510(k) · Cleared
ClearedJetstream G3 System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Pathway Medical Technologies, Inc. · MCW Catheter, Peripheral, Atherectomy
Cleared on Dec 1, 2009 as substantially equivalent to K092332 (Pathway Medical Technologies, Inc., 2009) after a 26-day review. Its family tree reaches back 12 generations to 1987. 3 of the 59 devices in its family, counting itself, have a Class I recall. 24 later devices descend from it.
- Decided
- Dec 1, 2009
- Review
- 26 days
- Class
- Class II
- Product code
- MCW
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 12 generations to K871011 (1987). 34 ancestors in all, the oldest K780551 (1978). 24 later devices descend from it, 6 citing it directly. 3 of the 59 devices in this family, counting this one, have a Class I recall.
- Root · 12 generations backK871011 Simpson Peripheral Atherocath (TM)Devices For Vascular Intervention, Inc. · 1987
- 8 generations between
- 3 generations backK090325 Jetstream G2 System, Models PV20300 (Jetstream G2 Catheter and…Pathway Medical Technologies, Inc. · 2009
- 2 generations backK091509 Jetstream G2 NXT System, Model PV20300Pathway Medical Technologies, Inc. · 2009
- 1 generation backK092332 Jetstream G3 System, Model PV31300 ( Jetstream G3 Catheter and…Pathway Medical Technologies, Inc. · 2009
- This deviceK093456 Jetstream G3 SystemPathway Medical Technologies, Inc. · 2009
- Descendant · cites it directly · 23 of its ownK093918 Jetstream G3 L System Model PV31400 (Jetstream G3 L Catheter with…Pathway Medical Technologies, Inc. · 2010
- Descendant · cites it directly · 22 of its ownK100462 Jetstream G3 L System, PV Console, Models PV31400, PV31300, PVCN100Pathway Medical Technologies, Inc. · 2010
- Descendant · cites it directly · 20 of its ownK101334 Jetstream G3 SE System, Jetstream G3 SF System Model PV31230…Pathway Medical Technologies, Inc. · 2010
- Descendant · cites it directly · 19 of its ownK101221 Jetstream G3 System, Model PV31300, PV Console, Model PVCN100Pathway Medical Technologies, Inc. · 2010
- Descendant · 2 generations on · 15 of its ownK111229 Jetstream G3 (R) SF 1.6 SystemPathway Medical Technologies, Inc. · 2011
Show all 12 generations
- Root · 12 generations backK871011 Simpson Peripheral Atherocath (TM)Devices For Vascular Intervention, Inc. · 1987
- 11 generations backK024243 Reform Peripheral Catheter System, Model 02200; Reform Peripheral…Fox Hollow Technologies · 2003 · Recalled
- 10 generations backK043553 Silverhawk Peripheral Plaque Excision System, Models 02550,04800…Fox Hollow Technologies · 2005
- 9 generations backK053460 Modification to Silverhawk Peripheral Plaque Excision System, Models…Fox Hollow Technologies · 2006
- 8 generations backK061063 Modification to: Silverhawk Peripheral Plaque Excision System…Foxhollow Technologies, Inc. · 2006
- 7 generations backK081328 Pathway PV Atherectomy SystemPathway Medical Technologies, Inc. · 2008
- 6 generations backK082186 Modification to Pathway PV Atherectomy SystemPathway Medical Technologies, Inc. · 2008
- 5 generations backK082664 Jetstream Pathway PV Atherectomy System, Model PV10300, PVCN100Pathway Medical Technologies, Inc. · 2008
- 4 generations backK083837 Jetstream G2 System, Models PV20300 (Jetstream G2 Catheter and…Pathway Medical Technologies, Inc. · 2009
- 3 generations backK090325 Jetstream G2 System, Models PV20300 (Jetstream G2 Catheter and…Pathway Medical Technologies, Inc. · 2009
- 2 generations backK091509 Jetstream G2 NXT System, Model PV20300Pathway Medical Technologies, Inc. · 2009
- 1 generation backK092332 Jetstream G3 System, Model PV31300 ( Jetstream G3 Catheter and…Pathway Medical Technologies, Inc. · 2009
