 3 descendants cleared2022: 5 descendants cleared2026: 12 descendants cleared2024: 7 descendants cleared2025: 7 descendants cleared2021: 3 descendants cleared2020: 4 descendants cleared2023: 4 descendants clearedK854046 · Conical Collar Hip Prosthesis · Kirschner Medical Corp. · 1985K960303 · Fenning Femoral Component · Biomet, Inc. · 1996K030055 · Mallory/Head Total Hip System; HA Mallory/Head Total Hip System; BI-Metric Femoral Components · Biomet, Inc. · 2003 · recalledK050124 · Porous Coated Acetabular Components · Biomet, Inc. · 2005 · recalledK070369 · Regenerx Ringloc + Modular Acetabular Shells · Biomet, Inc. · 2007 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 4 other ancestors5 generations back: 9 other ancestors 1 generation forward: 3 descendants2 generations forward: 17 descendants3 generations forward: 23 descendants4 generations forward: 17 descendants5 generations forward: 3 descendants6 generations forward: 2 descendants K070369 · Regenerx Ringloc + Modular Acetabular Shells · Biomet, Inc. · 2007 · recalledK050124 · Porous Coated Acetabular Components · Biomet, Inc. · 2005 · recalledK030055 · Mallory/Head Total Hip System; HA Mallory/Head Total Hip System; BI-Metric Femoral Components · Biomet, Inc. · 2003 · recalledK960303 · Fenning Femoral Component · Biomet, Inc. · 1996K854046 · Conical Collar Hip Prosthesis · Kirschner Medical Corp. · 1985−50+5
  1. Root · 5 generations backK854046 Conical Collar Hip ProsthesisKirschner Medical Corp. · 1985
  2. 4 generations backK960303 Fenning Femoral ComponentBiomet, Inc. · 1996
  3. 3 generations backK030055 Mallory/Head Total Hip System; HA Mallory/Head Total Hip System…Biomet, Inc. · 2003 · Recalled
  4. 2 generations backK050124 Porous Coated Acetabular ComponentsBiomet, Inc. · 2005 · Recalled
  5. 1 generation backK070369 Regenerx Ringloc + Modular Acetabular ShellsBiomet, Inc. · 2007 · Recalled
  6. This deviceK093235 Ringloc + Hybrid Acetabular SystemBiomet, Inc. · 2010
  7. Descendant · cites it directly · 53 of its ownK121874 G7 Acetabular SystemBiomet Manufacturing Corp · 2012 · Recalled
  8. Descendant · 2 generations on · 35 of its ownK140669 G7 Osseoti Acetublar ShellsBiomet, Inc. · 2014
  9. Descendant · 2 generations on · 17 of its ownK142882 G7 Freedom and Offset Liners, Freedom Head, Size 32Biomet, Inc. · 2015
  10. Descendant · 3 generations on · 8 of its ownK221479 Persona Personalized Knee SystemZimmer, Inc. · 2022
  11. Descendant · 2 generations on · 8 of its ownK182321 MobileLink Acetabular Cup SystemWaldemar Link GmbH & Co. KG · 2019

Showing 5 of 65 descendants. See all descendants

Descendants cleared per year

2012: 22014: 42015: 32016: 12017: 12018: 32019: 92020: 42021: 32022: 52023: 42024: 72025: 72026: 12
2012high 12 (2026) · low 0 (2013)2026

