510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Synchron Systems Enzymatic Creatinine (CR-E) Reagent, Model A60298

K093458510(k) SpecialCleared Dec 4, 200928-day reviewClass II21 CFR 862.1225Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · JFY Enzymatic Method, Creatinine

Cleared on Dec 4, 2009 as substantially equivalent to K091742 (Beckman Coulter, Inc., 2009) after a 28-day review. Its family tree reaches back 1 generation to 2009. It has been recalled once.

Decided
Dec 4, 2009
Review
28 days
Class
Class II
Product code
JFY
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K091742 (2009). No later device cites it as a predicate yet.

  1. Root · 1 generation backK091742 Synchron Systems Enzymatic Creatinine (CR-E) Reagent, Model A60298Beckman Coulter, Inc. · 2009 · Recalled
  2. This deviceK093458 Synchron Systems Enzymatic Creatinine (CR-E) Reagent, Model A60298Beckman Coulter, Inc. · 2009
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K091742Synchron Systems Enzymatic Creatinine (CR-E) Reagent, Model A60298Beckman Coulter, Inc.20091
0 (this device)K093458Synchron Systems Enzymatic Creatinine (CR-E) Reagent, Model A60298Beckman Coulter, Inc.2009—
FIG. 1 — Lineage of K093458, 2009–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JFY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JFY device

  1. #1 K252619 Qscheck Uisacrscore 12.4 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K091742Synchron Systems Enzymatic Creatinine (CR-E) Reagent, Model A60298
Summary page 1
…s substantial equivalence to the Beckman Coulter SYNCHRON Systems Enzymatic Creatinine (CR-E) reagent currently in commercial distribution, FDA 510(k) Number K091742. Candidate(s) Predicate Manufacturer Docket SYNCHRON Systems…
Beckman Coulter, Inc.20091

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 3, 2017Class IIZ-2080-2017Under Investigation by firmBeckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.

How does its category compare?

All JFY devices

JFY Enzymatic Method, Creatinine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · JFY

2in 2021

2007: 82008: 42009: 32011: 22014: 22016: 12017: 12018: 22019: 22021: 22026 (partial year): 1
2007high 8 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JFY

59in 2009

2007: 193 days median2008: 170 days median2009: 59 days median
2007high 193 (2007) · low 59 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JFY

1in 2013

2008: 22009: 32010: 32013: 1
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JFY

61in 2025

2007: 12008: 42009: 222010: 222011: 422012: 242013: 392014: 42015: 72016: 162017: 162018: 342019: 562020: 342021: 572022: 92023: 362024: 342025: 61
2007high 61 (2025) · low 1 (2007)2025

MAUDE reports naming product code JFY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007819301
2008417024
2009359322
20100—322
20112—042
20120—024
20130—139
20142—04
20150—07
20161—016
20171—016
20182—034
20192—056
20200—034
20212—057
20220—09
20230—036
20240—034
20250—061
20261—0—

What is it indicated for?

CR-E reagent, when used in conjunction with SYNCHRON® Systems, UniCel® DxC Systems and SYNCHRON® Systems AQUA CAL 1 and 2 and SYNCHRON CX® Calibrator Level 1 and 2, is intended for the quantitative determination of creatinine (CR-E concentration in human serum, plasma or urine (urine is not available on the SYNCHRON CX® PRO Systems).

Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093458: Synchron Systems Enzymatic Creatinine (CR-E) Reagent, Model A60298.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093458

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026