510(k) Report

510(k) · Cleared

Cleared

Alligator-HD Retrieval Device (Ard-HD), Model FA-88840-XX

K093977510(k) TraditionalCleared Apr 22, 2010119-day reviewClass II21 CFR 870.1250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Chestnut Medical Technologies, Inc. · DQY Catheter, Percutaneous

Cleared on Apr 22, 2010 as substantially equivalent to K043580 (Chestnut Medical Technologies, Inc., 2005) and 2 other predicates after a 119-day review. Its family tree reaches back 2 generations to 1991. 1 of the 7 devices in its family, counting itself, has a Class I recall.

Decided
Apr 22, 2010
Review
119 days
Class
Class II
Product code
DQY
Recalls
None
Predicates
3

Its longest line of descent runs back 2 generations to K914067 (1991). 6 ancestors in all. No later device cites it as a predicate yet. 1 of the 7 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year cleared 2005: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor1993: 2 other ancestorsK914067 · Retrieval Device · Target Therapeutics · 1991K970668 · The Amplantz Goose Neck Microsnare · Microvena Corp. · 199819901995200020052010
Other ancestors, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 2 other ancestors K970668 · The Amplantz Goose Neck Microsnare · Microvena Corp. · 1998K914067 · Retrieval Device · Target Therapeutics · 19910
  1. Root · 2 generations backK914067 Retrieval DeviceTarget Therapeutics · 1991
  2. 1 generation backK970668 The Amplantz Goose Neck MicrosnareMicrovena Corp. · 1998
  3. This deviceK093977 Alligator-HD Retrieval Device (Ard-HD), Model FA-88840-XXChestnut Medical Technologies, Inc. · 2010

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K914067Retrieval DeviceTarget Therapeutics1991—
−1K970668The Amplantz Goose Neck MicrosnareMicrovena Corp.1998—
0 (this device)K093977Alligator-HD Retrieval Device (Ard-HD), Model FA-88840-XXChestnut Medical Technologies, Inc.2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199320
200210
200511
FIG. 1 — Lineage of K093977, 1991–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQY device

  1. #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K043580Alligator Retrieval Device (Ard)
Summary page 1
…D Retrieval Device is intended for use in the peripheral, neuro and cardiovasculature for foreign body retrieval. Predicate Device: AlligatorTm Retrieval Device; 510k # K043580, Boston Scientific, Target In-Time Retrieval Device; Sl1Ok# K014109, Microvena Corp. Amplatz Microsnare, Sl1Ok# K970668 Comparis…
Chestnut Medical Technologies, Inc.2005Class I
*K014109IN-Time Retrieval Device
Summary page 1
…cardiovasculature for foreign body retrieval. Predicate Device: AlligatorTm Retrieval Device; 510k # K043580, Boston Scientific, Target In-Time Retrieval Device; Sl1Ok# K014109, Microvena Corp. Amplatz Microsnare, Sl1Ok# K970668 Comparison to Predicate Devices to Support Substantial Equivalence Determina…
Boston Scientific, Target2002—
*K970668The Amplantz Goose Neck Microsnare
Summary page 1
…cate Device: AlligatorTm Retrieval Device; 510k # K043580, Boston Scientific, Target In-Time Retrieval Device; Sl1Ok# K014109, Microvena Corp. Amplatz Microsnare, Sl1Ok# K970668 Comparison to Predicate Devices to Support Substantial Equivalence Determination: The information presented in the 510k shows t…
Microvena Corp.1998—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K043580Alligator Retrieval Device (Ard)Chestnut Medical Technologies, Inc. · 2005 · Direct predicate · ◆ in the 2005 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 19 years and passes through 3 companies; a new generation was cleared every 11.9 years on median.
  • 1 of the 6 ancestors has a recall of any class; 1 has a Class I recall.

How does its category compare?

All DQY devices

DQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · DQY

21in 2025

2007: 162008: 312009: 372010: 222011: 362012: 302013: 322014: 192015: 272016: 222017: 262018: 252019: 302020: 202021: 182022: 182023: 182024: 162025: 212026 (partial year): 11
2007high 37 (2009) · low 16 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQY

115in 2025

2007: 85 days median2008: 66 days median2009: 73 days median2010: 74 days median2011: 71 days median2012: 87 days median2013: 92 days median2014: 127 days median2015: 118 days median2016: 117 days median2017: 113 days median2018: 106 days median2019: 111 days median2020: 120 days median2021: 81 days median2022: 74 days median2023: 114 days median2024: 87 days median2025: 115 days median2026 (partial year): 30 days median
2007high 127 (2014) · low 66 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQY

1in 2025

2007: 62008: 122009: 172010: 322011: 62012: 22013: 36, 1 Class I2014: 32015: 32016: 22, 7 Class I2017: 82018: 22019: 12020: 22021: 22023: 172024: 12025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 36 (2013) · low 0 (2022)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · DQY

3,056in 2025

2007: 3132008: 4962009: 8192010: 1,2932011: 1,2112012: 1,1912013: 1,3522014: 1,4972015: 2,3372016: 2,2522017: 2,4892018: 2,8552019: 2,9022020: 2,8892021: 3,0062022: 2,6432023: 3,0492024: 3,1902025: 3,056
2007high 3,190 (2024) · low 313 (2007)2025

MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200716856313
2008316612496
2009377317819
20102274321,293
2011367161,211
2012308721,191
2013329236 (1)1,352
20141912731,497
20152711832,337
20162211722 (7)2,252
20172611382,489
20182510622,855
20193011112,902
20202012022,889
2021188123,006
2022187402,643
202318114173,049
2024168713,190
2025211151 (1)3,056
202611302 (1)—

What is it indicated for?

The AlligatorTM-HD Retrieval Device is intended for use in the peripheral, neuro and

cardiovasculature for foreign body retrieval.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093977: Alligator-HD Retrieval Device (Ard-HD), Model FA-88840-XX.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093977

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026