Most influential descendants

Showing 5 of 65 · See all
  • K121874G7 Acetabular SystemBiomet Manufacturing Corp · 2012 · cites it directly53descendants
  • K140669G7 Osseoti Acetublar ShellsBiomet, Inc. · 2014 · 2 generations on35descendants
  • K142882G7 Freedom and Offset Liners, Freedom Head, Size 32Biomet, Inc. · 2015 · 2 generations on17descendants
  • K221479Persona Personalized Knee SystemZimmer, Inc. · 2022 · 3 generations on8descendants
  • K182321MobileLink Acetabular Cup SystemWaldemar Link GmbH & Co. KG · 2019 · 2 generations on8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K854046Conical Collar Hip ProsthesisKirschner Medical Corp.1985—
−4K960303Fenning Femoral ComponentBiomet, Inc.1996—
−3K030055Mallory/Head Total Hip System; HA Mallory/Head Total Hip System; BI-Metric…Biomet, Inc.20033
−2K050124Porous Coated Acetabular ComponentsBiomet, Inc.20051
−1K070369Regenerx Ringloc + Modular Acetabular ShellsBiomet, Inc.200719
0 (this device)K093235Ringloc + Hybrid Acetabular SystemBiomet, Inc.2010—
+1K121874G7 Acetabular SystemBiomet Manufacturing Corp20123
+2K140669G7 Osseoti Acetublar ShellsBiomet, Inc.2014—
+2K142882G7 Freedom and Offset Liners, Freedom Head, Size 32Biomet, Inc.2015—
+3K221479Persona Personalized Knee SystemZimmer, Inc.2022—
+2K182321MobileLink Acetabular Cup SystemWaldemar Link GmbH & Co. KG2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199210
199460
199510
199610
199710
199810
199920
200020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201220
201440
201530
201610
201710
201830
201990
202040
202130
202250
202340
202470
202570
2026120
FIG. 1 — Lineage of K093235, 1985–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002379M2A 28MM Ringloc Liner
Summary page 1
…KWA, LPH, LZO Legally Marketed Devices To Which Substantial Equivalence Is Claimed: M2a Ringloc® Acetabular Liners - K002379 Regenerex Ringloc®+ Modular Acetabular Shell - K070369 Device Description: The Porous Pl…
Biomet, Inc.2000—
*K070369Regenerx Ringloc + Modular Acetabular Shells
Summary page 1
…M2a Ringloc® Acetabular Liners - K002379 Regenerex Ringloc®+ Modular Acetabular Shell - K070369 Device Description: The Porous Plasma Spray (PPS) Ringloc®± Acetabular System is a series of acetabular shells that incorporate t…
Biomet, Inc.200719

Also cleared against K002379: 4 other devices →

What descends from it? (65)

See all 65

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121874G7 Acetabular SystemBiomet Manufacturing Corp201253
K112897Renovis Surgical Hip Joint Replacement ProsthesisRenovis Surgical Technologies, LLC20128
K160895Initia Total Hip System & BIOCERAM AZUL (R) HEADKyocera Medical Corporation20162

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 29, 2017Class IIZ-0312-2018Employee errorDevice was distributed with the etching missing from the product.
Jul 28, 2015Class IIZ-2483-2015Under Investigation by firmThe device was assembled without screw hole plugs.
Jan 9, 2015Class IIZ-1123-2015Packaging process controlBiomet recalled the Mallory Head 4 Finned Acetabular Shell 50mm (Lot 364600) and Ringloc + Acetabular Shell Limited Hole Finned 52 mm (Lot 427840) following a mix up in packaging with the two.
Apr 14, 2011Not classifiedZ-2425-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2433-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2447-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2445-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2421-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2428-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2422-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2430-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2444-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2432-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2441-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2424-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2431-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2426-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2435-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2423-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2438-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2442-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2446-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2434-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2427-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2436-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2440-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2437-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2443-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2439-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…
Apr 14, 2011Not classifiedZ-2429-2011Employee errorThe firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 2 companies; a new generation was cleared every 3.1 years on median.
  • 4 of the 21 ancestors have a recall of any class.
  • Of its descendants, K121874 (Biomet Manufacturing Corp, 2012) has gone furthest, with 53 descendants of its own.
  • 4 other devices also cite its predicate K002379.

How does its category compare?

All KWA devices

KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component): FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · KWA

4in 2012

2007: 82008: 92009: 72010: 62011: 112012: 4
2007high 11 (2011) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWA

76in 2012

2007: 74 days median2008: 57 days median2009: 139 days median2010: 167 days median2011: 101 days median2012: 76 days median
2007high 167 (2010) · low 57 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWA

19in 2023

2008: 132010: 842011: 282012: 212013: 12016: 12017: 12018: 12019: 22020: 12023: 19
2007high 84 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWA

1,216in 2025

2007: 2132008: 7622009: 8202010: 3,3482011: 17,7372012: 24,6012013: 20,0202014: 16,4942015: 12,3392016: 7,8722017: 6,8472018: 6,3882019: 7,6282020: 4,5382021: 2,0642022: 1,6172023: 1,4742024: 1,1682025: 1,216
2007high 24,601 (2012) · low 213 (2007)2025

MAUDE reports naming product code KWA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078740213
200895713762
200971390820
20106167843,348
2011111012817,737
20124762124,601
20130—120,020
20140—016,494
20150—012,339
20160—17,872
20170—16,847
20180—16,388
20190—27,628
20200—14,538
20210—02,064
20220—01,617
20230—191,474
20240—01,168
20250—01,216
20260—0—

What is it indicated for?

1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.

2. Rheumatoid arthritis

3. Correction of functional deformity

4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.

5. Revision procedures where other treatment or devices have failed.

The Porous Plasma Spray (PPS) Ringloc®± Acetabular System is intended for uncemented use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093235: Ringloc + Hybrid Acetabular System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093235

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026