- This deviceK093456 Jetstream G3 SystemPathway Medical Technologies, Inc. · 2009
Showing 5 of 24 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 24 · See all- K093918Jetstream G3 L System Model PV31400 (Jetstream G3 L Catheter with Control Pod); PVCN100…Pathway Medical Technologies, Inc. · 2010 · cites it directly23descendants
- K100462Jetstream G3 L System, PV Console, Models PV31400, PV31300, PVCN100Pathway Medical Technologies, Inc. · 2010 · cites it directly22descendants
- K101334Jetstream G3 SE System, Jetstream G3 SF System Model PV31230 (Jetstream G3 SE Catheter…Pathway Medical Technologies, Inc. · 2010 · cites it directly20descendants
- K101221Jetstream G3 System, Model PV31300, PV Console, Model PVCN100Pathway Medical Technologies, Inc. · 2010 · cites it directly19descendants
- K111229Jetstream G3 (R) SF 1.6 SystemPathway Medical Technologies, Inc. · 2011 · 2 generations on15descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −12 | K871011 | Simpson Peripheral Atherocath (TM) | Devices For Vascular Intervention, Inc. | 1987 | — |
| −11 | K024243 | Reform Peripheral Catheter System, Model 02200; Reform Peripheral… | Fox Hollow Technologies | 2003 | 1 |
| −10 | K043553 | Silverhawk Peripheral Plaque Excision System, Models 02550,04800, 05200… | Fox Hollow Technologies | 2005 | — |
| −9 | K053460 | Modification to Silverhawk Peripheral Plaque Excision System, Models 02550… | Fox Hollow Technologies | 2006 | — |
| −8 | K061063 | Modification to: Silverhawk Peripheral Plaque Excision System, Models… | Foxhollow Technologies, Inc. | 2006 | — |
| −7 | K081328 | Pathway PV Atherectomy System | Pathway Medical Technologies, Inc. | 2008 | — |
| −6 | K082186 | Modification to Pathway PV Atherectomy System | Pathway Medical Technologies, Inc. | 2008 | — |
| −5 | K082664 | Jetstream Pathway PV Atherectomy System, Model PV10300, PVCN100 | Pathway Medical Technologies, Inc. | 2008 | — |
| −4 | K083837 | Jetstream G2 System, Models PV20300 (Jetstream G2 Catheter and Control Pod)… | Pathway Medical Technologies, Inc. | 2009 | — |
| −3 | K090325 | Jetstream G2 System, Models PV20300 (Jetstream G2 Catheter and Control Pod)… | Pathway Medical Technologies, Inc. | 2009 | — |
| −2 | K091509 | Jetstream G2 NXT System, Model PV20300 | Pathway Medical Technologies, Inc. | 2009 | — |
| −1 | K092332 | Jetstream G3 System, Model PV31300 ( Jetstream G3 Catheter and Control Pod).… | Pathway Medical Technologies, Inc. | 2009 | — |
| 0 (this device) | K093456 | Jetstream G3 System | Pathway Medical Technologies, Inc. | 2009 | — |
| +1 | K093918 | Jetstream G3 L System Model PV31400 (Jetstream G3 L Catheter with Control… | Pathway Medical Technologies, Inc. | 2010 | — |
| +1 | K100462 | Jetstream G3 L System, PV Console, Models PV31400, PV31300, PVCN100 | Pathway Medical Technologies, Inc. | 2010 | — |
| +1 | K101334 | Jetstream G3 SE System, Jetstream G3 SF System Model PV31230 (Jetstream G3 SE… | Pathway Medical Technologies, Inc. | 2010 | — |
| +1 | K101221 | Jetstream G3 System, Model PV31300, PV Console, Model PVCN100 | Pathway Medical Technologies, Inc. | 2010 | — |
| +2 | K111229 | Jetstream G3 (R) SF 1.6 System | Pathway Medical Technologies, Inc. | 2011 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1978 | 1 | 0 |
| 1988 | 3 | 0 |
| 1989 | 1 | 0 |
| 1990 | 2 | 0 |
| 1991 | 1 | 0 |
| 1993 | 1 | 0 |
| 1995 | 1 | 0 |
| 1996 | 2 | 0 |
| 1997 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 2 | 0 |
| 2004 | 1 | 0 |
| 2005 | 1 | 0 |
| 2006 | 2 | 0 |
| 2007 | 1 | 0 |
| 2009 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2010 | 4 | 0 |
| 2011 | 3 | 0 |
| 2012 | 2 | 0 |
| 2013 | 1 | 0 |
| 2014 | 4 | 0 |
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
| 2017 | 1 | 0 |
| 2018 | 3 | 1 |
| 2021 | 3 | 1 |
| 2025 | 1 | 1 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MCW device
- #1 K233668 Auryon Atherectomy Catheter 1.5 mm XL…score 17.6 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K092332 | Jetstream G3 System, Model PV31300 ( Jetstream G3 Catheter and Control Pod)., PVCN100…Summary page 1…MCW Predicate Devices: Manufacturer: Pathway Medical Technologies, Inc. (1) Jetstream Q313! System (K092332) Indications for Use: The Jetstream G3'T" System is intended for use in atherectomy of the peripheral vasculature and to break… | 2009 |
What descends from it? (24)
See all 246 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K093918 | Jetstream G3 L System Model PV31400 (Jetstream G3 L Catheter with Control Pod); PVCN100… | 2010 | 23 |
| K100462 | Jetstream G3 L System, PV Console, Models PV31400, PV31300, PVCN100 | 2010 | 22 |
| K101334 | Jetstream G3 SE System, Jetstream G3 SF System Model PV31230 (Jetstream G3 SE Catheter… | 2010 | 20 |
| K101221 | Jetstream G3 System, Model PV31300, PV Console, Model PVCN100 | 2010 | 19 |
| K120242 | Jetstream Navitus L System | 2012 | 14 |
| K130637 | Jetstream XC 2.4/3.4 System; Jetstream XC 2.1/3.0 System; Jetstream SC 1.85 System… | 2013 | 5 |
Has anything in its family been recalled? (3 Class I)
No FDA recall names this device.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K172315Straub Endovascular SystemStraub Medical AG · 2018 · 3 generations on · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
- K211738Rotarex Atherectomy SystemC.R. Bard, Inc. · 2021 · 4 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
- K242757Rotarex Atherectomy SystemBard Peripheral Vascular, Inc. · 2025 · 5 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 3 months on median.
- 6 of the 34 ancestors have a recall of any class.
- Of its descendants, K093918 (Pathway Medical Technologies, Inc., 2010) has gone furthest, with 23 descendants of its own.
- 5 other devices also cite its predicate K092332.
How does its category compare?
All MCW devicesMCW Catheter, Peripheral, Atherectomy: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.
FDA decisions per year · MCW
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MCW
46in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MCW
2in 2024
Class I (4 since 2007) Class II/III not classified
Adverse event reports per year · MCW
677in 2025
MAUDE reports naming product code MCW, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 5 | 43 | 2 | 50 |
| 2008 | 6 | 33 | 1 | 115 |
| 2009 | 8 | 27 | 3 | 105 |
| 2010 | 6 | 30 | 1 | 167 |
| 2011 | 6 | 39 | 0 | 204 |
| 2012 | 5 | 29 | 0 | 194 |
| 2013 | 5 | 30 | 0 | 134 |
| 2014 | 7 | 117 | 0 | 210 |
| 2015 | 6 | 30 | 0 | 377 |
| 2016 | 5 | 60 | 2 | 409 |
| 2017 | 7 | 91 | 2 | 415 |
| 2018 | 8 | 103 | 0 | 537 |
| 2019 | 3 | 112 | 0 | 618 |
| 2020 | 2 | — | 0 | 631 |
| 2021 | 6 | 119 | 2 (2) | 734 |
| 2022 | 6 | 69 | 2 (2) | 566 |
| 2023 | 4 | 104 | 1 | 574 |
| 2024 | 3 | 41 | 2 | 677 |
| 2025 | 3 | 46 | 0 | 677 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
The Jetstream G3™ System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries ≥ 3.0 mm in diameter. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K093456 · printed Nov 4, 2026
Cite this page
510(k) Report. “K093456: Jetstream G3 System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093456
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026